Equipment and Calibration History

View Equipment |

View history entry 3195 for Licor LI-7500A sn: 1594

Date: 2017-11-13
From: Bouldin Corn - Gas Analyzer CO2/H2O
To: Lab - none
Title: Removed
Notes: Removed from Bouldin Corn because it was calibrated on LI-7550 with embedded software v. 8.7.5 and the LI-7550 at the tower is embedded software v 8.5.0, so we were worried this software mismatch would produce wonky data. Replaced by sn 75H-2176.
Entered by: Joe Verfaillie