Skip to main content
rice field banner biometlab logo
   
   
 
   
   
   
   
   
 
 

Equipment and Calibration

View History

Equipment Entry

Instrument:Gas Analyzer
Manufacture:Licor
Model No:LI-7500
Serial No:1653
Tag No:
Acquisition Date:0000-00-00
Project:Borrowed
Location on 2011-11-03:Returned,
Notes: Borrowed from Frank for an emergency replacement at Varia

Documents

Manual: 7500SoftwareUpdate.pdf
Manual: LI-7500_fieldnote_solar_sensitivity.pdf
Manual: LI-7500_letter_explaining_solar_sensitivity.pdf
Manual: LI-7500_Manual_V4_200410.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
444 2011-11-03 Lab none Calibration Lab none Last calibration: 2011-03-16
AGC=52 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.12kPa, Tair=18.45C, Tdew=7.99C
Zero Air at 700psi, CO2 in Air 410.49 at 1900 Tdewspan=13.0
CO2 zero 0.9237 -> 0.9232 Offset: -2.05ppm
CO2 span 1.0118 -> 1.0103 Offset: 0.86ppm
H2O zero 0.8762 -> 0.8756 Offset: -3.7ppm
H2O span 0.9831 -> 0.9407 Offset: 0.98ppm
Notes: Frank's briefly borrowed for use on Vaira.
 
 