View history entry 2550 for Licor LI-7500A sn: 75H-2182

Date: 2016-12-19
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2016-08-30 AGC=87.2 SDM=6 DAC1=- DAC2=- Ambient P=101.31kPa, Tair=20.29C, Tdew=-0.86C Zero Air at 300psi, God CO2 421.22 at 400 Tdewspan=16.00 CO2 zero 0.8831 -> 0.8837 Offset: 2.75ppm CO2 span 1.0142 -> 1.0207 Offset: -3.48ppm H2O zero 0.8464 -> 0.8457 Offset: -4.2ppm H2O span 1.023 -> 1.0186 Offset: 0.09ppm Notes: