Equipment and Calibration History

View Equipment |

View history entry 2576 for Gill WM 1590 sn: 153805

Date: 2017-01-10
From: Lab - none
To: Lab - none
Title: Upgrade
Notes: Upgraded firmware to correct for W bug. Old SN: Y153805 New SN: X153805 Upgrade file: Y153802-Y153806-m5.gcu Old firmware: 2329-601-03 New firmware: 2329-701-01 All Checksum fields report PASS. Log file: WMUpgradeLog_Y153805_170110_093737.txt Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting. --------------------------------- THE FOLLOWING FIELDS HAVE CHANGED Baud Rate : 19200 Baud Rate : 57600 Output rate : P1 1 Hz Output rate : P8 20 Hz THE NEW CONFIGURATION IN FULL ==== DETAILS OF DETECTED ANEMOMETER======================== Serial Number : X153805 ---- SW VERSION-------------------------------------------- Anemometer Firmware Version: 2329-701 ---- HW DETAILS OF ANEMOMETER------------------------------ Detected Anemometer Class : WindMaster 20Mhz Number of AXIS on this unit : 3 DAC resolution of attached unit : 14 Number of DAC channels : 4 Maximum sampling rate : (8) P8 20 Hz ---- USER CONFIGURATION OF ANEMOMETER---------------------- Baud Rate : 19200 Baud Rate : 57600 Output Message Format : M1 UV(W) Output Message Velocity Units : m/s Output Message Terminator : Retries : ON Instantaneous Sampling : OFF Factory Calibration : ON Physical Communications : Auto Resolution settings : High resolution Alignment settings : North to spar Averaging settings : Averaging inactive Output rate : P1 1 Hz Output rate : P8 20 Hz Polar angle minimum magnitude: 0.050 m/s Power-On Message : Display Power On Message ASCII Format : Comma Separated (CSV) SOS and Sonic Temperature display : SOS Analogue output mode : -5V to 5V Unit Identifier : 'Q' Output Channel 1 : U output (full range = 30m/s) Output Channel 2 : V output (full range = 30m/s) Output Channel 3 : W output (full range = 5m/s) Output Channel 4 : Speed of Sound Synchronised Polling : Off
Entered by: Daphne Szutu