Equipment and Calibration History
Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 [40] 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70Ser No. | Date | Loc From | Title | Loc To | Notes |
---|---|---|---|---|---|
GMP222 K1310102 | 2017-11-29 | Vaira CO2 probe East Profile -0.04 | Removed | Lab none | Removed from the 4cm location in the East profile at Vaira. This probe was giving erratic values |
LI-7500A 0065 | 2017-11-28 | Lab none | Calibration | Lab none | Last calibration: 2017-05-30 AGC=100.0 SDM=0 DAC1=0 DAC2=0 Ambient P=100.5kPa, Tair=23.0C, Tdew=8.7C Zero Air at 300psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.8901 -> .8906 Offset: 2.12ppm CO2 span 1.013 -> 1.0163 Offset: -1.35ppm H2O zero 0.7087 -> .7083 Offset: -3.0ppm H2O span 1.0166 -> 1.0126 Offset: 0.08ppm Notes: Calibrated on 7550 software version 8.7.5 |
LI-7500A 1594 | 2017-11-28 | Lab none | Calibration | Lab none | Last calibration: 2017-09-27 AGC=99.9 SDM=0 DAC1=0 DAC2=0 Ambient P=100.5kPa, Tair=23.0C, Tdew=8.7C Zero Air at 300psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.8638 -> .8653 Offset: 6.09ppm CO2 span 1.0168 -> 1.0172 Offset: -.23ppm H2O zero 0.8878 -> .8868 Offset: -5.6ppm H2O span 1.0047 -> 1.0000 Offset: 0.1ppm Notes: Calibrated on 7550 software version 8.7.5 |
LI-7500A 75H-2668 | 2017-11-28 | Lab none | Calibration | Lab none | Last calibration: 2017-07-25 AGC=99.9 SDM=0 DAC1=0 DAC2=0 Ambient P=100.6kPa, Tair=21.7C, Tdew=8.9C Zero Air at 300psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.9395 -> .9392 Offset: -1.56ppm CO2 span 1.0159 -> 1.0147 Offset: 0.64ppm H2O zero 0.9047 -> 0.9066 Offset: 11.0ppm H2O span 1.0107 -> 1.0125 Offset: -0.04ppm Notes: Calibrated on 7550 software version 8.7.5 |
LI-7700 TG1-0224 | 2017-11-27 | Lab none | Calibration | Lab none | Calibration Location: Lab Last calibration: 2017-03-22 RSSI=71%, Optical RH=7% Ambient P=100.9kPa, Tair=22.4C Zero Air at 350psi, LBL Air 1.9497 at 1750 CH4 Zero Offset: 0.016ppm CH4 Span Offset: 0.005ppm Notes: |
LI-7500A 0418 | 2017-11-27 | Lab none | Calibration | Lab none | Last calibration: 2017-07-02
AGC=97.1 SDM=0 DAC1=0 DAC2=0 Ambient P=100.9kPa, Tair=22.8C, Tdew=6.8C Zero Air at 350psi, LBL Air 408.97 at 1750 Tdewspan=16.00 CO2 zero 0.8995 -> .8998 Offset: .87ppm CO2 span 1.0066 -> 1.006 Offset: .4ppm H2O zero 0.8436 -> .8437 Offset: .2ppm H2O span 0.9847 -> .9821 Offset: .05ppm Notes: Calibrated on 7550 software version 8.5.0 |
LI-7500A 0073 | 2017-11-27 | Lab none | Calibration | Lab none | Last calibration: 2017-06-22
AGC=100.6 SDM=0 DAC1=0 DAC2=0 Ambient P=101.0kPa, Tair=21.6C, Tdew=6.8C Zero Air at 350psi, LBL Air 408.97 at 1750 Tdewspan=16.00 CO2 zero 0.9572 -> .9571 Offset: -0.028ppm CO2 span 1.0084 -> 1.0118 Offset: -2.11ppm H2O zero 0.7307 -> .7306 Offset: -0.3ppm H2O span 1.0045 -> 1.0000 Offset: 0.1ppm Notes: Calibrated on 7550 software version 8.5.0 |
PTB110 H0160093 | 2017-11-16 | Lab none | Tested | Lab none | Consistently tested as 0V out. Vin was 12V and it was wired correctly. Looking at recent TW data (where this sensor was last) from Mar/Apr 2017, pressure data seems reasonable. Not sure where/why/when it got damaged. |
LI-7700 TG1-0357 | 2017-11-10 | Lab none | Installed | East End Gas Analyzer CH4 | Installed at EE to replace sn TG1-0155, which was brought back to the lab for calibration. |
LI-7700 TG1-0155 | 2017-11-10 | East End Gas Analyzer CH4 | Removed | Lab none | Removed from EE and brought back to the lab for calibration. It was replaced by sn TG1-0357. |
LI-7500A 75H-2176 | 2017-11-13 | Lab none | Installed | Bouldin Corn Gas Analyzer CO2/H2O | Installed on Bouldin Corn replacing sn 1594 for calibration. |
LI-7500A 1594 | 2017-11-13 | Bouldin Corn Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Bouldin Corn because it was calibrated on LI-7550 with embedded software v. 8.7.5 and the LI-7550 at the tower is embedded software v 8.5.0, so we were worried this software mismatch would produce wonky data. Replaced by sn 75H-2176. |
SD130BN 0030F4D21491 | 2017-07-12 | Lab none | Installed | Bouldin Corn Camera | This camera was put up on the Bouldin Corn Tower. It was attached to the southwest corner of the scaffolding just below the top railing and connected the to NL115 on the CR1000 to record the images on a CFM card. The CR1000 IP was set to 192.168.2.100 and the FTP user=Tower and pw=biomet. |
SD130BN 0030F4D214A4 | 2016-11-03 | Lab none | Installed | Bouldin Alfalfa Camera | This Stardot camera was mounted on the Bouldin Alfalfa tower. It is facing NW off the north side of the tower scaffolding. It is currently mounted near the top of the railing due to a short cable. It will be moved to near the top of the post later. |
LI-7500A 75H-2176 | 2017-11-09 | Lab none | Calibration | Lab none | Last calibration: 2017-06-12 AGC=89.6 SDM=0 DAC1=0 DAC2=0 Ambient P=100.6kPa, Tair=22.5C, Tdew=12.2C Zero Air at 400psi, LBL Air 408.97 at 1800 Tdewspan=16.00 CO2 zero 0.869 -> 0.869 Offset: 0ppm CO2 span 1.016 -> 1.0225 Offset: -3.37ppm H2O zero 0.8572 -> .8570 Offset: -1.3ppm H2O span 1.0022 -> 1.0047 Offset: -0.06ppm Notes: Calibrated on AIU-0674 (version 8.5.0) |
LI-7700 TG1-0224 | 2017-11-09 | Lab none | Repair | Lab none | Replaced thermocouple assembly on 7700. Temperature reading changed from 9999 to 24.5 deg C. |
LI-7700 TG1-0357 | 2017-11-09 | Lab none | Calibration | Lab none | Calibration Location: Lab Last calibration: 2017-03-23 RSSI=72.7%, Optical RH=23% Ambient P=100.6kPa, Tair=21.8C Zero Air at 400psi, LBL Air 1.9497 at 1800 CH4 Zero Offset: -0.1ppm CH4 Span Offset: -0.0197ppm Notes: Optics RH was 23%, so I flushed the head cavity with zero air and replaced the desiccant. We just changed the desiccant last time (March 2017) so I was expecting a RH ~5%. Let's keep an eye on the optics RH. |
SD130BN 0030F4D2148D | 2017-10-26 | Lab none | Installed | Twitchell Alfalfa Camera | Reinstalled at Twitchell Alfalfa. It is missing the plastic slide that clips into the enclosure - to be installed next time. |
SD130BN 0030F4D2148D | 2017-10-19 | Factory | Returned | Lab none | Returned from repair at StarDot |
SD130BN 0030F4D2148D | 2017-10-16 | Factory | Repair | Factory | Repaired by StarDot. Replaced CCD Board. |
SD130BN 0030F4D2148D | 2017-09-27 | Lab none | Shipped | Factory | Sent to StarDot for repair on RMA END-092217AAW1 |
SD130BN 0030F4D2148D | 2017-09-26 | Twitchell Alfalfa Camera | Removed | Lab none | Removed from Twitchell Alfalfa because it has been recording bad images with psychedelic colors. |
LI-7700 TG1-0223 | 2017-10-19 | Lab none | Installed | Bouldin Alfalfa Gas Analyzer CH4 | Installed at BA (now that the East End temporary towers are done, we have an available LI-7700 to put up here) |
eosFD FD20160001 | 2017-10-27 | Lab none | Repair | Factory | Sent back to Eosense for repair. |
eosFD FD20160001 | 2017-10-13 | Twitchell Alfalfa Forced Diffusion | Removed | Lab none | Removed from TA because it was giving bad readings (200? ppm +/- 200? ppm). |
LI-7700 TG1-0423 | 2017-10-26 | Lab none | Installed | Mayberry Gas Analyzer CH4 | Installed at MB to replace sn TG1-0357 |
LI-7700 TG1-0357 | 2017-10-26 | Mayberry Gas Analyzer CH4 | Removed | Lab none | Removed from MB and brought back to the lab for calibration |
PS-15M 12501584 | 2017-10-19 | East End-Temp1 Solar Controller | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
KD140SX-UFBS 11ZV0M0856 | 2017-10-19 | East End-Temp1 Solar Panel B | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
KD140SX-UFBS 11ZV0M0835 | 2017-10-19 | East End-Temp1 Solar Panel A | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
HMP45AC W0730036 | 2017-10-19 | East End-Temp1 RH & Temp | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
HFP01 10822 | 2017-10-19 | East End-Temp1 Heat Flux Plate B | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
CS450 20010770 | 2017-10-19 | East End-Temp1 Water Level | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
AM25T 2856 | 2017-10-19 | East End-Temp1 Multiplexer Thermocouples | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
CR6 2114 | 2017-10-19 | East End-Temp1 Data Logger Met | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
PAR-LITE 040412 | 2017-10-19 | East End-Temp1 PAR Sensor Reflected | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
NR-LITE 990352 | 2017-10-19 | East End-Temp1 Net Radiometer | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
WM 1590 153803 | 2017-10-19 | East End-Temp1 Anemometer | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
LI-7500A 0065 | 2017-10-19 | East End Gas Analyzer T1 CO2/H2O | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
LI-7700 TG1-0423 | 2017-10-19 | East End-Temp1 Gas Analyzer CH4 | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
LI-7550 AIU-0650 | 2017-10-19 | East End-Temp1 Interface LI-7550 | Removed | Lab none | Removed with decommissioning of East End Open Water tower (temporary tower 1) |
LI-7700 TG1-0421 | 2017-10-26 | Lab none | Installed | Sherman Wetland Gas Analyzer CH4 | Installed at SW to replace sn TG1-0224. |
LI-7700 TG1-0224 | 2017-10-26 | Sherman Wetland Gas Analyzer CH4 | Removed | Lab none | Removed from SW tower for calibration and because the temperature sensor was broken (reading 9999). Replaced by sn TG1-0421. |
LI-7500A 0042 | 2001-04-18 | Lab none | Installed | Tonzi-Floor | Installed on the Floor tower to replace existing IRGA for calibration. |
eosFD 20160003 | 2017-10-18 | Lab none | Installed | Tonzi | Installed near the Tree 1 soil CO2 profile for three month experiment |
GP-001 GP20140018 | 2017-08-16 | Lab none | Installed | West Pond CO2 Probe B | Installed at WP to replace a previous bad eosGP |
LI-7550 AIU-1569 | 2017-07-14 | Sherman Wetland Interface LI-7550 | Upgrade | Sherman Wetland Interface LI-7550 | Upgraded LI-7550 software from 8.0.0 to 8.7.5. |
eosFD 20160006 | 2017-10-18 | Lab none | Installed | Vaira Forced Diffusion East | Installed at Vaira near the East soil CO2 profile. The chamber is on a collar, fastened down with cable ties and power with 12V from a DC-DC power converter. |
eosFD 20160005 | 2017-10-18 | Lab none | Installed | Vaira Forced Diffusion West | Installed at Vaira near the West soil CO2 profile. The chamber is on a collar, fastened down with cable ties and power with 12V from a DC-DC power converter. |
LI-7500A 1594 | 2017-10-19 | Lab none | Installed | Bouldin Corn Gas Analyzer CO2/H2O | Installed on Bouldin Corn to replace sn 75H-2668 for calibration. |
LI-7700 TG1-0423 | 2017-10-23 | Lab none | Calibration | Lab none | Calibration Location: Lab Last calibration: 2017-05-31 RSSI=74.6%, Optical RH=3% Ambient P=101.3kPa, Tair=27.9C Zero Air at 425psi, AMP Air 1.9497 at 1850 CH4 Zero Offset: -0.055ppm CH4 Span Offset: 0.0197ppm Notes: |
LI-7500A 75H-2668 | 2017-10-19 | Bouldin Corn Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Bouldin Corn for calibration. Replaced by sn 1594 |
LI-7550 AIU-1570 | 2017-08-23 | Bouldin Corn Interface LI-7550 | Upgrade | Bouldin Corn Interface LI-7550 | Upgraded the 7550 firmware from v8.0.0 to v8.7.5. I confirmed the 7550 time, time zone, sonic analog settings, and CO2 and CH4 logging config. |
LAI-2050 PCA-1546 | 2017-10-18 | Borrowed | Returned | Lab none | Returned by Sebastian Wolf to the lab |
LAI-2000 PCA-1478 | 2017-10-18 | Borrowed | Returned | Lab none | Returned by Sebastian Wolf sewolf@ethz.ch to the lab |
LAI-2000 PCA-0846 | 2017-10-18 | Borrowed | Returned | Lab none | Returned by Sebastian Wolf sewolf@ethz.ch to the lab |
LI-7700 TG1-0424 | 2017-10-13 | Lab none | Installed | West Pond Gas Analyzer CH4 | Installed on West Pond to replace sn TG1-0223 for calibration |
LI-7700 TG1-0223 | 2017-10-13 | West Pond Gas Analyzer CH4 | Removed | Lab none | Removed from West Pond for calibration. Replaced by sn TG1-0424 |
LI-7500A 75H-2150 | 2017-10-13 | Lab none | Installed | West Pond Gas Analyzer CO2/H2O | Installed on West Pond to replace sn 0418 for calibration |
LI-7500A 0418 | 2017-10-13 | West Pond Gas Analyzer CO2/H2O | Removed | Lab none | Removed from West Pond for calibration. Replaced by sn 75H-2150 |
LI-7700 TG1-0421 | 2017-10-16 | Lab none | Calibration | Lab none | Calibration Location: Lab Last calibration: 2017-07-13 RSSI=77%, Optical RH=3% Ambient P=100.4kPa, Tair=29.7C Zero Air at 500psi, AMP Air 1.9497 at 2000 CH4 Zero Offset: 0.020ppm CH4 Span Offset: -0.0137ppm Notes: |
LI-7700 TG1-0223 | 2017-10-16 | Lab none | Calibration | Lab none | Calibration Location: Lab Last calibration: 2017-03-23 RSSI=79.2%, Optical RH=2% Ambient P=100.5kPa, Tair=28.7C Zero Air at 500psi, AMP Air 1.9497 at 2000 CH4 Zero Offset: -0.015ppm CH4 Span Offset: -0.0177ppm Notes: |
HFP01 000137 | 2017-10-13 | East End-Temp3 Heat Flux Plate B | Removed | Lost none | Cable removed from East End Boardwalk tower (temporary tower #3) with decommissioning of the site and brought back to the lab. The cable was broken and the sensor is lost to the wetland. |
HFP01SC 00245 | 2017-10-13 | East End-Temp3 Heat Flux Plate A | Removed | Lab none | Removed from East End Boardwalk tower (temporary tower #3) with decommissioning of the site. The cable has been chewed, and the cable-head connection needs to be repaired. |
HFP01 6649 | 2017-09-26 | East End Heat Flux Plate C | Lost | Disposed none | Cable removed from EE because the cable had been chewed through. The sensor is still lost to the wetland. |
PS-15M 12501588 | 2017-10-16 | East End-Temp2 Solar Controller | Removed | Lab none | Removed from the East End Temporary Tower #2 - Closed Canopy with the decommissioning of the site. |
LI-7500A 75H-2150 | 2017-10-11 | Lab none | Calibration | Lab none | Last calibration: 2017-05-30 AGC=96.8 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=23.9C, Tdew=5.5C Zero Air at 500psi, AMP CO2 408.97 at 2000 Tdewspan=16.0 CO2 zero 0.9024 -> 0.9018 Offset: -2.42ppm CO2 span 1.0194 -> 1.0070 Offset: 6.38ppm H2O zero 0.8664 -> 0.8649 Offset: -10.1ppm H2O span 0.9975 -> 1.0090 Offset: 0.24ppm Notes: |
LI-7700 TG1-0424 | 2017-10-11 | Lab none | Calibration | Lab none | Calibration Location: Lab Last calibration: 2017-03-22 RSSI=76.7%, Optical RH=4% Ambient P=100.5kPa, Tair=23.7C Zero Air at 500psi, AMP CH4 1.9497 at 2000 CH4 Zero Offset: 0.042ppm CH4 Span Offset: -0.001ppm Notes: |
LI-7500A 75H-2176 | 2017-09-26 | East End-Temp2 Gas Analyzer CO2/H2O | Removed | Lab none | Removed from East End Temporary Tower 2 - closed canopy with the decommissioning of the tower. |
U24-002-C 11020258 | 2017-09-26 | East End-Temp2 Conductivity Sensor | Removed | Lab none | Removed from East End Temporary Tower 2 - Closed Canopy with the decommissioning of the tower. |
miniDOT 1219 | 2017-09-26 | East End-Temp2 DO Sensor | Removed | Lab none | Removed from East End Temporary Tower 2 - Close Canopy with the decommissioning of the tower. |
CMTD-As4s0 3317052234210 | 2017-06-23 | Lab none | Installed | East End-Temp3 Solar Controller | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
NR-LITE 041556 | 2017-06-23 | Lab none | Installed | East End-Temp3 Net Radiometer | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
PAR-LITE 050596 | 2017-06-23 | Lab none | Installed | East End-Temp3 PAR Sensor Reflected | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
GP-001 GP20160012 | 2017-09-21 | Lab none | Installed | Vaira CO2 probe West Profile -0.16 | Installed in the Vaira West profile at -16cm |
GP-001 GP20160117 | 2017-09-21 | Lab none | Installed | Vaira CO2 probe West Profile -0.08 | Installed in the Vaira West profile at -8cm |
GP-001 GP20170028 | 2017-09-21 | Lab none | Installed | Vaira CO2 probe West Profile -0.04 | Installed in the Vaira West profile at -4cm |
GP-001 GP20160042 | 2017-09-21 | Lab none | Installed | Vaira CO2 probe West Profile -0.02 | Installed in the Vaira West profile at -2cm |
GMP222 K1350102 | 2017-09-21 | Vaira CO2 probe East Profile -0.04 | Moved | Vaira CO2 probe East Profile -0.16 | Moved from East profile at -4cm to East profile at -16cm |
GMP222 Y2009109 | 2017-09-21 | Vaira CO2 probe East Profile -0.02 | Moved | Vaira CO2 probe East Profile -0.08 | Moved from East profile at -2cm to East profile at -8cm |
GMP222 K1310103 | 2017-09-21 | Vaira CO2 probe North Profile -0.04 | Moved | Vaira CO2 probe East Profile -0.02 | Moved from North profile at -4cm to East profile at -2cm |
GMT222 G0230062 | 2017-09-21 | Vaira CO2 probe North Profile -0.02 | Moved | Vaira CO2 probe North Profile -0.16 | Moved from North profile at -2cm to North profile at -16cm |
GMP222 Y2030049 | 2017-09-21 | Vaira CO2 probe East Profile -0.08 | Moved | Vaira CO2 probe North Profile -0.02 | Moved from East profile at -8cm to North profile at -2cm |
GMP222 K1350101 | 2017-10-05 | Lab none | Installed | Vaira CO2 probe North Profile -0.04 | Installed at Vaira in the North profile at the -4cm location |
GMP222 K1310102 | 2017-10-05 | Lab none | Installed | Vaira CO2 probe East Profile -0.04 | Installed at Vaira in the East profile at the -4cm location |
CMTD-As4s0 3317052234210 | 2017-10-04 | East End-Temp3 Solar Controller | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
KD140SX-UFBS 11ZV0M0855 | 2017-10-04 | East End-Temp3 Solar Panel A | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
KD140SX-UFBS 11ZV0M0836 | 2017-10-04 | East End-Temp3 Solar Panel B | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
CR1000 23865 | 2017-10-04 | East End-Temp3 Data Logger Flux | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
NL115 6761 | 2017-10-04 | East End-Temp3 Interface NL115 | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
CR1000 50889 | 2017-10-04 | East End-Temp3 Data Logger Met | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
AM16-32A 1238 | 2017-10-04 | East End-Temp3 Multiplexer | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
HMP60 G4540021 | 2017-10-04 | East End-Temp3 RH & Temp | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
NR-LITE 041556 | 2017-10-04 | East End-Temp3 Net Radiometer | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
PAR-LITE 050596 | 2017-10-04 | East End-Temp3 PAR Sensor Reflected | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
CS451 20010046 | 2017-10-04 | East End-Temp3 Water Level | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
LI-7700 TG1-0424 | 2017-10-04 | East End-Temp3 Gas Analyzer CH4 | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
LI-7500 0976 | 2017-10-04 | East End-Temp3 Gas Analyzer CO2/H2O | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
WM 1590 120907 | 2017-10-04 | East End-Temp3 Anemometer | Removed | Lab none | Removed with the decommissioning of Temporary Tower #3 - Boardwalk site. |
LI-7550 AIU-1570 | 2017-10-04 | Bouldin Corn Interface LI-7550 | Removed | Lab none | Removed from the Bouldin Corn tower for testing in the lab for bad headers and 7700 drop outs during wash cycle. Replaced by sn AIU-0614. |