Equipment and Calibration History

View Equipment |

View history entry 2282 for Licor LI-820 sn: CGA-556

Date: 2016-05-26
From: Tonzi-Tower -
To: Tonzi-Tower -
Title: Calibration
Notes: Last calibration was 2016-02-09 Cell T = 52.2, Pressure = 92.0 Zero Air offset = 16.4ppm Span CO2 405ppm offset = -19.4ppm Zero 0.9423 -> 0.9461 Span 0.8220 -> 0.8499 Changed Pressure output range from 1.33331998E3 to 115kPa
Entered by: Joe Verfaillie