LI-7500A 75H-2667 | 2021-09-29 | Lab | Calibration | Lab | Last calibration: 2021-06-07 AGC=101 SDM= DAC1= DAC2= Ambient P=100.5kPa, Tair=26.7C, Tdew=11.2C Zero Air at 1475psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00 CO2 zero 0.9422 -> .9411 Offset: -4.00ppm CO2 span 1.0122 -> 1.0174 Offset: -3.75ppm H2O zero 0.8404 -> .8413 Offset: 5.3ppm H2O span 1.0198 -> 1.0281 Offset: -.18ppm Notes: |
LI-7500A 0041 | 2021-09-23 | Lab none | Installed | Bouldin Corn Gas Analyzer CO2/H2O | Installed at Bouldin Corn to replace sn 75H-2150 for calibration. |
LI-7500A 75H-2150 | 2021-09-23 | Bouldin Corn Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Bouldin Corn for calibration. Replaced by sn 0041. |
LI-7500A 75H-2176 | 2021-09-23 | Lab none | Installed | Hill Slough Gas Analyzer CO2/H2O | Installed at Hill Slough to replace sn 75H-2667 for calibration. |
LI-7500A 75H-2667 | 2021-09-23 | Hill Slough Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Hill Slough for calibration. Replaced by sn 75H-2176. |
LI-7500A 0041 | 2021-09-21 | Lab | Calibration | Lab | Last calibration: 2021-05-25 AGC=98 SDM= DAC1= DAC2= Ambient P=100.3kPa, Tair=26.6C, Tdew=11.3C Zero Air at 1700psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00 CO2 zero 0.9966 -> .9667 Offset: .29ppm CO2 span 1.0137 -> 1.0145 Offset: -.44ppm H2O zero 0.8206 -> .8224 Offset: 10.9ppm H2O span 1.0266 -> 1.0228 Offset: .09ppm Notes: Calibrated zero and CO2 span on 9/20. There was an issue with the dew point generator where it was producing a dewpoint ~16C when the temp switch was set to 20.00C. I calibrated the H2O span the next day on 9/21. We fixed the problem by adding more water in the condenser block on the back of the machine. The water level looked ok while the cap was on and the machine was running, but once I turned the machine off and took the cap off, the water level dropped to below the "Min" line. After I added water and turned it back on, the dewpoint generator got to a reasonable dewpoint and it seemed stable. |
LI-7500A 75H-2176 | 2021-09-21 | Lab | Calibration | Lab | Last calibration: 2021-06-07 AGC=100 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=29C, Tdew=11.4C Zero Air at 1550psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00 CO2 zero 0.8673 -> .867 Offset: -1.56ppm CO2 span 1.0067 -> 1.0023 Offset: 3.07ppm H2O zero 0.8536 -> .8526 Offset: -5.5ppm H2O span 0.9148 -> .9091 Offset: .14ppm Notes: |
LI-7500A 0418 | 2021-09-21 | Lab | Calibration | Lab | Last calibration: 2021-09-20 AGC=99.3 SDM= DAC1= DAC2= Ambient P=100.7kPa, Tair=29C, Tdew=11.1C Zero Air at 1600psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00 CO2 zero .9004 -> .8992 Offset: -4.8ppm CO2 span 1.0016 -> .9972 Offset: 3.22ppm H2O zero .8444 -> .8490 Offset: 27.3ppm H2O span .9808 -> .9774 Offset: .08ppm Notes: I swapped the chemicals immediately before this calibration run. The chemicals were a tad sticky at the top but after a bit of poking with a screwdriver it was easy to dump the chemicals out into the waste jar. I don't think the chemicals needed to be changed necessarily. Before inserting the new chemical bottles, I ran some zero air into the headspace. This run was just to see what the numbers looked like. I will let it run on a bench for a week and calibrate again next week. |
LI-7500A 0418 | 2021-09-20 | Lab | Calibration | Lab | Last calibration: 2021-09-13 AGC=98 SDM= DAC1= DAC2= Ambient P=100.3kPa, Tair=26.6C, Tdew=11.3C Zero Air at 1700psi, at Tdewspan= CO2 zero 0.8992 -> Offset: ppm CO2 span 0.9956 -> Offset: ppm H2O zero 0.8484 -> Offset: 26ppm H2O span 0.9749 -> Offset: ppm Notes: Checked the zero and the H2O offset is still around 26 mmol/m3. I'm going to try changing the chemicals and calibrating again. |
LI-7500A 0418 | 2021-09-13 | Lab | Calibration | Lab | Last calibration: 2021-06-01 AGC=96.8 SDM= DAC1= DAC2= Ambient P=99.9kPa, Tair=26.4C, Tdew=14.6C Zero Air at 1725psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00 CO2 zero 0.9004 -> .8992 Offset: -4.69ppm CO2 span 1.0016 -> .9956 Offset: -4.2ppm H2O zero 0.8444 -> .8484 Offset: 23.0ppm H2O span 0.9808 -> .9749 Offset: .15ppm Notes: Large H2O zero offset. Check again next week. |
LI-7500A 0035 | 2021-09-13 | Lab | Calibration | Lab | Last calibration: 2021-06-01 AGC=100 SDM= DAC1= DAC2= Ambient P=100.0kPa, Tair=26.3C, Tdew=14.0C Zero Air at 1750psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00 CO2 zero 0.9375 -> .9373 Offset: -.82ppm CO2 span 1.0204 -> 1.0163 Offset: -2.87ppm H2O zero 0.8985 -> .8991 Offset: 3.6ppm H2O span 1.0268 -> 1.0291 Offset: -.05ppm Notes: |
LI-7500A 0065 | 2021-09-08 | Lab none | Calibration | Lab none | Last calibration: 2021-06-01
AGC=102 SDM= DAC1= DAC2= Ambient P=100.3kPa, Tair=27.75C, Tdew=15.5C Zero Air at 1875psi, AMP CA04119 524.77 at 1700 Tdewspan=20.00 CO2 zero 0.8913 -> .8911 Offset: -1.15ppm CO2 span 1.0201 -> 1.0211 Offset: .7ppm H2O zero 0.7305 -> .7313 Offset: 5.7ppm H2O span 1.0158 -> 1.0143 Offset: .03ppm Notes: Showing Carlos how to calibrate 7500s |
S2-411 1050 | 2021-09-15 | Cal Kit Narrow Band NDVI Reflected | Moved | Cal Kit Narrow Band NDVI Reflected | Moved from MB to the new Gilbert Tract site to continue the radiometer intercomparison there. |
S2-411 1049 | 2021-09-15 | Cal Kit Narrow Band NDVI Incoming | Moved | Cal Kit Narrow Band NDVI Incoming | Moved from MB to the new Gilbert Tract site to continue the radiometer intercomparison there. |
PAR-LITE 060831 | 2021-09-15 | Cal Kit PAR Sensor Reflected | Moved | Cal Kit PAR Sensor Incoming | Moved from MB to the new Gilbert Tract GT site to continue the radiometer intercomparison there. |
CNR4 121038 | 2021-09-15 | Mayberry | Moved | Gilbert Tract none | Moved from MB to the new Gilbert Tract site to continue the radiometer intercomparison there. |
CR1000 23865 | 2021-09-15 | Cal Kit Data Logger | Moved | Cal Kit Data Logger | Moved from MB to the new Gilbert Tract (GT) site to continue the radiometer intercomparison there. |
RPI0W RPI0W-003 | 2021-09-15 | Lab none | Installed | Mayberry none | Installed at MB on the wooden post next to the Stardot camera to continue the camera intercomparison. Joe has updated the firmware so the camera skips nighttime photos and takes a concurrent photo with fixed white balance (and maybe other improvements?) |
RPI0W RPI0W-003 | 2021-08-04 | Mayberry | Removed | Lab none | Removed from MB during a pause in the camera intercomparison. We plan to reinstall the camera after updating the firmware to skip nighttime photos. |
LI-7500A 0035 | 2021-09-15 | Lab none | Installed | Bouldin Alfalfa Gas Analyzer CO2/H2O | Installed at Bouldin Alfalfa to replace sn 75H-2176 for calibration. |
LI-7500A 75H-2176 | 2021-09-15 | Bouldin Alfalfa Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Bouldin Alfalfa for calibration. Replaced by sn 0035. |
HMP155 T2930114 | 2021-09-03 | Lab none | Installed | Bouldin Alfalfa none | Installed on the main BA tower (lower location) for the advection experiment |
HMP45AC U3030041 | 2021-09-03 | Bouldin Alfalfa | Removed | Lab none | Removed from BA bouncer tower, where it was the lower T/RH sensor. It was replaced by sn T2820183. |
HMP155 T2820180 | 2021-09-03 | Lab none | Installed | Bouldin Alfalfa none | Installed on the BA bouncer tower for the advection experiment in the upper T/RH location. |
HMP45AC T1250012 | 2021-09-03 | Bouldin Alfalfa | Removed | Lab none | Removed from BA bouncer tower, where it was the upper T/RH sensor. It was replaced by sn T2820180. |
HMP45AC U3030041 | 2021-08-04 | Lab none | Installed | Bouldin Alfalfa none | Installed on the BA bouncer tower next to the lower sonic for the advection experiment. It was wired to U11/U12. |
HMP45AC T1250012 | 2021-08-04 | Lab none | Installed | Bouldin Alfalfa none | Installed on the BA bouncer tower next to the upper sonic for the advection experiment. It was wired to U9/U10. |
eosFD FD20190027 | 2021-09-01 | Lab none | Installed | Tonzi-SoilCO2 Forced Diffusion Open | Installed at Tonzi-SoilCO2 to replace sn FD20180026 for testing it the lab and possible repair. |
eosFD FD20180026 | 2021-09-01 | Tonzi-SoilCO2 Forced Diffusion Open | Removed | Lab none | Removed from Tonzi-SoilCO2 due to consistent offset between soil and atmosphere chambers. This resulted in small fluxes when they should be pretty much zero. Replaced by sn FD20190027. |
LI-7500A 75H-2669 | 2021-09-01 | Lab none | Installed | Tonzi-Floor Gas Analyzer CO2/H2O | Installed at Tonzi-Floor to replace sn 0418 for calibration. |
LI-7500A 0418 | 2021-09-01 | Tonzi-Floor Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Floor for calibration. Replaced by sn 75H-2669. |
LI-7500A 75H-2514 | 2021-09-01 | Lab none | Installed | Tonzi-Tower Gas Analyzer CO2/H2O | Installed at Tonzi-Tower to replace sn 0035 for calibration. |
LI-7500A 0035 | 2021-09-01 | Tonzi-Tower Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Tonzi-Tower for calibration. Replaced by sn 75H-2514. |
LI-7500A 0042 | 2021-09-01 | Lab none | Installed | Vaira Gas Analyzer CO2/H2O | Installed at Vaira to replace sn 0065 for calibration. |
LI-7500A 0065 | 2021-09-01 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Vaira for calibration. Replaced by sn 0042. |
LI-7500A 0073 | 2021-08-31 | Lab | Calibration | Lab | Last calibration: 2021-06-01 AGC=99.0 SDM=0 DAC1=- DAC2=- Ambient P=99.7kPa, Tair=25.2C, Tdew=12.8C Zero Air at 1800psi, AMP CA04119 524.77 at 1700 Tdewspan=20.0 CO2 zero 0.9568 -> 0.9570 Offset: 0.3ppm CO2 span 1.0071 -> 1.0075 Offset: -0.07ppm H2O zero 0.7324 -> 0.7337 Offset: 10.1ppm H2O span 1.0037 -> 1.0078 Offset: -0.12ppm Notes: I loaded the config from the 03/15 calibration not the 06/01 config however this sensor was not used in the field between these calibrations because we are watching it due to noisy signals. Again the calibrations seem to be right on, but the noise around the set values is two the three times greater than our other IRGAs. I'm planning to send it in for retrofit. |
LI-7500A 0042 | 2021-08-31 | Lab | Calibration | Lab | Last calibration: 2021-05-25 AGC=99.8 SDM=0 DAC1=- DAC2=- Ambient P=99.7kPa, Tair=26.4C, Tdew=13.15C Zero Air at 1800psi, AMP CA04119 524.77 at 1700 Tdewspan=20.0 CO2 zero 0.9481 -> 0.9479 Offset: -0.78ppm CO2 span 1.0041 -> 1.0063 Offset: -1.62ppm H2O zero 0.8437 -> 0.8448 Offset: 7.1ppm H2O span 1.0149 -> 1.0163 Offset: -0.04ppm Notes: |
LI-7500A 75H-2669 | 2021-08-31 | Lab | Calibration | Lab | Last calibration: 2021-03-01 AGC=100.8 SDM=0 DAC1=- DAC2=- Ambient P=99.7kPa, Tair=26.2C, Tdew=13.6C Zero Air at 1800psi, AMP CA04119 524.77 at 1700 Tdewspan=20.0 CO2 zero 0.919 -> 0.9188 Offset: -0.85ppm CO2 span 1.0059 -> 1.0127 Offset: -4.77ppm H2O zero 0.8984 -> 0.9005 Offset: 13.3ppm H2O span 1.0033 -> 1.0086 Offset: -0.12ppm Notes: |
LI-7500A 75H-2514 | 2021-08-31 | Lab | Calibration | Lab | Last calibration: 2021-03-01 AGC=101.2 SDM=0 DAC1=- DAC2=- Ambient P=99.7kPa, Tair=26.2C, Tdew=13.1C Zero Air at 1800psi, AMP CA04119 524.77 at 1700 Tdewspan=20.0 CO2 zero 0.9048 -> 0.9046 Offset: -0.45ppm CO2 span 0.9984 -> 1.0082 Offset: -6.77ppm H2O zero 0.8389 -> 0.8399 Offset: 7.1ppm H2O span 1.0061 -> 1.0133 Offset: -0.16ppm Notes: |
SD130BN 0030F4D214A5 | 2016-06-06 | Lab none | Installed | Sherman Wetland Camera | Installed on 8ft 4x4 post on the southeast corner of the scaffolding |
LI-190/R Q100421 | 2021-08-25 | Lab none | Installed | Hill Slough PAR Sensor Diffuse | Installed at Hill Slough as part of the rotating shadow band for dfPAR. It replaced sn Q100420. |
LI-190/R Q100420 | 2021-08-25 | Hill Slough PAR Sensor Diffuse | Removed | Lab none | Removed from Hill Slough dfPAR and brought back to the lab for testing because it was mostly NAN or 0 with some sporadic data. It was replaced by sn Q100421. |
WMP 1352 000020 | 2021-08-30 | Borrowed none | Returned | Lab none | Returned by Sarah Daniel from LBL FLEXLAB. This is the final item returned. |
GP-001 GP20160070 | 2021-08-26 | Lab none | Notes | Lab none | Partly cleaned this sensor and tried to take the cable off. However, the connector at the top of the sensor twists loose from the body of the sensor so I don't think it's worth keeping. I'm not sure if the connector was already loose, or if I loosened it trying to table the cable off. |
S2-411 1050 | 2021-08-17 | Cal Kit Narrow Band NDVI Reflected | Moved | Cal Kit Narrow Band NDVI Reflected | Moved from EE to MB after completion of radiometer intercomparison at EE |
S2-411 1049 | 2021-08-17 | Cal Kit Narrow Band NDVI Incoming | Moved | Cal Kit Narrow Band NDVI Incoming | Moved from EE to MB after completion of radiometer intercomparison at EE |
PAR-LITE 060831 | 2021-08-17 | Cal Kit PAR Sensor Reflected | Moved | Cal Kit PAR Sensor Incoming | Moved from EE to MB after completion of radiometer intercomparison at EE |
CNR4 121038 | 2021-08-17 | East End | Moved | Mayberry none | Moved from EE to MB after completion of radiometer intercomparison at EE |
CR1000 23865 | 2021-08-17 | East End | Moved | Mayberry none | Moved from EE to MB after completion of radiometer intercomparison at EE |
HMP45AC U3030041 | 2021-08-02 | Lab none | Tested | Lab none | Ran for a few hours in the lab in the afternoon next to another HMP45, sn T1250012. There is an offset in RH and AirT, but they track ok. |
NOAA-CMDL CA03923 | 2021-08-16 | Factory none | Returned | Lab none | Returned from LBNL by Sebastien after refill and calibration. |
NOAA-CMDL CA03923 | 2021-07-22 | Factory none | Calibration | Factory none | Calibrated by Ameriflux at LBNL CO2: 488.75 +/- 0.02ppm, CH4: 3386.5 +/- 0.4ppb, H2O -6.24 +/- 1.34ppm. 2000psi. |
NOAA-CMDL CA04396 | 2021-08-16 | Factory none | Returned | Lab none | Returned from LBNL by Sebastien after refill and calibration. |
NOAA-CMDL CA04396 | 2021-08-09 | Factory none | Calibration | Factory none | Calibrated by Ameriflux at LBNL CO2: 417.61 +/- 0.02ppm, CH4: 1931.7 +/- 0.2ppb, H2O -0.63 +/- 0.86ppm |
NOAA-CMDL CA04396 | 2021-05-26 | Lab none | Shipped | Factory none | Sent to Ameriflux at LBL for refill. |
NOAA-CMDL CA03923 | 2021-05-26 | Lab none | Shipped | Factory none | Sent to Ameriflux at LBL for refill. |
PAR-LITE 050596 | 2005-07-11 | Factory none | Calibration | Factory none | Initial factory calibration 5.27 uV/(uE/m2/s) |
PAR-LITE 050597 | 2005-07-11 | Factory none | Calibration | Factory none | Initial factory calibration 5.35 uV/(uE/m2/s) |
GMP221 Y0850004 | 2021-08-11 | Tonzi-SoilCO2 CO2 Probe Tree2 Profile -0.16 | Removed | Lab none | Removed from -16cm in the soil co2 profile near LiuKang's tree. This profile has not been running for a while due to a break in the power line. This style of measurements are being replaced by forced diffusion chambers. |
GMT221 F4960020 | 2021-08-11 | Tonzi-SoilCO2 CO2 Probe Tree2 Profile -0.08 | Removed | Lab none | Removed from -8cm in the soil co2 profile near LiuKang's tree. This profile has not been running for a while due to a break in the power line. This style of measurements are being replaced by forced diffusion chambers. |
GMP222 K1350103 | 2021-08-11 | Tonzi-SoilCO2 CO2 Probe Tree2 Profile -0.04 | Removed | Lab none | Removed from -4cm in the soil co2 profile near LiuKang's tree. This profile has not been running for a while due to a break in the power line. This style of measurements are being replaced by forced diffusion chambers. |
GMP222 Z2720014 | 2021-08-11 | Tonzi-SoilCO2 CO2 Probe Tree2 Profile -0.02 | Removed | Lab none | Removed from -2cm in the soil co2 profile near LiuKang's tree. This profile has not been running for a while due to a break in the power line. This style of measurements are being replaced by forced diffusion chambers. |
CR10X 13214 | 2021-08-11 | Tonzi-SoilCO2 Data Logger Tree 2 | Removed | Lab none | Removed from the soil co2 profile near LiuKang's tree. This profile has not been running for a while due to a break in the power line. This style of measurements are being replaced by forced diffusion chambers. |
Mark I A000251 | 2021-08-11 | Tonzi Weather Station | Removed | Lab none | Removed from the Tower top due to not sending data and one of the incoming sensors was removed by woodpeckers. |
PQS 1 110340 | 2021-08-11 | Tonzi PAR Sensor Under-A | Removed | Lab none | Removed from Tonzi near z-cam 2 because the data seems not so useful. |
eosFD FD20180026 | 2021-08-11 | Tonzi-SoilCO2 Forced Diffusion Open | Maintenance | Tonzi-SoilCO2 Forced Diffusion Open | Performance check done. This was a double check of the last check since the CO2 concentrations were so far from ambient. The two chambers were still offset: Average Flux: 1.3, 0.37 STDEV Average ATM: 343.7ppm, 41.5C Average Soil: 408.5ppm, 41.5C Need to ask Eosense again. |
BB-Fab BB-TT | 2021-07-29 | Tonzi-Tower Battery Bank | Maintenance | Tonzi-Tower Battery Bank | I cleaned the battery terminals, replaced the affected battery cables (positive terminals) and sealed all the connections. |
WM 1590 152607 | 2021-08-06 | Lab none | Returned | Returned none | Returned to Tibisay. |
eosFD FD20180048 | 2021-07-29 | Tonzi-SoilCO2 Forced Diffusion Tree 1 | Maintenance | Tonzi-SoilCO2 Forced Diffusion Tree 1 | I unfastened this chamber and laid it on its side for a performance check. Average Flux: 0.0, 0.04 STDEV Average ATM: 242.5ppm, 51.9C Average Soil: 243.0ppm, 51.7C I reset the baseline CO2 concentration to 400ppm. This chamber does not have a sun shield and even though it was in the shade under the trees its temperatures were very high. |
eosFD FD20190027 | 2021-08-04 | Lab none | Maintenance | Lab none | This chamber was removed from the Mayberry boom because its readings looked bad. In the lab I cleaned it up. There are what look like tooth or claw marks on most of the filter windows but no breaks. The initial performance check had the first of the five measurement way off which skewed everything, but its concentrations were about -140ppm. I reset the concentrations to 400ppm and did a second performance check. The second check was much better: Average Flux: -0.1, 0.05 STDEV Average ATM: 378.1ppm, 32.0C Average Soil: 374.8ppm, 32.0C I reset the baseline CO2 concentration to 400ppm again. I think this chamber can be swapped back into the field as needed. |
eosFD FD20180047 | 2021-07-21 | Lab none | Installed | Mayberry Forced Diffusion Boom | Installed at Mayberry to replace sn FD20190027 because 27's numbers looked bad. |
eosFD FD20190027 | 2021-07-21 | Mayberry Forced Diffusion Boom | Removed | Lab none | Removed from Mayberry because the numbers looked bad. Replaced by sn FD20180047. |
ML2x 54/007 | 2018-06-13 | Twitchell Alfalfa Soil Moisture Probe SoilCO2 A | Removed | Lab none | Removed with Twitchell Alfalfa site decommissioning. |
ML2x 54/001 | 2012-08-02 | Lab none | Installed | Sherman Island Soil Moisture Probe SoilCO2 D | Installed at Sherman Island next to the East soil CO2 profile. This sensor was connected to the fourth position on the data logger. |
ML2x 139/052 | 2012-08-02 | Lab none | Installed | Sherman Island Soil Moisture Probe SoilCO2 C | Installed at Sherman Island next to the East soil CO2 profile. This sensor was connected to the third position on the data logger. |
ML2x 54/007 | 2014-08-07 | Sherman Island Soil Moisture Probe SoilCO2 B | Removed | Lab none | Removed from Sherman Island with the rest of the soil co2 system for use at Twitchell Alfalfa. |
ML2x 54/007 | 2012-07-19 | Lab none | Installed | Sherman Island Soil Moisture Probe SoilCO2 B | Installed next to the west soil co2 profile at -2cm |
ML2x 64/072 | 2004-10-22 | Lab none | Installed | Vaira Soil Moisture Probe E | Installed at Vaira to replace sn 54/007 that was reading low. |
ML2x 54/007 | 2004-10-22 | Vaira Soil Moisture Probe E | Removed | Lab none | Removed from Vaira due to low readings. Replaced by sn 64/072 |
ML2x 54/007 | 2002-08-01 | Vaira Soil Moisture Probe E | Location | Vaira Soil Moisture Probe E | From a field note by Liu Kang, this sensor is in the fifth location at Vaira. It is noted that sn 54/020 was listed in this location previously but it is unknown when the swap was made. |
S2-411 1050 | 2021-07-21 | Cal Kit Narrow Band NDVI Reflected | Moved | Cal Kit Narrow Band NDVI Reflected | Moved from West Pond to East End at the conclusion of the radiometer intercomparison at WP |
S2-411 1049 | 2021-07-21 | Cal Kit Narrow Band NDVI Incoming | Moved | Cal Kit Narrow Band NDVI Incoming | Moved from West Pond to East End at the conclusion of the radiometer intercomparison at WP |
PAR-LITE 060831 | 2021-07-21 | Cal Kit PAR Sensor Reflected | Moved | Cal Kit PAR Sensor Incoming | Moved from WP to EE at the conclusion of the radiometer inter-comparison at WP |
CNR4 121038 | 2021-07-21 | West Pond | Moved | East End none | Moved from West Pond to East End at the conclusion of the radiometer intercomparison at WP |
CR1000 23865 | 2021-07-21 | West Pond | Moved | East End none | Moved from West Pond to East End at the conclusion of the radiometer intercomparison at WP |
WM 1590 153805 | 2021-07-20 | Lab none | Installed | Bouldin Alfalfa-ADEX Anemometer Bouncer Lower | This sonic was installed as part of the upwind tower for the advection experiment. It was installed at 242cm agl on a cross arm on the tripod tower 283m northwest of the main tower. The sonic heading is 19deg and it is about 0.5m northwest of the tripod center post. |
WM 1590 153806 | 2021-07-20 | Bouldin Alfalfa Anemometer | Removed | Lab none | Removed from Bouldin Alfalfa and replaced by sn 074417 so that the advection experiment could run with four inter-compared sonics. |
LI-190/R Q100420 | 2015-05-08 | Factory none | Calibration | Factory none | Initial factory calibration: 7.06 uA/(1000 mmol/m2/s). If using with 604 Ohm mV adapter: -234.54 (umol/m2/s)/mV. |
PS-15 03220377 | 2021-07-20 | Lab none | Installed | Bouldin Alfalfa-ADEX Solar Controller Bouncer | Installed as part of the upwind tower for the advection experiment. It was installed at the tripod tower 283m northwest of the main tower to control one solar panel and one battery. |
CR6 2115 | 2021-07-20 | Lab none | Installed | Bouldin Alfalfa-ADEX Data Logger Bouncer Flux | Installed as part of the upwind tower for the advection experiment. It was installed at the tripod tower 283m northwest of the main tower to collect data from two sonics and fine wire TC. |
WM 1590 152603 | 2021-07-20 | Lab none | Installed | Bouldin Alfalfa-ADEX Anemometer Bouncer Upper | This sonic was installed as part of the upwind tower for the advection experiment. It was installed at 393cm agl on the top of a tripod tower 283m northwest of the main tower. Its heading is 19deg |
WM 1590 152601 | 2021-07-20 | Lab none | Installed | Bouldin Alfalfa-ADEX Anemometer Main Lower | Installed on a cross bar mounted on the NW corner of the scaffolding at 242cm agl for the advection experiment. The data from this sonic is input and digitized by the 7700. Its heading is 253deg. |
WM 1590 074417 | 2021-07-20 | Lab none | Installed | Bouldin Alfalfa Anemometer | Installed at Bouldin Alfalfa to replace sn 152806 for advection experiemnt. This sonic was compare to the other three used for the advection experiement. |
eosFD FD20160001 | 2021-07-20 | Bouldin Alfalfa Forced Diffusion A | Maintenance | Bouldin Alfalfa Forced Diffusion A | Performance check done. Referenced reset to 400ppm. Average Flux: -0.1, 0.07 STDEV Average ATM: 402.1ppm, 48.5C Average Soil: 398.1ppm, 48.2C |
eosFD FD20190028 | 2021-07-20 | Hill Slough Forced Diffusion A | Maintenance | Hill Slough Forced Diffusion A | Performance check done. Referenced reset to 400ppm. Average Flux: -0.0, 0.03 STDEV Average ATM: 172.2ppm, 34.6C Average Soil: 171.6ppm, 34.5C |
eosFD FD20180067 | 2021-07-20 | Bouldin Corn Forced Diffusion C | Maintenance | Bouldin Corn Forced Diffusion C | Performance check done. Referenced reset to 400ppm. Average Flux: -0.4, 0.09 STDEV Average ATM: 551.1ppm, 39.5C Average Soil: 534.8ppm, 39.5C Large difference between internal chambers. |
WM 1590 152607 | 2021-07-16 | Lab none | Damaged | Lab none | Today I noticed that this sonic is missing pin 8 in its round connector that is used for the first analog output (usually U). |
eosFD FD20190027 | 2020-08-11 | Mayberry | Removed | Lab none | Removed from MB for a thorough cleaning |
MNKID-B KB04456 | 2021-07-15 | Lab none | Upgrade | Lab none | The screen on this unit seemed to be black but at an extreme viewing angle the text could be seen on top of solid dark LCD boxes. It seems to be intermittent and not controllable with the LCD back light and contrast controls. I updated the firmware to v1856. v1864 is the newest firmware, but I could not get it to load that version. I also set it up to spit out serial data at 57600 baud but still haven't figured out how to decode it. |
eosFD FD20180068 | 2021-07-13 | Vaira Forced Diffusion West | Maintenance | Vaira Forced Diffusion West | Performance check done. Referenced to 400ppm. Average Flux: -0.0, 0.08 STDEV Average ATM: 610.7ppm, 49.6C Average Soil: 609.8ppm, 49.8C |