Equipment and Calibration History

View Equipment |

View history entry 451 for Licor LI-6262 sn: IRG3-1079

Date: 2011-11-18
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Calibrated in differential mode for check of Frank's new secondary CO2 standard. Ambient temperature was 21.5C, pressure was 100.45kPa. Using N2 and god bottle 402.69ppm CO2 (1850psi). Flowing N2 in both sample and reference CO2 = -0.30umol/mol adjusted to zero. H2O = -0.15 Flowing N2 in reference and god bottel in sample CO2 = 401.8ppm adjusted to 402.69ppm. Flowing N2 in reference and new secondary standard in sample, I wrote down 75 readings and made an average. The new CO2 secondary standard has 398.22ppm CO2.
Entered by: