View history entry 3539 for Licor LI-7500A sn: 75H-2180

Date: 2018-06-25
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-01-10 AGC=100.3 SDM=0 DAC1=- DAC2=- Ambient P=100.5kPa, Tair=26.3C, Tdew=13.2C Zero Air at 300psi, AMP 408.97 at 1350 Tdewspan=15.99 CO2 zero 0.9147 -> 0.9141 Offset: -3.0ppm CO2 span 1.0362 -> 1.0267 Offset: 4.93ppm H2O zero 0.8864 -> 0.8865 Offset: 0.5ppm H2O span 1.0168 -> 1.0274 Offset: -0.21ppm Notes: Calibrated by Joe