Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 [50] 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70
Ser No.DateLoc FromTitleLoc ToNotes
LI-820 CGA-556 2016-02-23 Lab none Installed Tonzi-Tower Reinstalled in the Tower CO2 profiles system, but the pump was not working and was removed. The IRGA was left unpowered.
GMM222 L2010070 2016-02-23 Lab none Installed Vaira CO2 probe North Profile -0.08 Installed in the Vaira North profile at -8cm after lab calibration.
GMM222 L2010068 2016-02-23 Lab none Installed Vaira CO2 probe North Profile -0.04 Installed in the Vaira North profile at -4cm after lab calibration.
GMT222 G0230062 2016-02-23 Lab none Installed Vaira CO2 probe North Profile -0.02 Installed in the Vaira North profile at -2cm after lab calibration.
GMP222 K1350101 2016-02-23 Lab none Installed Vaira CO2 probe West Profile -0.08 Installed in the Vaira West profile at -8cm after lab calibration.
GMP222 K1350102 2016-02-23 Lab none Installed Vaira CO2 probe West Profile -0.04 Installed in the Vaira West profile at -4cm after lab calibration.
GMP222 K1350103 2016-02-23 Lab none Installed Vaira CO2 probe West Profile -0.02 Installed in the Vaira West profile at -2cm after lab calibration.
GMP222 Y2030049 2016-02-23 Lab none Installed Vaira CO2 probe East Profile -0.08 Installed in the Vaira East profile at -8cm after lab calibration. Using a new cable.
GMP222 K1310103 2016-02-23 Lab none Installed Vaira CO2 probe East Profile -0.04 Installed in the Vaira East profile at -4cm after lab calibration.
GMP222 Y2009109 2016-02-23 Lab none Installed Vaira CO2 probe East Profile -0.02 Installed in the Vaira East profile at -2cm after lab calibration.
LI-7500 0976 2016-02-23 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira tower to replace sn 0041 for calibration.
LI-7500A 0041 2016-02-23 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira tower for calibration. Replaced by sn 0976.
LI-7500A 75H-2514 2016-02-11 Lab none Installed East End Gas Analyzer CO2/H2O Installed on the East End tower to replace sn 75H-2176 for calibration
LI-7500A 75H-2176 2016-02-11 East End Gas Analyzer CO2/H2O Removed Lab none Removed from the East End tower for calibration. Replaced by sn 75H-2514
LI-7500A 75H-2180 2016-02-11 Lab none Installed Twitchell Alfalfa Gas Analyzer CO2/H2O Installed on the Alfalfa Tower to replace sn 1594 for calibration.
LI-7500A 1594 2016-02-11 Twitchell Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from the Alfalfa Tower for calibration. Replaced by sn 75H-2180.
GMT221 F4960018 2016-02-11 Lab none Installed Twitchell Alfalfa Installed at Alfalfa Profile A at 8cm. The existing probe sn Y0710013 was removed due to low readings.
GMP221 Y0710013 2016-02-11 Twitchell Alfalfa Removed Lab none Removed from Alfalfa Profile A at 8cm due to low readings. Replaced by sn F4960018
GMT221 F4960020 2016-02-04 Tonzi CO2 Probe Tree2 Profile -0.08 Removed Lab none Removed from Tonzi Tree 2 profile at -8cm for calibration
GMP222 Z2720012 2016-02-04 Tonzi CO2 Probe Tree2 Profile -0.04 Removed Lab none Removed from Tonzi Tree 2 profile at -4cm for calibration
GMP222 Z2720014 2016-02-04 Tonzi CO2 Probe Tree2 Profile -0.02 Removed Lab none Removed from Tonzi Tree 2 profile at -2cm for calibration
GMT221 G0150036 2016-02-04 Tonzi CO2 Probe Tree1 Profile -0.08 Removed Lab none Removed from Tonzi Tree profile at -8cm for calibration
GMP222 Z2720013 2016-02-04 Tonzi CO2 Probe Tree1 Profile -0.04 Removed Lab none Removed from Tonzi Tree profile at -4cm for calibration
GMM222 L2010069 2016-02-04 Tonzi CO2 Probe Tree1 Profile -0.02 Removed Lab none Removed from Tonzi Tree profile at -2cm for calibration
GMP222 K1310104 2016-02-04 Tonzi CO2 Probe Open Profile -0.08 Removed Lab none Removed from Tonzi Open profile at -8cm for calibration
GMP222 K1310103 2016-02-04 Tonzi CO2 Probe Open Profile -0.04 Removed Lab none Removed from Tonzi Open profile at -4cm for calibration
GMP222 K1310102 2016-02-04 Tonzi CO2 Probe Open Profile -0.02 Removed Lab none Removed from Tonzi Open profile at -2cm for calibration
GMM222 L2010070 2016-02-04 Vaira CO2 probe North Profile -0.08 Removed Lab none Removed from Vaira North profile at -8cm for calibration
GMM222 L2010068 2016-02-04 Vaira CO2 probe North Profile -0.04 Removed Lab none Removed from Vaira North profile at -4cm for calibration
GMT222 G0230062 2016-02-04 Vaira CO2 probe North Profile -0.02 Removed Lab none Removed from Vaira North profile at -2cm for calibration
GMP222 K1350101 2016-02-04 Vaira CO2 probe West Profile -0.08 Removed Lab none Removed from Vaira West profile at -8cm for calibration
GMP222 K1350102 2016-02-04 Vaira CO2 probe West Profile -0.04 Removed Lab none Removed from Vaira West profile at -4cm for calibration
GMP222 K1350103 2016-02-04 Vaira CO2 probe West Profile -0.02 Removed Lab none Removed from Vaira West profile at -2cm for calibration
GMP222 Z2720015 2016-02-04 Vaira CO2 probe East Profile -0.04 Removed Lab none Removed from Vaira East profile at -4cm for calibration
GMP221 L2058003 2016-01-21 Twitchell Alfalfa Removed Lab none Removed from Alfalfa set A at 8cm due to maxed out readings.
WM 1590 153805 2016-01-12 Lab none Installed Twitchell Alfalfa Anemometer Installed to replace old sonic--brand new
WM 1590 120908 2016-01-12 Twitchell Alfalfa Anemometer Removed Lab none Removed from field for calibration
WM 1590 153804 2016-01-12 Lab none Installed East End Anemometer Installed at East End, brand new
WM 1590 124215 2016-01-12 East End Anemometer Removed Lab none Removed from tower for calibration. The SOS was reading unexpectedly low.
LI-7500A 75H-2150 2015-12-28 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed on the Mayberry Tower to replace sn 75H-2514 for calibration.
LI-7500A 75H-2514 2015-12-28 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from the Mayberry tower for calibration. Replaced by sn 75H-2150
FGGA 09-0010 2010-04-08 Twitchell Rice Calibration Twitchell Rice Last calibration: 2010-03-11
Cell Pressure=134.99Torr, Cell Temperature=26.64C, Ring Time=14.014usec
2nd Standard at 1900psi CH4 Span=3.276ppm, CO2 Span=433.0ppm
CH4 = 3.302 -> 3.305 Offset: -0.026ppm
CO2 = 436 -> 439 Offset: -3ppm
Notes: Seems to have made it worse?
FGGA 09-0010 2010-03-03 Lab none Calibration Lab none Last calibration: 2010-01-14
Cell Pressure=138.16Torr, Cell Temperature=20.17C, Ring Time=13.987usec
God CH4 at psi CH4 Span=1.8978ppm, CO2 Span=382.50ppm
CH4 = 1.834 -> 1.903 Offset: 0.064ppm
CO2 = 354 -> 383 Offset: 28.5ppm
Notes: Testing with CH4 standard 4.978ppm CH4 and 3ppm CO2 LGR reading: 4.785ppm +- 9ppb CH4 CO2 reading -1.00 Maybe out of range Testing with CO2 standard 597.4ppm CO2 ??ppm CH4 LGR reading: 2.701ppm +- 9ppb CH4 LGR reading: 582ppm +-4ppm CO2 This analyzer seems noisier than the single channel. Will check other FGGA analyzer after swap.
LI-7500A 75H-2669 2015-12-23 Twitchell Rice-Trailer Removed Lab none Removed with the temporary tower set up near the building at the rice project for the LGR COS experiment.
LI-7700 TG1-0424 2015-12-23 Twitchell Rice-Trailer Removed Lab none Removed with the temporary tower set up near the building at the rice project for the LGR COS experiment.
LI-7550 AIU-1570 2015-12-23 Twitchell Rice-Trailer Removed Lab none Removed with the temporary tower set up near the building at the rice project for the LGR COS experiment. This box had a significant amount of water inside. There was some corrosion on the power connectors but it looks like it was running up until I turned it off. I blew out all the connectors and stuffed it full of desiccant in the lab.
425A 0843 2015-12-23 Tonzi Removed Lab none Removed from the 2nd from top location of the wind profile near the Floor tower.
NL115 13053 2015-12-23 Tonzi Removed Lab none Removed from the box for the wind profile near the Floor tower.
CR1000 66250 2015-12-23 Tonzi Removed Lab none Removed from the box for the wind profile near the Floor tower.
425A 0846 2015-12-23 Tonzi Removed Lab none Removed from the top location of the wind profile near the Floor tower.
425A 0839 2015-12-23 Tonzi Removed Lab none Removed from the 3rd from bottom location of the wind profile near the Floor tower.
425A 0852 2015-12-23 Tonzi Removed Lab none Removed from the bottom location of the wind profile near the Floor tower.
SATI-3K 000603 2012-09-21 Borrowed Borrowed Lab none Borrowed from Robin in Allen Goldstein's lab for an inter-comparison
PAR-LITE 010163 2015-03-05 Lab none Installed Mayberry PAR Sensor Reflected Veg Installed on the Mayberry tower (over tules) but it was reading high when it was connected to the multiplexor. The mux channel was even read high when jumped - so may be bad channel on mux. The PAR sensor was left on the tower and the mux or program will be changed to deal with the bad channel.
HP-127-1.0-1.3-71P P01 2015-12-21 Lab none Repair Lab none The red lead looked chewed and was broken at its junction to the plate. I soldered a new red wire directly to the junction and a new black wire about 3cm from the plate. Both wires were covered with heat shrink and the junctions to the plate were painted with black Rustoleum paint.
BB-Fab BB-TW 2014-09-10 Twitchell Rice Upgrade Twitchell Rice We moved the three old group 8D out of the way and installed the four group 27 batteries. I had two older batteries in the lab and bought two more. We also installed a new solar charger under the solar panels.
LI-7500A 0065 2015-12-07 Lab none Calibration Lab none Last calibration: 2015-04-20
AGC=76.5 SDM=1 DAC1=- DAC2=-
Ambient P=100.92kPa, Tair=19.81C, Tdew=14.18C
Zero Air at 650psi, God CO2 421.22 at 700 Tdewspan=11.09
CO2 zero 0.8897 -> 0.8900 Offset: 1.20ppm
CO2 span 1.0074 -> 1.0164 Offset: -4.88ppm
H2O zero 0.7055 -> 0.7097 Offset: 31.2ppm
H2O span 1.0071 -> 1.0231 Offset: -0.32ppm
Notes:
LI-7500A 75H-2150 2015-12-07 Lab none Calibration Lab none Last calibration: 2015-07-15
AGC=98.6 SDM=1 DAC1=- DAC2=-
Ambient P=100.94kPa, Tair=19.78C, Tdew=12.44C
Zero Air at 650psi, God CO2 421.22 at 700 Tdewspan=11.09
CO2 zero 0.9021 -> 0.9019 Offset: -1.16ppm
CO2 span 1 -> 1.0114 Offset: -6.07ppm
H2O zero 0.8671 -> 0.8671 Offset: 0.3ppm
H2O span 0.9986 -> 1.0258 Offset: -0.52ppm
Notes:
GP-001 GP20150100 2015-12-08 Lab none Installed West Pond CO2 Probe B This 8% probe was installed at West Pond to replace the 4% probe sn GP20140022
GP-001 GP20140022 2015-12-08 West Pond CO2 Probe B Removed Lab none This 4% probe was removed from West Pond. It was replaced by an 8% probe sn GP20150100
LI-7500A 0065 2015-12-08 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed on the West Pond tower to replace sn 75H-2180 for calibration.
LI-7500A 75H-2180 2015-12-08 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from the West Pond tower for calibration. Replaced by sn 0065
GP-001 GP20150109 2015-12-08 Lab none Installed East End CO2 probe This 8% probe was installed at East End to replace a 4% probe sn GP20140021
GP-001 GP20140021 2015-12-08 East End CO2 probe Removed Lab none This 4% CO2 probe was removed from East End. It was replaced by an 8% probe sn GP20150109
KD140SX-UFBS 11ZV0M1127 2015-12-08 Lab none Installed East End Solar Panel C This solar panel was added to the existing two panels.
WM 1590 153805 2015-12-08 Lab none Installed Twitchell Rice-Trailer Installed near the trailer at Twitchell Rice as a temporary site to test a COS analyzer from LGR
LI-7500A 75H-2669 2015-12-08 Lab none Installed Twitchell Rice-Trailer Installed near the trailer at Twitchell Rice as a temporary site to test a COS analyzer from LGR
LI-7550 AIU-1570 2015-12-08 Lab none Installed Twitchell Rice-Trailer Installed near the trailer at Twitchell Rice as a temporary site to test a COS analyzer from LGR
LI-7700 TG1-0424 2015-12-08 Lab none Installed Twitchell Rice-Trailer Installed near the trailer at Twitchell Rice as a temporary site to test a COS analyzer from LGR
CS511 DO-8 2015-11-25 Lab none Installed Twitchell Rice DO Sensor Reinstalled at the Rice tower after being removed for fall draw down.
HP-127-1.0-1.3-71P P04 2015-12-01 Lab none Repair Lab none The red wire had broken right where it met the plate. I was able solder on a new wire. I covered the wires with heat shrink and painted the connections to the plate.
HP-127-1.0-1.3-71P P03 2015-12-01 Lab none Repair Lab none I repaired a wire that was chewed by a gopher. I covered the wires with heat shrink and painted the connections between the wires and the plate.
HP-127-1.0-1.3-71P P04 2015-11-20 Vaira Removed Lab none Removed from Vaira at -16cm. Looks like the insulation had cracked and the wire rotted.
HP-127-1.0-1.3-71P P03 2015-11-20 Vaira Removed Lab none Removed from Vaira at -8cm. A gopher had chewed its wire.
LI-7500A 75H-2182 2015-11-25 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Installed on the Twitchell Rice tower to replace sn 0065 for calibration
LI-7500A 0065 2015-11-25 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none Removed from the Twitchell Rice tower for calibration. REplaced by sn 75H-2182
LI-7500A 1594 2015-11-25 Lab none Installed Twitchell Alfalfa Gas Analyzer CO2/H2O Installed on the Alfalfa Tower to replace sn 75H-2150 for calibration
LI-7500A 75H-2150 2015-11-25 Twitchell Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from the Alfalfa Tower for calibration. Replaced by sn 1594
HP-127-1.0-1.3-71P P01 2015-12-02 Vaira Removed Lab none Removed from Vaira at -2cm. The red wire was broken not clear whether corrosion or animal (or shovel) damage
HP-127-1.0-1.3-71P P04 2015-12-02 Lab none Installed Vaira Reinstalled at Vaira at -16cm. This was spliced onto the existing cable in the field.
HP-127-1.0-1.3-71P P03 2015-12-02 Lab none Installed Vaira Reinstalled at Vaira at -8cm. This was spliced onto the existing cable in the field.
CS511 DO-8 2015-11-23 Lab none Serviced Lab none Serviced probe to go back in the field this week. The new calibration is:
M = 9.45/36.1 in (mg/L)/mV (aka ppm/mV) (@ 18C).
GMT221 G0150037 2014-08-07 Sherman Island Removed Lab none Removed with the rest of the soil CO2 system from the Sherman site to be redeployed at the Alfalfa site on Twitchell
WMP 1352 000020 2001-07-02 Lab none Installed Tonzi-Tower Installed on the Tonzi Tower to replace existing sonic that has had suspicious performance.
WMP 1352 000024 2001-01-05 Lab none Installed Vaira Anemometer Installed on the Vaira Tower to replace sn 000019 that was going off scale at midday.
WMP 1352 000019 2001-01-05 Vaira Anemometer Removed Lab none Removed from the Vaira Tower due to going off scale at midday. Replaced by sn 000024
WMP 1352 000019 2000-10-22 Lab none Installed Vaira Anemometer Installed on the Vaira Tower with the initial setup of the site.
LI-7500A 75H-2182 2015-11-10 Lab none Calibration Lab none Last calibration: 2015-07-15
AGC=87.4 SDM=6 DAC1=- DAC2=-
Ambient P=100.94kPa, Tair=21.68C, Tdew=6.22C
Zero Air at 700psi, God CO2 421.22 at 700 Tdewspan=11.08
CO2 zero 0.8825 -> 0.8833 Offset: 3.48ppm
CO2 span 1.0064 -> 1.0197 Offset: -7.02ppm
H2O zero 0.8458 -> 0.8464 Offset: 4.0ppm
H2O span 1.0016 -> 1.0242 Offset: -0.43ppm
Notes:
LI-7500A 1594 2015-11-16 Lab none Calibration Lab none Last calibration: 2015-04-20
AGC=101.9 SDM=1 DAC1=- DAC2=-
Ambient P=101.42kPa, Tair=20.60C, Tdew=-2.88C
Zero Air at 700psi, God CO2 421.22 at 700 Tdewspan=11.08
CO2 zero 0.8658 -> 0.8662 Offset: 1.43ppm
CO2 span 1.0027 -> 1.0243 Offset: -11.42ppm
H2O zero 0.8855 -> 0.8821 Offset: -18.0ppm
H2O span 0.9935 -> 1.0212 Offset: -0.54ppm
Notes: This is the first calibration of this head after it was upgraded to the new 7500A interface and had its chemicals changed. Old coefficients from last Licor calibration loaded were loaded during the upgrade. Last lab calibration coefficients: CO2_zero = 0.8648, CO2_span = 1.0178, H2O_zero = 0.8985, H2O_span = 1.0087.
GMT221 G0150037 2014-06-17 Lab none Installed Sherman Island Installed in Sherman profile A at -16cm to replace sn F1340006 that has been reading low.
GMP221 K1340006 2014-06-17 Sherman Island Removed Sherman Island Removed from Sherman profile A at -16cm due to low readings. Replaced by sn G0150037
GMP221 K1340006 2014-04-10 Lab none Installed Sherman Island Installed in Sherman profile A at -16cm to replace sn G0150035
GMT221 G0150035 2014-04-10 Sherman Island Removed Lab none Removed from Sherman profile A at -16cm, replaced with sn K1340006
GMP221 K1340007 2014-04-10 Lab none Installed Sherman Island Installed in Sherman profile A at -8cm to replace sn F4960019
GMT221 F4960019 2014-04-10 Sherman Island Removed Lab none Removed from Sherman profile A at -8cm, replaced with sn K1340007
GMP221 K1340005 2014-04-10 Lab none Installed Sherman Island Installed in Sherman profile A at -2cm to replace sn H1630066
GMP222 H1630066 2014-04-10 Sherman Island Removed Lab none Removed from Sherman profile A at -2cm, replaced with sn K1340005
ML2x 64/068 2014-04-10 Sherman Island Location Lab none Third probe on data logger, notes claim it is at -10cm
ML2x 70/002 2014-04-10 Sherman Island Location Sherman Island First probe on data logger, notes claim it is at -20cm
GMT221 F4960018 2014-01-23 Sherman Island Location Sherman Island Sherman Profile B -8cm