Equipment and Calibration History

View Equipment |

View history entry 2661 for Licor LI-7700 sn: TG1-0224

Date: 2017-03-22
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Calibration Location: Lab Last calibration: 2016-10-27 RSSI=67.3%, Optical RH=22% Ambient P=100.9kPa, Tair=22.7C Zero Air at 1700psi, NOAA CH4 1.89919 at 1750 CH4 Zero Offset: 0ppm CH4 Span Offset: 0.001ppm Notes: Changed dessicant
Entered by: