Equipment and Calibration History

View Equipment |

View history entry 3541 for Licor LI-7500A sn: 0035

Date: 2018-06-25
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-01-12 AGC=99.2 SDM=0 DAC1=- DAC2=- Ambient P=100.5kPa, Tair=25.8C, Tdew=12.7C Zero Air at 350psi, AMP 408.97 at 1350 Tdewspan=15.99 CO2 zero 0.9378 -> 0.9374 Offset: -1.8ppm CO2 span 1.0288 -> 1.0296 Offset: 5.83ppm H2O zero 0.8978 -> 0.8979 Offset: -0.1ppm H2O span 1.0296 -> 1.0261 Offset: 0.08ppm Notes: Calibrated by Joe
Entered by: