Skip to main content
rice field banner biometlab logo
   
   
 
   
   
   
   
   
 
 

Equipment and Calibration

View History

Equipment Entry

Instrument:CO2 Probe
Manufacture:Vaisala
Model No:GMP222
Serial No:K1310104
Tag No:
Acquisition Date:2014-04-03
Project:DOE
Location on 2022-10-05:Lab,
Notes: 0-1% range

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
6027 2022-10-05 Tonzi-SoilCO2 CO2 Probe Open Profile -0.02 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Open -2cm
3054 2017-09-14 Lab none Installed Tonzi CO2 Probe Open Profile -0.02 Installed at Tonzi in the Open profile at -2cm after lab calibration. This is a new installation with new holes dug.
3016 2017-08-25 Tonzi CO2 Probe Tree2 Profile -0.04 Removed Lab none Removed from Tonzi Tree 2 profile at -4 cm position for calibration
2195 2016-02-23 Lab none Installed Tonzi CO2 Probe Tree2 Profile -0.04 Installed in the Tonzi Tree 2 profile at -4cm after lab calibration.
2159 2016-02-04 Tonzi CO2 Probe Open Profile -0.08 Removed Lab none Removed from Tonzi Open profile at -8cm for calibration
2072 2014-12-09 Tonzi CO2 Probe Open Profile -0.08 Location Tonzi CO2 Probe Open Profile -0.08 Tonzi Open profile at -8cm
2015 2014-04-08 Lab none Installed Tonzi CO2 Probe Open Profile -0.02 Installed at Tonzi in the Open profile at -2cm. not clear if it replaced another probe.
1448 2014-04-07 Lab none Calibration Lab none Trueppm = Probeppm*1.008 + 43.09; linear relationship between the ppm reported by the probe and the known ppm of the calibration tanks. Tanks: zero air (0ppm), 2238ppm, 3056ppm, and 10000ppm. Ta=25.8C, P=101.9kPa. Because the lab conditions were close to factory conditions, no temperature or pressure corrections were applied. Done by: Joe.
 
 

Login