View history entry 2660 for Licor LI-7700 sn: TG1-0424

Date: 2017-03-22
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Calibration Location: Lab Last calibration: 2015-07-23 RSSI=84.6%, Optical RH=26% Ambient P=100.7kPa, Tair=23.2C Zero Air at 1700psi, NOAA CH4 1.89919 at 1750 CH4 Zero Offset: 0.05ppm CH4 Span Offset: -0.004ppm Notes: Changed dessicant. Previous calibration was at Licor.