Equipment and Calibration

View History

Equipment Entry

Instrument:CO2 Probe
Manufacture:Vaisala
Model No:GMP343
Serial No:N2210009
Tag No:NULL
Acquisition Date:2017-06-06
Project:Kearney
Location on 2024-04-29:Lab, NULL
Notes: Calibrated over range 0-1000ppm; Certificate report no. H301-17220030

Documents

Calibration: cal_Vaisala_GMP343_N2210009.pdf
Calibration: cal_Vaisala_GMP343_20240220_AllFour.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
6695 2024-04-29 Lab none Tested Lab none I connected a DB9 connector white to pin 2, green to pin 3, and power to a lithium power pack, pink to +12V and gray to G and a spare brown wire between G and pin5 on the DB9. This allowed me to communicate with the probe via a serial terminal. I managed to get it to tell me its error notices and it confessed that its IR source has failed. Vaisala's web site has a automated repair quote process that had a flat repair fee for about 0 and a calibration fee of about 0. I'll send it for repair.
6686 2024-04-22 Tonzi CO2 Probe Tower Profile 6 Removed Lab none Not sending sensible data. Not sure why. Brought in for testing.
6582 2024-02-27 Lab none Installed Tonzi CO2 Probe Tower Profile 6 Reinstalled in the upper-mid location of the Tower co2 profile after testing in the lab.
6562 2024-02-09 Tonzi CO2 Probe Tower Profile 6 Removed Lab none Removed from the upper mid location on the Tower due to bad readings and to be calibrated
3295 2018-01-15 Lab none Installed Tonzi Installed at the 3rd lowest location in the tower co2 profile - not running yet due to data logger failure.