Equipment and Calibration History

View Equipment |

View history entry 3381 for Licor LI-7700 sn: TG1-0224

Date: 2018-04-04
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Calibration Location: Lab Last calibration: 2017-11-27 RSSI=73%, Optical RH=14% Ambient P=100.7kPa, Tair=23.9C Zero Air at 100psi, LBL Air 1.9497 at 1700 CH4 Zero Offset: -.002ppm CH4 Span Offset: .167ppm Notes: Powered up twice successfully and synced waveforms within 20 seconds. Actual laser temperature = 25.6C, set laser temperature = 26.C.
Entered by: