Ser No.DateLoc FromTitleLoc ToNotes
WM 1590 153805 2017-01-17 Lab none Installed Twitchell Rice Anemometer Installed on Twitchell Rice tower to replace sn 120906. We also replaced the sonic cable.
WM 1590 120906 2017-01-19 Twitchell Rice Anemometer Removed Lab none Removed from Twitchell Rice tower and returned to lab for firmware upgrade. It was replaced by sn 153805.
WM 1590 124215 2017-01-17 Lab none Installed East End Anemometer Installed on East End tower to replace sn 153804.
WM 1590 153804 2017-01-17 East End Anemometer Removed Lab none Returned to lab for firmware upgrade. It was replaced by sn 124215.
WM 1590 120908 2017-01-17 Lab none Installed West Pond Anemometer Installed on West Pond tower to replace sn 120907.
WM 1590 120907 2017-01-17 West Pond Anemometer Removed Lab none Removed from West Pond and returned to lab for firmware upgrade. It was replaced by sn 120908.
LI-7550 AIU-0674 2017-01-05 Lab none Shipped Factory Sent to Licor on RMA 00199505 with 75H-2668. This box had its door left open in the field and got rain inside. It was functional but throwing errors - probably a bad temperature controller. It was sent with the head to Licor for repair.
LI-7500A 75H-2668 2017-01-05 Lab none Shipped Factory Sent to Licor on RMA 00199505 with AIU-0674. This head was on the box when its door was left open in the field and got rain inside. Subsequent lab calibration looked a little funny so it was sent with the box to Licor for a check.
LI-7500A 0418 2017-01-12 Lab none Calibration Lab none Last calibration: 2016-08-15
AGC=49 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.44kPa, Tair=22.75C, Tdew=6.33C
Zero Air at 1800psi, God CO2 421.22 at 400 Tdewspan=16.00
CO2 zero 0.9012 -> 0.9009 Offset: ppm
CO2 span 1.007 -> 1.0062 Offset: ppm
H2O zero 0.8558 -> 0.8570 Offset: 0.17ppm
H2O span 0.9911 -> 0.9788 Offset: 0.02ppm
Notes: Using a new tank of Zero Air. Forgot to write offsets for CO2. The serial port on bottom of Licor box has missing pin. Connect to laptop using a different cable that has one serial end and one green-guillotine-port end. Connect the green-guillotine-port end inside the box for the "Serial" connection.
WM 1590 153805 2017-01-10 Lab none Upgrade Lab none Upgraded firmware to correct for W bug.

Old SN: Y153805
New SN: X153805
Upgrade file: Y153802-Y153806-m5.gcu
Old firmware: 2329-601-03
New firmware: 2329-701-01

All Checksum fields report PASS.

Log file: WMUpgradeLog_Y153805_170110_093737.txt

Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting.

---------------------------------
THE FOLLOWING FIELDS HAVE CHANGED

Baud Rate : 19200
Baud Rate : 57600
Output rate : P1 1 Hz
Output rate : P8 20 Hz

THE NEW CONFIGURATION IN FULL

==== DETAILS OF DETECTED ANEMOMETER========================

Serial Number : X153805

---- SW VERSION--------------------------------------------
Anemometer Firmware Version: 2329-701

---- HW DETAILS OF ANEMOMETER------------------------------
Detected Anemometer Class : WindMaster 20Mhz
Number of AXIS on this unit : 3
DAC resolution of attached unit : 14
Number of DAC channels : 4
Maximum sampling rate : (8) P8 20 Hz

---- USER CONFIGURATION OF ANEMOMETER----------------------
Baud Rate : 19200
Baud Rate : 57600
Output Message Format : M1 UV(W)
Output Message Velocity Units : m/s
Output Message Terminator :
Retries : ON
Instantaneous Sampling : OFF
Factory Calibration : ON
Physical Communications : Auto
Resolution settings : High resolution
Alignment settings : North to spar
Averaging settings : Averaging inactive
Output rate : P1 1 Hz
Output rate : P8 20 Hz
Polar angle minimum magnitude: 0.050 m/s
Power-On Message : Display Power On Message
ASCII Format : Comma Separated (CSV)
SOS and Sonic Temperature display : SOS
Analogue output mode : -5V to 5V
Unit Identifier : 'Q'
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 3 : W output (full range = 5m/s)
Output Channel 4 : Speed of Sound
Synchronised Polling : Off

NULL 0030F4D0A167 2011-10-18 Lab none Installed Tonzi Installed on the Tonzi tower to cover the same area as the existing webcam. The camera is mounted about 20m agl and is pointing about 310deg true and 20deg down from horizontal. Horizontal FOV is about 70deg.
NULL 0030F4D0A168 2011-10-13 Lab none Installed Vaira Installed on the Vaira radiation tower to cover the same area as the existing webcam. The camera is about 3m agl, pointed at about 290deg true and pointed down 10deg from horizontal. FOV is about 70deg.
WM 1590 124215 2017-01-06 Lab none Upgrade Lab none Upgraded firmware to correct for W bug.

Old SN: Y124215
New SN: X124215
Upgrade file: Y074417-Y124215-m.gcu, Y012345-s.gcu, X120908 X124215-m.gcu
Old firmware: 2329-114
New firmware: 2329-116-01

All Checksum fields report PASS.

There are multiple upgrade .gcu files. During the initial upgrade using file "Y074417-Y124215-m.gcu," the sensor was set at 57,600 baud rate. This baud rate caused an error in the upgrade, so the SN reset to Y012345. Gill sent me the second upgrade file, "Y-12345-s.gcu," which reset the sensor to SN X124215. Finally, I used the third upgrade file, "X120908 X124215-m.gcu," to upgrade the firmware.

Log file: WMUpgradeLog_X124215_170106_123318.txt

Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting.

-------

THE FOLLOWING FIELDS HAVE CHANGED

Baud Rate : 19200
Baud Rate : 57600
Output rate : P1 1 Hz
Output rate : P8 20 Hz
Output Channel 3 : W output (full range = 30m/s)
Output Channel 3 : W output (full range = 5m/s)

THE NEW CONFIGURATION IN FULL

==== DETAILS OF DETECTED ANEMOMETER========================

Serial Number : X124215

---- SW VERSION --------------------------------------------
Anemometer Firmware Version: 2329-116

---- HW DETAILS OF ANEMOMETER------------------------------
Detected Anemometer Class : WindMaster 20Mhz
Number of AXIS on this unit : 3
DAC resolution of attached unit : 14
Number of DAC channels : 4
Maximum sampling rate : (8) P8 20 Hz

---- USER CONFIGURATION OF ANEMOMETER----------------------
Baud Rate : 19200
Baud Rate : 57600
Output Message Format : M1 UV(W)
Output Message Velocity Units : m/s
Output Message Terminator :
Retries : ON
Instantaneous Sampling : OFF
Factory Calibration : ON
Physical Communications : Auto
Resolution settings : High resolution
Alignment settings : North to spar
Averaging settings : Averaging inactive
Output rate : P1 1 Hz
Output rate : P8 20 Hz
Polar angle minimum magnitude: 0.050 m/s
Power-On Message : Display Power On Message
ASCII Format : Comma Separated (CSV)
SOS and Sonic Temperature display : SOS
Analogue output mode : -5V to 5V
Unit Identifier : 'Q'
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 3 : W output (full range = 30m/s)
Output Channel 3 : W output (full range = 5m/s)
Output Channel 4 : Speed of Sound

WM 1590 120908 2017-01-06 Lab none Upgrade Lab none Upgraded firmware to correct for W bug.

Old SN: Y120908
New SN: X120908
Upgrade file: Y074417-Y124215-m.gcu, Y012345-s.gcu, X120908 X124215-m.gcu
Old firmware: 2329-114
New firmware: 2329-116-01

All Checksum fields report PASS.

There are multiple upgrade .gcu files. During the initial upgrade using file "Y074417-Y124215-m.gcu," the sensor was set at 57,600 baud rate. This baud rate caused an error in the upgrade, so the SN reset to Y012345. Gill sent me the second upgrade file, "Y-12345-s.gcu," which reset the sensor to SN X120908. Finally, I used the third upgrade file, "X120908 X124215-m.gcu," to upgrade the firmware.

Log file: WMUpgradeLog_X120908_170106_122011.txt

Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting.

-------

THE FOLLOWING FIELDS HAVE CHANGED

Baud Rate : 19200
Baud Rate : 57600
Output rate : P1 1 Hz
Output rate : P8 20 Hz

THE NEW CONFIGURATION IN FULL

==== DETAILS OF DETECTED ANEMOMETER========================

Serial Number : X120908

---- SW VERSION--------------------------------------------
Anemometer Firmware Version: 2329-116

---- HW DETAILS OF ANEMOMETER------------------------------
Detected Anemometer Class : WindMaster 20Mhz
Number of AXIS on this unit : 3
DAC resolution of attached unit : 14
Number of DAC channels : 4
Maximum sampling rate : (8) P8 20 Hz

---- USER CONFIGURATION OF ANEMOMETER----------------------
Baud Rate : 19200
Baud Rate : 57600
Output Message Format : M1 UV(W)
Output Message Velocity Units : m/s
Output Message Terminator :
Retries : ON
Instantaneous Sampling : OFF
Factory Calibration : ON
Physical Communications : Auto
Resolution settings : High resolution
Alignment settings : North to spar
Averaging settings : Averaging inactive
Output rate : P1 1 Hz
Output rate : P8 20 Hz
Polar angle minimum magnitude: 0.050 m/s
Power-On Message : Display Power On Message
ASCII Format : Comma Separated (CSV)
SOS and Sonic Temperature display : SOS
Analogue output mode : -5V to 5V
Unit Identifier : 'Q'
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 3 : W output (full range = 5m/s)
Output Channel 4 : Speed of Sound
WM 1590 153802 2017-01-05 Lab none Installed Twitchell Alfalfa Anemometer Installed at Twitchell Alfalfa to replace sn 153805. Brand new.
WM 1590 153805 2017-01-05 Twitchell Alfalfa Anemometer Removed Lab none Removed from Twitchell Alfalfa tower for firmware upgrade. Replaced by sn 153802.
LI-7500A 75H-2514 2016-12-29 Lab none Installed Twitchell Alfalfa Gas Analyzer CO2/H2O Installed at Twitchell Alfalfa with new 7550 box sn AIU-1557 to replace damaged system
LI-7550 AIU-1557 2016-12-29 Lab none Installed Twitchell Alfalfa Interface LI-7550 Installed at Twitchell Alfalfa after old AIU was damaged in rain. Installed together with 7500 head sn 75H-2514
LI-7500A 75H-2180 2016-12-29 Lab none Installed Sherman Wetland Gas Analyzer CO2/H2O Installed at Sherman Wetland after calibration to replace sn 1594
LI-7500A 1594 2016-12-29 Sherman Wetland Gas Analyzer CO2/H2O Removed Lab none Removed from Sherman Wetland for calibration. Replaced by sn 75H-2180
WM 1590 153802 2017-01-03 Lab none Upgrade Lab none Upgraded firmware to correct for W bug.

Old SN: Y153802
New SN: X153802
Upgrade file: Y153802-Y153806-m5.gcu
Old firmware: 2329-601
New firmware: 2329-701-01

All Checksum fields report PASS.

Log file: WMUpgradeLog_Y153802_170103_110126.txt

Config file below. If there are two entries in a field, the first entry is the old setting and the second entry is the new setting.

-------

THE FOLLOWING FIELDS HAVE CHANGED

Baud Rate : 19200
Baud Rate : 57600
Output Message Format : M2 Polar
Output Message Format : M1 UV(W)
Resolution settings : Normal resolution
Resolution settings : High resolution
Output rate : P1 1 Hz
Output rate : P8 20 Hz
SOS and Sonic Temperature display : Neither
SOS and Sonic Temperature display : SOS
Analogue output mode : 0v to 5V
Analogue output mode : -5V to 5V
Output Channel 1 : Polar angle (wrap mode = 360 degree wrap)
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : Polar Magnitude (full range = 5m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 4 : Status (status mode = Good signal high, bad signal low)
Output Channel 4 : Speed of Sound

THE NEW CONFIGURATION IN FULL

==== DETAILS OF DETECTED ANEMOMETER ========================

Serial Number : X153802

---- SW VERSION --------------------------------------------
Anemometer Firmware Version: 2329-701

---- HW DETAILS OF ANEMOMETER ------------------------------
Detected Anemometer Class : WindMaster 20Mhz
Number of AXIS on this unit : 3
DAC resolution of attached unit : 14
Number of DAC channels : 4
Maximum sampling rate : (8) P8 20 Hz

---- USER CONFIGURATION OF ANEMOMETER ----------------------
Baud Rate : 19200
Baud Rate : 57600
Output Message Format : M2 Polar
Output Message Format : M1 UV(W)
Output Message Velocity Units : m/s
Output Message Terminator :
Retries : ON
Instantaneous Sampling : OFF
Factory Calibration : ON
Physical Communications : Auto
Resolution settings : Normal resolution
Resolution settings : High resolution
Alignment settings : North to spar
Averaging settings : Averaging inactive
Output rate : P1 1 Hz
Output rate : P8 20 Hz
Polar angle minimum magnitude: 0.050 m/s
Power-On Message : Display Power On Message
ASCII Format : Comma Separated (CSV)
SOS and Sonic Temperature display : Neither
SOS and Sonic Temperature display : SOS
Analogue output mode : 0v to 5V
Analogue output mode : -5V to 5V
Unit Identifier : 'Q'
Output Channel 1 : Polar angle (wrap mode = 360 degree wrap)
Output Channel 1 : U output (full range = 30m/s)
Output Channel 2 : Polar Magnitude (full range = 5m/s)
Output Channel 2 : V output (full range = 30m/s)
Output Channel 3 : W output (full range = 5m/s)
Output Channel 4 : Status (status mode = Good signal high, bad signal low)
Output Channel 4 : Speed of Sound
Synchronised Polling : Off

LI-7500A 75H-2180 2016-12-23 Lab none Calibration Lab none Last calibration: 2016-08-30
AGC=99.5 SDM=6 DAC1=- DAC2=-
Ambient P=99.4kPa, Tair=16.12C, Tdew=9.33C
Zero Air at 300psi, God CO2 421.22 at 400 Tdewspan=14.00
CO2 zero 0.9142 -> 0.9145 Offset: 1.55ppm
CO2 span 1.0227 -> 1.0287 Offset: -3.1ppm
H2O zero 0.8858 -> 0.8836 Offset: -13.6ppm
H2O span 1.0211 -> 1.0124 Offset: 0.17ppm
Notes:
LI-7500A 75H-2668 2016-12-23 Lab none Calibration Lab none Last calibration: 2016-12-06
AGC=102.2 SDM=6 DAC1=- DAC2=-
Ambient P=99.57kPa, Tair=14.95C, Tdew=C
Zero Air at 300psi, God CO2 421.22 at 400 Tdewspan=14.00
CO2 zero 0.9385 -> 0.9378 Offset: -2.65ppm
CO2 span 1.0171 -> 1.0000 Offset: 9.43ppm
H2O zero 0.9017 -> 0.9006 Offset: -7.5ppm
H2O span 1.0841 -> 0.9663 Offset: 2.45ppm
Notes: Recal after box failed in the field after getting rain inside because the door was left open. Pressure about 0.5kPa low. Large offsets. Waves with 1 to 2 min period in data at steady state. Sending to licor for check.
LI-7500A 75H-2514 2016-12-23 Lab none Calibration Lab none Last calibration: 2016-09-12
AGC=103.3 SDM=1 DAC1=- DAC2=-
Ambient P=99.98kPa, Tair=15.95C, Tdew=7.22C
Zero Air at 300psi, God CO2 421.22 at 400 Tdewspan=14.00
CO2 zero 0.9054 -> 0.9054 Offset: -0.1ppm
CO2 span 1.0092 -> 1.0135 Offset: -2.47ppm
H2O zero 0.8371 -> 0.8356 Offset: -10.3ppm
H2O span 1.0239 -> 1.0225 Offset: 0.04ppm
Notes: First calibration after upgrading embedded software to v8.0.0. Changed bandwidth to 10Hz for 20Hz sampling.
LI-7550 AIU-1557 2016-12-23 Lab none Upgrade Lab none I updated the embedded software from v7.5.0 to v8.0.0 and added settings for use at non-CH4 site (TA).
LI-7500A 75H-2668 2016-12-21 Twitchell Alfalfa Gas Analyzer CO2/H2O Removed Lab none This head was deployed with LI7550 box sn AIU-0674 when the box was damaged in the field. Could not determine in the field if the head was working so brought it back to the lab for testing.
LI-7550 AIU-0674 2016-12-21 Twitchell Alfalfa Interface LI-7550 Removed Lab none This 7550 box was damaged in the field. The door was left open and the circuit board got wet. Brought back to the lab for testing and repairs. LI7500 head sn 75H-2668 was deployed with the unit when it was damaged.
LI-7500A 75H-2182 2016-12-21 Lab none Installed Twitchell Rice Gas Analyzer CO2/H2O Installed at Twitchell Rice after calibration in the lab to replace 75H-2180.
LI-7500A 75H-2180 2016-12-21 Twitchell Rice Gas Analyzer CO2/H2O Removed Lab none Removed from Twitchell Rice for calibration in the lab. Replaced by 75H-2182.
LI-7500A 75H-2669 2016-12-21 Lab none Installed East End Gas Analyzer CO2/H2O Installed at East End after calibration in the lab to replace 75H-2514
LI-7500A 75H-2514 2016-12-21 East End Gas Analyzer CO2/H2O Removed Lab none Removed from East End for calibration in the lab. Replaced by 75H-2669.
LI-7500A 75H-2667 2016-12-21 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed at West Pond after calibration in the lab to replace 75H-2150
LI-7500A 75H-2150 2016-12-21 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from West Pond for calibration in the lab. Replaced by 75H-2267
LI-7500A 75H-2669 2016-12-19 Lab none Calibration Lab none Last calibration: 2016-07-19
AGC=104.5 SDM=6 DAC1=- DAC2=-
Ambient P=101.24kPa, Tair=21.21C, Tdew=0.0C
Zero Air at 300psi, God CO2 421.22 at 400 Tdewspan=16.00
CO2 zero 0.919 -> 0.9184 Offset: -2.14ppm
CO2 span 1.0141 -> 1.0119 Offset: 1.13ppm
H2O zero 0.9987 -> 0.8989 Offset: 25.1ppm
H2O span 1.0143 -> 1.0109 Offset: 0.07ppm
Notes:
LI-7500A 75H-2667 2016-12-19 Lab none Calibration Lab none Last calibration: 2016-08-03
AGC=107.0 SDM=6 DAC1=- DAC2=-
Ambient P=101.23kPa, Tair=21.56C, Tdew=-1.32C
Zero Air at 300psi, God CO2 421.22 at 400 Tdewspan=16.00
CO2 zero 0.9427 -> 0.9421 Offset: -2.08ppm
CO2 span 1.0156 -> 1.0156 Offset: -0.02ppm
H2O zero 0.8333 -> 0.8378 Offset: 28.1ppm
H2O span 1.015 -> 1.0214 Offset: -0.13ppm
Notes:
LI-7500A 75H-2182 2016-12-19 Lab none Calibration Lab none Last calibration: 2016-08-30
AGC=87.2 SDM=6 DAC1=- DAC2=-
Ambient P=101.31kPa, Tair=20.29C, Tdew=-0.86C
Zero Air at 300psi, God CO2 421.22 at 400 Tdewspan=16.00
CO2 zero 0.8831 -> 0.8837 Offset: 2.75ppm
CO2 span 1.0142 -> 1.0207 Offset: -3.48ppm
H2O zero 0.8464 -> 0.8457 Offset: -4.2ppm
H2O span 1.023 -> 1.0186 Offset: 0.09ppm
Notes:
CS655 17135 2016-11-22 Lab none Installed Bouldin Alfalfa Soil Moisture Probe B Installed at the Bouldin Alfalaf Tower to replace the on destroyed by the mower. It was buried horizontally at about 2.5cm depth. The data logger program was not updated for the new SDM address until 2016-12-21
HFP01 11334 2016-11-22 Lab none Installed Bouldin Alfalfa Heat Flux Plate B Reinstalled at the Bouldin Alfalfa tower. This SHF plate was installed about 1m west of the tower at about 1cm deep.
WM 1590 124215 2016-12-19 Lab none Maintenance Lab none This was outputting sonic T instead of SOS. I changed its settings to output SOS. Everything else looked okay. Its firmware is 2329-114 - so far not update tool to fix w bug for this version.
WMP 1352 000021 2016-10-25 Lab none Loaned Loaned Loaned to Sam in Whendee Silver's lab.
LI-7500A 0073 2016-12-07 Lab none Installed Tonzi-Floor Installed on the Tonzi Tower to replaced sn 0976 for calibration.
LI-7500 0976 2016-12-07 Tonzi-Floor Removed Lab none Removed from the Tonzi Floor for calibration. Replaced by sn 0073.
LI-7500A 0035 2016-12-07 Lab none Installed Tonzi-Tower Installed on the Tonzi Tower to replaced sn 0418 for calibration.
LI-7500A 0418 2016-12-07 Tonzi-Tower Removed Lab none Removed from the Tonzi Tower for calibration. Replaced by sn 0035.
LI-7500A 0041 2016-12-07 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira Tower to replace sn 0042 for calibration.
LI-7500A 0042 2016-12-07 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira Tower for calibration. Replaced by sn 0041.
LI-7500A 0065 2016-12-07 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed on the Bouldin Alfalfa tower to replace sn 75H-2667 for calibration.
LI-7500A 75H-2667 2016-12-07 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Alfalfa for calibration. Replaced by sn 0065
LI-7500A 75H-2668 2016-12-06 Lab none Installed Twitchell Alfalfa Gas Analyzer CO2/H2O Installed on the Twitchell Alfalfa tower to replace sn 75H-2182 for calibration
CS451 14011129 2016-11-17 Lab none Installed Vaira Pond Installed at the main pond at Varia. It was installed in the pond on a wooden stake at 11.75ft of water.
CR10X 31307 2016-11-17 Lab none Installed Vaira Pond Installed at the main pond at Varia. This will collect water level data.
LI-7500A 75H-2182 2016-12-06 Twitchell Alfalfa Gas Analyzer CO2/H2O Calibration Lab none Removed from the Twitchell Alfalfa tower for calibration. Replaced by sn 75H-2668.
LI-7500A 0073 2016-12-05 Lab none Calibration Lab none Last calibration: 2016-05-05
AGC=59 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.65kPa, Tair=20.82C, Tdew=8.53C
Zero Air at 400psi, God CO2 421.22 at 450 Tdewspan=16.00
CO2 zero 0.9579 -> 0.9576 Offset: ppm
CO2 span 1.0029 -> 1.0707 Offset: -0.30ppm
H2O zero 0.7385 -> 0.7391 Offset: 0.09ppm
H2O span 1.0159 -> 1.0255 Offset: -0.02ppm
Notes:
LI-7500A 0041 2016-12-05 Lab none Calibration Lab none Last calibration: 2016-05-05
AGC=51 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.66kPa, Tair=20.72C, Tdew=8.12C
Zero Air at 400psi, God CO2 421.22 at 450 Tdewspan=16.00
CO2 zero 1.0002 -> 1.0004 Offset: 0.0ppm
CO2 span 1.0128 -> 1.0124 Offset: 0.36ppm
H2O zero 0.8257 -> 0.8255 Offset: -1.0ppm
H2O span 1.0037 -> 1.0000 Offset: -0.01ppm
Notes: Had trouble connecting sensor to software at first. Software froze in middle so re-started software to do calibration.
LI-7500A 0035 2016-12-05 Lab none Calibration Lab none Last calibration: 2016-05-05
AGC=50 SDM=7 DAC1=0-1000 DAC2=10-25
Ambient P=100.73kPa, Tair=20.92C, Tdew=9.8C
Zero Air at 400psi, God CO2 421.22 at 450 Tdewspan=16.00
CO2 zero 0.9388 -> 0.9390 Offset: 0.89ppm
CO2 span 1.0216 -> 1.0265 Offset: -2.1ppm
H2O zero 0.9018 -> 0.9025 Offset: 4.1ppm
H2O span 0.9928 -> 1.0029 Offset: -0.27ppm
Notes:
LI-7500A 0065 2016-12-06 Lab none Calibration Lab none Last calibration: 2016-05-13
AGC=74.7 SDM=6 DAC1= DAC2=
Ambient P=100.29kPa, Tair=16.04C, Tdew=0.60C
Zero Air at 350psi, God CO2 421.22 at 400 Tdewspan=16.02
CO2 zero 0.8898 -> 0.8899 Offset: 0.36ppm
CO2 span 1.0149 -> 1.0142 Offset: 0.43ppm
H2O zero 0.7086 -> 0.7086 Offset: 0.2ppm
H2O span 1.0139 -> 1.0127 Offset: 0.01ppm
Notes: H2O span done later in the day.
LI-7500A 75H-2668 2016-12-06 Lab none Calibration Lab none Last calibration: 2016-05-20
AGC=103.3 SDM=6 DAC1= DAC2=
Ambient P=100.29kPa, Tair=15.36C, Tdew=3.00C
Zero Air at 350psi, God CO2 421.22 at 400 Tdewspan=16.02
CO2 zero 0.9385 -> 0.9385 Offset: 0.0ppm
CO2 span 1.0136 -> 1.0171 Offset: -1.92ppm
H2O zero 0.8969 -> 0.9017 Offset: 30.2ppm
H2O span 1.0146 -> 1.0841 Offset: -1.41ppm
Notes:
PRB-K Type #8 00000FEDDAA1 2016-11-17 Lab none Installed Tonzi Installed at Tonzi near the main tower to replace sn 00000128B8072 that has two broken pins
PRB-K Type #8 00000128B8072 2016-11-17 Tonzi Removed Lab none Removed from Tonzi near the main tower. This probe has two broken pins.
FMA RMT-200 05-0010 2014-05-20 Lab none Damaged Lab none After being moved, but not treated roughly, the analyzer stopped reporting CH4. The ring time is variable from 2 to 20usec and the spectrum looks odd. Cleaning the mirrors made no difference. LGR thinks there is something wrong with the laser. At this time there is an 8 week wait for repairs and a minimum cost of 00 for cleaning and calibration - any repairs would be an additional cost. We don't need this analyzer right now and will wait to repair it.
U24-002-C 11020258 2016-11-16 Lab none Installed Mayberry Conductivity Sensor B The HOBO conductivity sensor was put in the water near the existing conductivity sensor at the southeast corner of the scaffolding for an intercomparison with the Campbell CS547A sensor b/c we're technically out of the CS547A range. It is tied to the tower with some rope.
CR10X 31307 2016-11-15 Lab none Maintenance Lab none The internal lithium battery was dead. I replaced the battery.
PQS 1 140457 2016-11-02 Vaira PAR Sensor Incoming Cal Removed Lab none Removed from Vaira with the radiometer inter-comparison system that was on a stand alone tripod.
NR01 2420 2016-11-02 Vaira Removed Lab none Removed from Vaira with the radiometer inter-comparison system that was on a stand alone tripod.
NR01 1720 2016-11-02 Vaira Removed Lab none Removed from Vaira with the radiometer inter-comparison system that was on a stand alone tripod.
CR6 2113 2016-11-02 Vaira Removed Lab none Removed from Vaira with the radiometer inter-comparison system that was on a stand alone tripod.
AM16-32A 1238 2016-11-02 Vaira Removed Lab none Removed from Vaira with the radiometer inter-comparison system that was on a stand alone tripod.
LI-7700 TG1-0224 2016-10-27 Lab none Loaned Loaned Loaned to Frank Anderson. He will use it temporarily at his Nisqually Marsh site while his 7700 is tested for no signal problems.
LI-7700 TG1-0224 2016-10-27 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2014-03-07
RSSI=79.3%, Optical RH=17%
Ambient P=100.4kPa, Tair=22.8C
Zero Air at 400psi, CH4 2nd 4.995 at 500
CH4 Zero Offset: 0.042ppm
CH4 Span Offset: 0.057ppm
Notes: Last calibration at Licor after replacement of laser due to unknown failure.
LI-7500A 75H-2176 2016-10-25 Lab none Calibration Lab none Last calibration: 2016-06-03
AGC=91.2 SDM=6 DAC1=- DAC2=-
Ambient P=100.28kPa, Tair=20.11C, Tdew=12.3C
Zero Air at 400psi, God CO2 421.22 at 450 Tdewspan=16.01
CO2 zero 0.8688 -> 0.8689 Offset: 0.22ppm
CO2 span 1.0175 -> 1.0176 Offset: -0.12ppm
H2O zero 0.8571 -> 0.8563 Offset: -4.9ppm
H2O span 1.0281 -> 1.0134 Offset: 0.31ppm
Notes:
DO6400/T DO-06 2016-10-25 Lab none Maintenance Lab none This sensor was serviced in the lab. The probe was taken apart and cleaned with vinegar and DI water. The membrane was replace and the probe was reassembled with fresh solution. In ambient air it was reading 80.0mV
DO6400/T DO-06 2016-10-26 Lab none Installed East End DO Sensor Re-installed at East End tower after testing in the lab.
LI-7500A 75H-2176 2016-10-26 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at Mayberry tower after lab calibration to replace 75H-2669
LI-7500A 75H-2669 2016-10-26 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from Mayberry tower for calibration. Replaced by 75H-2176
GP-001 GP20150111 2016-10-26 Lab none Installed Mayberry CO2 Probe B Installed probe at original 'mid' level with other water sensors. This level is at water surface right now due to low water table. This probe has its cable connector painted and covered with silicon sealant.
PQS 1 140457 2016-10-20 Twitchell Rice Moved Vaira PAR Sensor Incoming Cal Moved with radiometer test tripod from Twitchell Rice to Vaira Ranch to compare against Ione sensors.
NR01 2420 2016-10-20 Twitchell Rice Moved Vaira Moved with radiometer test tripod from Twitchell Rice to Vaira Ranch to compare against Ione sensors.
NR01 1720 2016-10-20 Twitchell Rice Moved Vaira Moved with radiometer test tripod from Twitchell Rice to Vaira Ranch to compare against Ione sensors.
CR6 2113 2016-10-20 Twitchell Rice Moved Vaira Moved with radiometer test tripod from Twitchell Rice to Vaira Ranch to compare against Ione sensors.
AM16-32A 1238 2016-10-20 Twitchell Rice Moved Vaira Moved with radiometer test tripod from Twitchell Rice to Vaira Ranch to compare against Ione sensors.
AM16/32A 1236 2016-10-13 Tonzi Removed Lab none Removed from the non-functioning sap flow station north of the road
CR10X 47772 2016-10-13 Tonzi Removed Lab none Removed from the non-functioning sap flow station north of the road
AM16/32A 1430 2016-10-13 Tonzi Removed Lab none Removed from the non-functioning sap flow station within the Tower cow fence
AM16/32A 1429 2016-10-13 Tonzi Removed Lab none Removed from the non-functioning sap flow station within the Tower cow fence
CR10X 48042 2016-10-13 Tonzi Removed Lab none Removed from the non-functioning sap flow station within the Tower cow fence
LI-820 CGA-556 2016-10-13 Tonzi-Tower Calibration Tonzi-Tower none Last calibration was 2016-09-13 Cell T = 52.2, Pressure = 96.2, N2 at 1150psi, CO2 406ppm at 1800
zero offset = -1.5ppm
Span offset = -7.5ppm
Zero 0.9429 -> 0.9424
Span 0.8368 -> 0.8492
DO6400/T DO-06 2016-10-11 East End DO Sensor Removed Lab none Removed from East End for service. It was out od the water, reading high and had bite marks on the cable.
miniDOT 1220 2016-10-11 Mayberry DO Sensor D Moved Mayberry DO Sensor F Due to the lowering of the water to flush the salt, this sensor was no longer in the water, but could not be removed because it was rusted to the chain. It was moved to the bottom/lower location by inverting the chain.
miniDOT 1221 2016-10-11 Mayberry DO Sensor F Removed Lab none Removed from Mayberry due to lowering water for flushing salt. This sensor was at the bottom location. The top sensor (sn 1220) was no longer in water, but could not be removed due to rust and was moved to the bottom location.
GP-001 GP20140024 2016-10-11 Mayberry CO2 Probe C Removed Lab none Removed from Mayberry due to odd readings and lowering water for flushing salt
GP-001 GP20140025 2016-10-11 Mayberry CO2 Probe B Removed Lab none Removed from Mayberry due to odd readings and lowering water for flushing salt
LI-7700 TG1-0357 2016-10-11 Mayberry Gas Analyzer CH4 Calibration Mayberry Gas Analyzer CH4 Calibration Location: Mayberry
Last calibration: 2016-04-13
RSSI=73.6%, Optical RH=13%
Ambient P=101.6kPa, Tair=18.3C
zero air at 850psi, AMP CH4 4.030 at 1900
CH4 Zero Offset: 0.020ppm
CH4 Span Offset: 0.025ppm
Notes: Windy.
LI-7700 TG1-0423 2016-10-11 Sherman Wetland Gas Analyzer CH4 Calibration Sherman Wetland Gas Analyzer CH4 Calibration Location: Sherman Wetland
Last calibration: 2016-05-20
RSSI=78.7%, Optical RH=23%
Ambient P=101.6kPa, Tair=16.4C
zero air at 850psi, AMP CH4 4.030 at 1900
CH4 Zero Offset: 0.026ppm
CH4 Span Offset: 0.023ppm
Notes: Windy.
LI-7700 TG1-0223 2016-10-11 West Pond Gas Analyzer CH4 Calibration West Pond Gas Analyzer CH4 Calibration Location: west pond
Last calibration: 2016-04-13
RSSI=71%, Optical RH=27%
Ambient P=101.3kPa, Tair=22C
zero air at 700psi, AMP CH4 4.030 at 1200
CH4 Zero Offset: 0.001ppm
CH4 Span Offset: -0.006ppm
Notes: Breezy.
LI-7700 TG1-0155 2016-10-11 East End Gas Analyzer CH4 Calibration East End Gas Analyzer CH4 Calibration Location: East End
Last calibration: 2016-04-13
RSSI=82%, Optical RH=21%
Ambient P=101.6kPa, Tair=21.1C
zero air at 800psi, AMP CH4 4.030 at 1400
CH4 Zero Offset: 0.020ppm
CH4 Span Offset: 0.010ppm
Notes: Breezy
LI-7700 TG1-0211 2016-10-11 Twitchell Rice Gas Analyzer CH4 Calibration Twitchell Rice Gas Analyzer CH4 Calibration Location: Twitchell Rice
Last calibration: 2016-04-13
RSSI=68.0%, Optical RH=26%
Ambient P=101.6kPa, Tair=20.8C
zero air at 850psi, AMP CH4 4.030 at 1600
CH4 Zero Offset: -0.01ppm
CH4 Span Offset: 0.021ppm
Notes: windy, rice just harvested, tower out of field.
LI-7500A 75H-2514 2016-09-12 Lab none Calibration Lab none Last calibration: 2016-06-03
AGC=103.2 SDM=6 DAC1=- DAC2=-
Ambient P=99.87kPa, Tair=18.64C, Tdew=9.66C
Zero Air at 400psi, God CO2 421.22 at 400 Tdewspan=16.0
CO2 zero 0.9047 -> 0.9054 Offset: 2.80ppm
CO2 span 1.0055 -> 1.0092 Offset: -2.07ppm
H2O zero 0.8368 -> 0.8371 Offset: -10.5ppm
H2O span 1.015 -> 1.0239 Offset: -0.18ppm
Notes:
LI-7500A 75H-2150 2016-09-12 Lab none Calibration Lab none Last calibration: 2016-05-13
AGC=99.0 SDM=6 DAC1=- DAC2=-
Ambient P=99.83kPa, Tair=19.51C, Tdew=10.68C
Zero Air at 400psi, God CO2 421.22 at 400 Tdewspan=10.6
CO2 zero 0.902 -> 0.9019 Offset: -0.58ppm
CO2 span 1.008 -> 1.0084 Offset: -0.29ppm
H2O zero 0.8668 -> 0.8662 Offset: -4.1ppm
H2O span 1.0109 -> 1.0155 Offset: -0.1ppm
Notes:
CS655 17134 2016-10-07 Lab none Disposed Disposed This sensors was hit by s mower. One tine was ripped off and the other twisted. The cable is cut near the body. Put in the dead sensor box in the lab.
WM 1590 153806 2016-10-06 Lab none Installed Bouldin Alfalfa Anemometer Reinstalled on the Bouldin Alfalfa tower replacing sn 124215. This sonic is working correctly. Spikes were caused by the soil moisture probes powering up and browning out the system.