View history entry 3308 for Licor LI-7550 sn: AIU-1570

Date: 2018-01-29
From: Lab - none
To: Lab - none
Title: Upgrade
Notes: Downgraded from embedded Licor software v8.7.5 to 8.5.0. Version 8.7.5 was causing issues with header/data mismatch. I checked 7550 time and time zone and CO2 and CH4 logging config.
Entered by: Daphne Szutu