View history entry 3538 for Licor LI-7500A sn: 75H-2669

Date: 2018-06-25
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-12-15 AGC=104.2 SDM=0 DAC1=- DAC2=- Ambient P=100.4kPa, Tair=26.1C, Tdew=11.9C Zero Air at 350psi, AMP 408.97 at 1350 Tdewspan=15.99 CO2 zero 0.9187 -> 0.9182 Offset: -2.11ppm CO2 span 1.0157 -> 1.0125 Offset: 1.68ppm H2O zero 0.8988 -> 0.8994 Offset: 3.3ppm H2O span 1.0033 -> 1.0062 Offset: -0.06ppm Notes: Calibrated by Joe
Entered by: