View history entry 5477 for Texas Electronics TE525 sn: 48102-611

Date: 2021-10-06
From: Tonzi -
To: Tonzi - none
Title: Calibration
Notes: Tested with our new method using the peristaltic pump and 237ml of water which should result in 50 tips. 1/2 turn of the screws (CW+, CCW-) should adjust the number of tips by 2-3% (1-1.5 tips). Got 50 tips – No adjustment
Entered by: Joe Verfaillie