Skip to main content
rice field banner biometlab logo
   
   
 
   
   
   
   
   
 
 

Equipment and Calibration

View History

Equipment Entry

Instrument:Multiplexer
Manufacture:Campbell Sci.
Model No:AM25T
Serial No:4852
Tag No:
Acquisition Date:0000-00-00
Project:NSF-Methane
Location on 2017-09-26:Lab,
Notes:

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
3092 2017-09-26 East End-Temp2 Multiplexer Thermocouples Removed Lab none Brought back to the lab with the decommissioning of the East End closed canopy temporary tower 2
2864 2017-06-14 Lab none Installed East End-Temp2 Multiplexer Thermocouples Installed on the East End Temporary Tower #2 Closed Canpoy location with the initial tower set up.
2631 2017-02-24 Factory Returned Lab none Returned from Campbell after repair for: 12V screw terminal damage, internal thermistor damage.
Campbell says: Able to verify customer problem. Replaced terminal connector (P22C) 12V connector. Replaced internal thermistor (Z16) due to leads being broken off. Bench tested unit, calibrated unit with documentation, final tested to verify correct operation.
2632 2017-02-21 Factory Calibration Factory Calibrated by Campbell. Good for two years.
1939 2015-09-08 Lab none Maintenance Lab none I took this multiplexer apart and cleaned it. Some of the screw terminals were tough to turn - I sprayed them with WD40. There was a little bit of corrosion on and near the 12V terminal.
1850 2015-05-12 Sherman Island Multiplexer Thermocouple Removed Lab none Removed with the decommissioning of the Sherman Cow Pasture site.
37 2010-02-05 Sherman Island Multiplexer Thermocouple Location Sherman Island Multiplexer Thermocouple Currently connected to the met CR10X sn47253 at Sherman. mostly soil temeraptures.
 
 

Login