Equipment and Calibration History

View Equipment |

View history entry 5087 for Global Water WL16U sn: 57278

Date: 2018-05-08
From: Lab - none
To: Lab - none
Title: Repair
Notes: I stripped off the old rotting insulation and then slid on a long continuous piece of heat shrink. I shrunk the whole length of the cable. I disassembled the data logging enclosure and kept only the data logger board and USB connector in the end cap. I cut off the twin 9V battery plugs and will power the sensor from solar connected rechargeable battery.
Entered by: Joe Verfaillie