Rec No. | Date | Loc From | Title | Loc To | Notes |
6695 |
2024-04-29 |
Lab none |
Tested |
Lab none |
I connected a DB9 connector white to pin 2, green to pin 3, and power to a lithium power pack, pink to +12V and gray to G and a spare brown wire between G and pin5 on the DB9. This allowed me to communicate with the probe via a serial terminal. I managed to get it to tell me its error notices and it confessed that its IR source has failed. Vaisala's web site has a automated repair quote process that had a flat repair fee for about 0 and a calibration fee of about 0. I'll send it for repair. |
6686 |
2024-04-22 |
Tonzi CO2 Probe Tower Profile 6 |
Removed |
Lab none |
Not sending sensible data. Not sure why. Brought in for testing. |
6582 |
2024-02-27 |
Lab none |
Installed |
Tonzi CO2 Probe Tower Profile 6 |
Reinstalled in the upper-mid location of the Tower co2 profile after testing in the lab. |
6562 |
2024-02-09 |
Tonzi CO2 Probe Tower Profile 6 |
Removed |
Lab none |
Removed from the upper mid location on the Tower due to bad readings and to be calibrated |
3295 |
2018-01-15 |
Lab none |
Installed |
Tonzi |
Installed at the 3rd lowest location in the tower co2 profile - not running yet due to data logger failure. |