Equipment and Calibration History

View Equipment |

View history entry 3313 for Campbell Sci. CR6 sn: 2115

Date: 2018-02-01
From: Lab - none
To: Lab - none
Title: Upgrade
Notes: Upgraded OS from CR6.Std.03 to CR6.Std.07.02. The older OS has a known issue where the CR6 is unresponsive until the internal lithium battery is removed and reinstalled. This new O.S. CR6.Std.07.02 requires updating your computer USB drivers using Device Config Utility 2.16 (or newer). I updated Device Config Utility on my computer from version 2.10 to 2.16. In the "Send OS" tab in Device Config Utility, the first step provides a link to update USB drivers. After upgrading the OS, I used the "Network Services" tab to enable FTP with the username tower and the usual password.
Entered by: Daphne Szutu