Equipment and Calibration History

View Equipment |

View history entry 3288 for Licor LI-7500A sn: 0035

Date: 2018-01-12
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-09-05 AGC=98.9 SDM=0 DAC1=0 DAC2=0 Ambient P=101.5kPa, Tair=20.5C, Tdew=10.2C Zero Air at 250psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.9373 -> .9378 Offset: 1.97ppm CO2 span 1.0084 -> 1.0288 Offset: -10.27ppm H2O zero 0.8976 -> .8978 Offset: 1.3ppm H2O span 1.0141 -> 1.0296 Offset: -.31ppm Notes:
Entered by: