Equipment and Calibration History

View Equipment |

View history entry 2224 for Licor LI-6400 sn: PSC-0785

Date: 2016-04-04
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Patty Oikawa 6400 was giving "IRGA's not ready" error in field Discovered mirrors needed to be cleaned AGC scores-- Before cleaning: CO2_S:4937 H2O_S:-1500 After cleaning: CO2_S=2652 H2O_S=399.7 *protocol for cleaning mirrors is in the cabinet in downstairs lab Then I zeroed and spanned: Zero: Before: CO2_S:13.8 H2O_S:-0.997 After: CO2_S:0.0 H2O_S:0.003 Span with 421.22ppm cal gas: Before: CO2_S: 425.4 After: CO2_S: 421.1 (setting 0.998) Span with 11.08C dew point: Before: H2O_S:12.18 After: H2O_S: 11.08 (setting 1.039)
Entered by: Patty Oikawa