View history entry 3540 for Licor LI-7500A sn: 1594

Date: 2018-06-25
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-04-04 AGC=99.9 SDM=0 DAC1=- DAC2=- Ambient P=100.5kPa, Tair=26.1C, Tdew=12.9C Zero Air at 350psi, AMP 408.97 at 1350 Tdewspan=15.99 CO2 zero 0.8651 -> 0.8648 Offset: -1.50ppm CO2 span 1.0185 -> 1.0181 Offset: 0.13ppm H2O zero 0.8854 -> 0.8869 Offset: 8.0ppm H2O span 1.0064 -> 1.0106 Offset: -0.09ppm Notes: Calibrated by Joe