Equipment and Calibration History

View Equipment |

View history entry 3235 for Licor LI-7500A sn: 75H-2669

Date: 2017-12-15
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-08-14 AGC=103.3 SDM=0 DAC1=0 DAC2=0 Ambient P=100.7kPa, Tair=22.2C, Tdew=7C Zero Air at 250psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.9187 -> .9187 Offset: .39ppm CO2 span 1.0144 -> 1.0157 Offset: -.73ppm H2O zero 0.8988 -> .8988 Offset: .2ppm H2O span 1.0091 -> 1.0033 Offset: .12ppm Notes:
Entered by: