Equipment and Calibration History

View Equipment |

View history entry 2670 for Licor LI-7700 sn: TG1-0357

Date: 2017-03-23
From: Mayberry - Gas Analyzer CH4
To: Mayberry - Gas Analyzer CH4
Title: Calibration
Notes: Calibration Location: Mayberry Last calibration: 2016-10-11 RSSI=77.3%, Optical RH=17% Ambient P=102.1kPa, Tair=17.0C Zero Air at 750psi, Ameriflux CH4 4.0304 at 600 CH4 Zero Offset: 0.032ppm CH4 Span Offset: -0.0277ppm Notes: Changed dessicant and flushed out head with dry air.
Entered by:
Field notes