Equipment and Calibration History

View Equipment |

View history entry 384 for Licor LI-6262 sn: IRG3-1079

Date: 2011-07-08
From: Lab - none
To: Lab - none
Title: Calibration
Notes: I calibrated the Li6262 in differential mode. I flowed zero air through a T into both the reference and sample cells at a flow just detectable on wetted lips. CO2 was low by about a 0.1 ppm CO2. I adjusted both CO2 zero and H2O zero. I then flowed 370.91ppm CO2 in air (god bottle) through the sample while leaving the zero air in the reference. The 6262 read 369.20+-0.05ppm. I adjusted the CO2 span to match the tank value. I tested the nearly empty 597.4ppm tank and it was true. I then test the new 400ppm CO2 in air tank and came up with a value of 405.85ppm and labeled the tank with this value.
Entered by: