Equipment and Calibration History

View Equipment |

View history entry 5478 for Texas Electronics TR-525I-R1 sn: 24872-999

Date: 2021-10-06
From: Tonzi -
To: Tonzi - none
Title: Calibration
Notes: Tested with our new method using the peristaltic pump and 237ml of water which should result in 50 tips. 1/2 turn of the screws (CW+, CCW-) should adjust the number of tips by 2-3% (1-1.5 tips). Got 54 tips – I turned the screws two turns CCW
Entered by: Joe Verfaillie
Field notes