View history entry 3346 for Licor LI-7500A sn: 75H-2182

Date: 2018-03-05
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2018-02-20 AGC=86.9 SDM=0 DAC1=0 DAC2=0 Ambient P=101.4kPa, Tair=25.1C, Tdew=1.2C Zero Air at 150psi, LBL Air 408.97 at 1700 Tdewspan=16.00 CO2 zero 0.8844 -> .8844 Offset: -0.07ppm CO2 span 1.0215 -> 1.0217 Offset: -.11ppm H2O zero 0.8446 -> .8453 Offset: 3.8ppm H2O span 1.0038 -> 1.0169 Offset: -.26ppm Notes: H2O span set with our dewpoint generator, which agrees with Patty's brand-new dewpoint generator