Skip to main content
rice field banner biometlab logo
   
   
 
   
   
   
   
   
 
 

Field Notes

<--2016-04-01 10:00:00 | 2016-04-28 09:25:00-->
Other sites visited today: Twitchell Alfalfa | West Pond | East End | Twitchell Rice

Mayberry: 2016-04-13 09:05:00 (DOY 104)
Author: Elke Eichelmann
Others: Joe Verfaillie

Summary: Calibrated LI7700, swapped LI7500, conductivity sensor not working

Arrived at 9:05. Sunny, clear, breezy with gusts. Tules and cattails are quite tall now, there is lots of thick grass on the berms, really need a truck to drive out here.

LI7500 readings (all H2O readings were taken in mmol/mol at this site today):
CO2 410.73
H2O 12.43
T 15.64
P 102.08
SS 96.253
SOS 341.797

LI7700 readings after cleaning (did not take a reading before cleaning today):
CH4: 1.911
RSSI: 78.29

Collected USB GHG data, camera pictures, and Met data. HMP pressure reading looks fine, reads
102kPa, in line with the reading from the LI7550 pressure sensor.
Ts_16 reading INF; CondTemp and TempCondA also reading INF, Conductivity reading also weird. Conductivity probe does not seem to be working correctly, removed probe sn 6097 and interface sn 3270 and brought back to the lab for testing.

Updated LI7550 sn AIU-1285 box firmware from v7.3.0 to v8.0.0
Refilled LI7700 wash water reservoir.

Calibrated LI7700 sn TG1-0357. Calibration tube fell in the water before we got started, was recovered, but dripping wet. Dried the tube as best as we could. Should not affect the CH4 calibration.
Calibration was done using zero air and the Ameriflux AMPCO2CH4-calibration gas at 1.9966ppm
After calibration LI7700 read:
CH4: 1.933
RSSI: 77.65

Swapped the LI7500 head, took down sn 75H-2150 for calibration in the lab, replaced with sn 75H-2182.
LI7500 readings (all H2O readings were taken in mmol/mol at this site today):
old (sn 2150) new (sn 2182)
CO2 420.21 411.72
H2O 11.51 11.03
T 16.65 17.03
P 102.08 102.23
SS 88.91 85.835
SOS 342.146 342.405

We left again at approximately 9:45

0 photos found

Graphs display:
11 sets found

7 MB_flux graphs found

Sonic Wind

LI7700 mV signals

Counts

LI7700 CH4 from analog signal

LI7500 CO2 and H2O

LI7700 CH4 from digital signal

CO2 in ppm

No data found for MB_met.
27 MB_tule graphs found

MB Tule Tower: Open Water 2 (3/1/11-3/23/17)

MB Tule Tower: Open Water 1 (3/1/11-2/21/13)

MB Tule Tower: Tule Temps 2 & 3 (3/1/11-3/19/14)

MB Tule Tower: Tule Temps 1 (3/1/11-3/19/14)

Mayberry Tule Tower: Solar Radiation

Mayberry Tule Tower: Disolved Oxygen

Mayberry Tule Tower: IRT and Sensor Temp

Mayberry Tule Tower: Volts and RH

LED NDVI Sensor

Water CO2

Water Pressure

Fluorometer Data

Mayberry pCO2 Profile

Water Temperatures

Air Temperatures

Air Pressure

4-Way Radiometer

MB Tule Tower: Open Water 1 (2/21/13-3/19/14)

MB Tule Tower: Tule Temps 1 (3/19/14-now)

MB Tule Tower: Open Water 1 (3/19/14-now)

MB Tule Tower: Tule Temps 2 & 3 (3/19/14-3/23/17)

MB Tule Tower: Tule Temps 2 (3/23/17-now)

MB Tule Tower: Tule Temps 3 (3/23/17-now)

SRS NDVI Channels

SRS NDVI

Net Radiation

Conductivity

Can't check MBWPEESWGT_mixed
1 MB_cam graphs found

Mayberry Autocam GCC

Can't check MBEEWPSWEPGT_cam
Can't check SIMBTWEE_met
1 MB_minidot graphs found

Mayberry miniDot Dissolved Oxygen

Can't check MBSWEEWPBCBA
1 MB_processed graphs found

Can't check MB_fd
1 MB_fd graphs found

Water Forced Diffusion CO2 Flux

 
 

Login
Select Notes:

This material is based upon work supported by the National Science Foundation and US Department of Energy. Any opinions, findings, conclusions, or recommendations expressed in the material are those of the author(s) and do not necessarily reflect the views of the supporters.