CS451 70011234 | 2017-06-23 | Sherman Wetland Water Level | Removed | Lab none | Removed from the Sherman Wetland site due to NaNs in the data stream. |
N940-5APE.W 03387648 | 2017-06-02 | Lab none | Loaned | Loaned | Loaned to Patty Oikawa at CSU East Bay for use with her LGR isotope system to be deployed at the Sherman wetland site. |
KD140SX-UFBS 11ZV0M0855 | 2017-06-23 | Lab none | Installed | East End-Temp3 Solar Panel A | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
KD140SX-UFBS 11ZV0M0836 | 2017-06-23 | Lab none | Installed | East End-Temp3 Solar Panel B | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
U24-002-C 11020258 | 2017-06-23 | Lab none | Installed | East End-Temp3 Conductivity Sensor | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
LI-7700 TG1-0424 | 2017-06-23 | Bouldin Alfalfa Gas Analyzer CH4 | Installed | East End-Temp3 Gas Analyzer CH4 | Moved from Bouldin Alfalfa to be installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
LI-7500 0976 | 2017-06-23 | Lab none | Installed | East End-Temp3 Gas Analyzer CO2/H2O | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
WM 1590 120907 | 2017-06-23 | Lab none | Installed | East End-Temp3 Anemometer | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
CS451 20010046 | 2017-06-23 | Lab none | Installed | East End-Temp3 Water Level | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
HFP01 000137 | 2017-06-23 | Lab none | Installed | East End-Temp3 Heat Flux Plate B | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
HFP01SC 00245 | 2017-06-23 | Lab none | Installed | East End-Temp3 Heat Flux Plate A | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
AM16-32A 1238 | 2017-06-23 | Lab none | Installed | East End-Temp3 Multiplexer | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
NL115 6761 | 2017-06-23 | Lab none | Installed | East End-Temp3 Interface NL115 | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
CR1000 50889 | 2017-06-23 | Lab none | Installed | East End-Temp3 Data Logger Met | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
CR1000 23865 | 2017-06-23 | Lab none | Installed | East End-Temp3 Data Logger Flux | Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site. |
KD140SX-UFBS 11ZV0M0828 | 2017-06-14 | East End-Temp2 none | Installed | East End-Temp2 Solar Panel A | Installed on the East End Temporary Tower #2 Closed Canpoy location with the initial tower set up. |
KD140SX-UFBS 11ZV0M0829 | 2017-06-14 | East End-Temp2 none | Installed | East End-Temp2 Solar Panel B | Installed on the East End Temporary Tower #2 Closed Canpoy location with the initial tower set up. |
PAR-LITE 050597 | 2017-06-14 | Lab none | Installed | East End-Temp2 PAR Sensor Reflected | Installed on the East End Temporary Tower #2 Closed Canpoy location with the initial tower set up. |
NR-LITE 021029 | 2017-06-14 | Lab none | Installed | East End-Temp2 Net Radiometer | Installed on the East End Temporary Tower #2 Closed Canpoy location with the initial tower set up. |
HMP45AC U3030046 | 2017-06-14 | Lab none | Installed | East End-Temp2 RH & Temp | Installed on the East End Temporary Tower #2 Closed Canopy location with the initial tower set up. |
CS451 70011236 | 2017-06-14 | Lab none | Installed | East End-Temp2 Water Level | Installed on the East End Temporary Tower #2 Closed Canpoy location with the initial tower set up. |
HFP01 10809 | 2017-06-14 | Lab none | Installed | East End-Temp2 Heat Flux Plate B | Installed on the East End Temporary Tower #2 Closed Canpoy location with the initial tower set up. |
HFP01 6499 | 2017-06-14 | Lab none | Installed | East End-Temp2 Heat Flux Plate A | Installed on the East End Temporary Tower #2 Closed Canpoy location with the initial tower set up. |
AM25T 4852 | 2017-06-14 | Lab none | Installed | East End-Temp2 Multiplexer Thermocouples | Installed on the East End Temporary Tower #2 Closed Canpoy location with the initial tower set up. |
CR6 2115 | 2017-06-14 | Lab none | Installed | East End-Temp2 Data Logger | Installed on the East End Temporary Tower #2 Closed Canpoy location with the initial tower set up. |
LI-7550 AIU-0614 | 2017-06-14 | Lab none | Installed | East End-Temp2 Gas Analyzer CO2/H2O | Installed on the East End Temporary Tower #2 Closed Canpoy location with the initial tower set up. |
PS-15M 12501584 | 2017-06-01 | Lab none | Installed | East End-Temp1 Solar Controller | Installed at the East End Temporary Tower #1 Open Water location with the initial set up of this site. |
KD140SX-UFBS 11ZV0M0835 | 2017-06-01 | Twitchell Rice 0 | Moved | East End-Temp1 Solar Panel A | Installed at the East End Temporary Tower #1 Open Water location with the initial set up of this site. |
CS450 20010770 | 2017-06-01 | Lab none | Installed | East End-Temp1 Water Level | Installed at the East End Temporary Tower #1 Open Water location with the initial set up of this site. |
HFP01 10822 | 2017-06-01 | Lab none | Installed | East End-Temp1 Heat Flux Plate B | Installed at the East End Temporary Tower #1 Open Water location with the initial set up of this site. |
HFP01 001604 | 2017-06-01 | Lab none | Installed | East End-Temp1 Heat Flux Plate A | Installed at the East End Temporary Tower #1 Open Water location with the initial set up of this site. |
CR6 2114 | 2017-06-01 | Lab none | Installed | East End-Temp1 Data Logger Met | Installed at the East End Temporary Tower #1 Open Water location with the initial set up of this site. |
AM25T 2856 | 2017-06-01 | Lab none | Installed | East End-Temp1 Multiplexer Thermocouples | Installed at the East End Temporary Tower #1 Open Water location with the initial set up of this site. |
U24-002-C 11020258 | 2017-05-25 | Mayberry Conductivity Sensor B | Removed | Lab none | Removed from Mayberry for use on the temporary tower at East End |
LI-7500A 0073 | 2017-06-23 | Lab none | Installed | Bouldin Alfalfa Gas Analyzer CO2/H2O | Installed at Bouldin Alfalfa to replace 75H-2667. |
LI-7500A 75H-2667 | 2017-06-23 | Bouldin Alfalfa Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Bouldin Alfalfa for calibration; replaced by 75H-0073. |
SRS-Ni NDVI 2046503404 | 2017-06-23 | Lab none | Installed | Sherman Wetland Narrow Band NDVI Reflected | Installed at Sherman Wetland to replace NDVI_out field stop sensor. |
LI-7500A 75H-2514 | 2017-06-23 | Lab none | Installed | Sherman Wetland Gas Analyzer CO2/H2O | Installed at Sherman Wetland to replace 75H-2669. |
LI-7500A 75H-2669 | 2017-06-23 | Sherman Wetland Gas Analyzer CO2/H2O | Removed | Lab none | Removed for calibration from Sherman Wetland; replaced by 75H-2514. |
LI-7500A 0418 | 2017-06-23 | Lab none | Calibration | Lab none | Slated to replace 75H-2668 at West Pond; measured unreasonable H20 levels and returned to lab for further calibration. |
LI-7500A 75H-2668 | 2017-06-23 | West Pond Gas Analyzer CO2/H2O | Adjusted | West Pond Gas Analyzer CO2/H2O | Slated to be removed for calibration but replacement 75H-0418 measured unreasonable H20 levels. |
miniDOT 1219 | 2017-06-23 | Mayberry DO Sensor D | Moved | East End-Temp3 DO Sensor | Removed MiniDOT (from upper position; it was 1 of 2 sensors in upper position) since water level has been really low and we probably decided 2 sensors was probably enough for a profile. Installed it at EE Temp Tower 3. |
LI-7500A 0073 | 2017-06-22 | Lab none | Calibration | Lab none | Last calibration: 2017-06-21 AGC=100.4 SDM= DAC1= DAC2= Ambient P=99.8kPa, Tair=25.8C, Tdew=14.5C Zero Air at 1450psi, NOAA Air 402.69 at 1600 Tdewspan=16.00 CO2 zero 0.9574 -> .9572 Offset: -1.08ppm CO2 span 1.0075 -> 1.0084 Offset: -.35ppm H2O zero 0.7305 -> .7307 Offset: 1.2ppm H2O span 1.0071 -> 1.0045 Offset: -.06ppm Notes: Matches well with calibration from previous day. |
LI-7500A 0073 | 2017-06-21 | Lab none | Calibration | Lab none | Last calibration: 2017-06-13 AGC=101.1 SDM= DAC1= DAC2= Ambient P=99.3kPa, Tair=24.2C, Tdew=12.7C Zero Air at 1450psi, NOAA Air 402.69 at 1600 Tdewspan=16.00 CO2 zero 0.9567 -> .9574 Offset: 2.42ppm CO2 span 1.0065 -> 1.0075 Offset: -.67ppm H2O zero 0.7342 -> .7305 Offset: -21.8ppm H2O span 1.0251 -> 1.0071 Offset: .4ppm Notes: Checking the zero and spans after upgrade from 7500 to 7500A. There are jumps in CO2 zero and span so I will check again tomorrow. |
LI-7500A 0418 | 2017-06-21 | Lab none | Calibration | Lab none | Last calibration: 2017-06-20 AGC=96.8 SDM= DAC1= DAC2= Ambient P=99.4kPa, Tair=24.7C, Tdew=12.2C Zero Air at 1450psi, NOAA Air 402.69 at 1600 Tdewspan=16.00 CO2 zero 0.8994 -> .8992 Offset: -.84ppm CO2 span 0.9961 -> .9985 Offset: -1.31ppm H2O zero 0.8423 -> .8432 Offset: 4.9ppm H2O span 0.9658 -> .9806 Offset: -0.33ppm Notes: Re-testing zero and spans after large change in H2O zero/span and a smaller change in CO2 zero after upgrade from 7500 to 7500A. The coefficients seem stable now. The changes in H2O and CO2 zeros after upgrade might be from changing out the old chemicals for fresh chemicals during the upgrade. |
LI-7500A 0418 | 2017-06-20 | Lab none | Calibration | Lab none | Last calibration: 2017-06-13 AGC=95.4 SDM= DAC1= DAC2= Ambient P=99.5kPa, Tair=26.9C, Tdew=13.2C Zero Air at 1500psi, NOAA Air 402.69 at 1650 Tdewspan=16.00 CO2 zero 0.9003 -> .8994 Offset: -3.25ppm CO2 span 1.0066 -> .9961 Offset: 5.59ppm H2O zero 0.841 -> .8423 Offset: 7.6ppm H2O span 1.2209 -> .9658 Offset: 5.25ppm Notes: Checking the zero and spans again because there was a large H2O zero offset last time. H2O zero is pretty stable, but there was a relatively large shift in H2O span. I'll re-test tomorrow. |
LI-7500A 75H-2514 | 2017-06-20 | Lab none | Calibration | Lab none | Last calibration: 2017-03-14 AGC=102.2 SDM= DAC1= DAC2= Ambient P=100.9kPa, Tair=25.7C, Tdew=13.2C Zero Air at 1500psi, NOAA Ait 402.69 at 1650 Tdewspan=16.00 CO2 zero 0.9046 -> .9041 Offset: -2.08ppm CO2 span 1.0139 -> 1.0115 Offset: 0.96ppm H2O zero 0.8368 -> .8378 Offset: 6.4ppm H2O span 1.0106 -> 1.0019 Offset: 0.20ppm Notes: |
SRS-Ni NDVI 2046503390 | 2017-06-05 | Lab none | Installed | East End Narrow Band NDVI Incoming | Installed at EE in the incoming position. |
LI-7700 TG1-0211 | 2017-06-14 | Lab none | Installed | East End-Temp2 Gas Analyzer CH4 | Installed with initial setup of East End closed vegetation temporary tower for the summer of 2017 |
LI-7500A 75H-2176 | 2017-06-14 | Lab none | Installed | East End-Temp2 Gas Analyzer CO2/H2O | Installed with initial setup of East End closed vegetation temporary tower for the summer of 2017 |
WM 1590 153805 | 2017-06-14 | Lab none | Installed | East End-Temp2 Anemometer | Installed with initial setup of East End closed vegetation temporary tower for the summer of 2017 |
LI-7500A 75H-2150 | 2017-06-14 | Lab none | Installed | East End Gas Analyzer CO2/H2O | Installed at East End after calibration in the lab to replace LI7500 sn 75H-2514. |
LI-7500A 75H-2514 | 2017-06-14 | East End Gas Analyzer CO2/H2O | Removed | Lab none | Removed LI7500 from East End for calibration, replaced with 75H-2150. |
LI-7500A 0418 | 2017-06-13 | Lab none | Calibration | Lab none | Last calibration: 2017-01-12 AGC=100 SDM= DAC1= DAC2= Ambient P=100.2kPa, Tair=23.2C, Tdew=7.69C Zero Air at 1500psi, NOAA Air 402.69 at 1600 Tdewspan=16.00 CO2 zero 0.9009 -> .9003 Offset: -2.33ppm CO2 span 1.0062 -> 1.0066 Offset: -0.23ppm H2O zero 0.857 -> .8410 Offset: -96.9ppm H2O span 0.9788 -> 1.2209 Offset: -4.6ppm Notes: Large H2O Zero offset. First calibration since upgrade from 7500 to 7500A. Copied coefficients for CO2 (A, B, C, D, E, XS, Z) and H2O (A, B, C, XS, Z) from sn00418_20160815post.l75 before the calibration. |
LI-7500A 0073 | 2017-06-13 | Lab none | Calibration | Lab none | Last calibration: 2017-04-07
AGC=100.3 SDM= DAC1= DAC2= Ambient P=100.2kPa, Tair=21.9C, Tdew=11.2C Zero Air at 1500psi, NOAA Air 402.69 at 1600 Tdewspan=16.00 CO2 zero 0.9578 -> .9567 Offset: -3.62ppm CO2 span 1.0036 -> 1.0065 Offset: -1.63ppm H2O zero 0.7388 -> .7342 Offset: -26.9ppm H2O span 1.0096 -> 1.0251 Offset: -0.36ppm Notes: First calibration since upgrade from 7500 to 7500A. Copied coefficients for CO2 (A, B, C, D, E, XS, Z) and H2O (A, B, C, XS, Z) from sn0073_20170118post.l75 before calibration. Took about 9-10 minutes for H2O to go down to 0, took about 1-2 minutes for CO2 to reach span, and took ~9 minutes for H2O to reach span. Set span T to 15.99 in software for a 16.00 Span Dew Point. |
LI-7500A 1594 | 2017-06-05 | Lab none | Installed | Twitchell Alfalfa Gas Analyzer CO2/H2O | Installed at Twitchell Alfalfa to replace sn 75H-2176. |
LI-7500A 75H-2176 | 2017-06-05 | Twitchell Alfalfa Gas Analyzer CO2/H2O | Removed | Lab none | Removed from Twitchell Alfalfa and brought back to the lab for calibration. It was replaced by sn 1594. |
BB-Fab BB-TT | 2017-01-17 | Tonzi | Maintenance | Tonzi | The Tower has been losing power over night recently. All the batteries were disconnected and tested them with the load tester. Four batteries tested bad. We removed them and reconnected the others. |
GP-001 GP20140017 | 2017-06-05 | Lab none | Installed | West Pond CO2 Probe B | Installed at West Pond to replace sn GP2015019. |
GP-001 GP20150109 | 2017-06-05 | West Pond CO2 Probe B | Removed | Lab none | Removed from West Pond because it was giving funky numbers after the water level went down. |
SRS-Ni NDVI 2046503376 | 2017-06-05 | Lab none | Installed | Mayberry Narrow Band NDVI Incoming | Installed in incoming position at Mayberry. |
SRS-Ni NDVI 1066703780 | 2017-06-05 | Lab none | Installed | West Pond Narrow Band NDVI Incoming | Installed in incoming position at West Pond. |
LI-7500A 0073 | 2017-06-12 | Lab none | Upgrade | Lab none | Upgraded to 7500A with the 7500-600 LI-7500 Head Upgrade kit from Licor. I put in new bottom assembly and greased o-ring, screws and greased o-rings, and two chemical bottles. I also adjusted the potentiometer until the waveform displayed about 15,000 counts for the minimum and 55,000 counts for the maximum(Diagnostics icon > Waveform tab). I confirmed the Signal Strength coefficients (B = 1.6, C = 3.2) and reset the signal strength to 100% after cleaning the windows (Calibration icon > Signal Strength tab > Reset Signal Strength). |
LI-7500A 0418 | 2017-06-12 | Lab none | Upgrade | Lab none | Upgraded to 7500A with the 7500-600 LI-7500 Head Upgrade kit from Licor. I put in new bottom assembly and greased o-ring, screws and greased o-rings, and two chemical bottles. I also adjusted the potentiometer until the waveform displayed about 15,000 counts for the minimum and 55,000 counts for the maximum(Diagnostics icon > Waveform tab). I confirmed the Signal Strength coefficients (B = 1.6, C = 3.2) and reset the signal strength to 100% after cleaning the windows (Calibration icon > Signal Strength tab > Reset Signal Strength). |
LI-7500A 75H-2176 | 2017-06-12 | Lab none | Calibration | Lab none | Last calibration: 2017-03-07 AGC=90.4 SDM= DAC1= DAC2= Ambient P=100.4kPa, Tair=19.8C, Tdew=8.83C Zero Air at 1500psi, NOAA Air 402.69 at 1600 Tdewspan=16.00 CO2 zero 0.8691 -> .8690 Offset: -.41ppm CO2 span 1.0195 -> 1.0160 Offset: 1.77ppm H2O zero 0.8561 -> .8572 Offset: 5.9ppm H2O span 1.0102 -> 1.0022 Offset: 0.18ppm Notes: |
LI-7500 0976 | 2017-06-12 | Lab none | Calibration | Lab none | Last calibration: 2017-04-07 AGC=57 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.7kPa, Tair=20.3C, Tdew=7.04C Zero Air at 1500psi, NOAA Air 402.69 at 1600 Tdewspan=16.00 CO2 zero 0.8594 -> .8597 Offset: 1.39ppm CO2 span 1.0097 -> 1.0077 Offset: 1.01ppm H2O zero 0.6348 -> .6335 Offset: -.23ppm H2O span 0.9689 -> 1.0310 Offset: -1.36ppm Notes: Set span temperature in software to 15.92 in order for dew point value to be 16.00 after pressing "Span" |
SRS-Ni NDVI 2046503357 | 2017-06-05 | Lab none | Installed | Bouldin Corn Narrow Band NDVI Reflected | Installed at BC tower to replace downward-facing NDVI sensor with field-stop lens. |
SRS-Nr NDVI 943103015 | 2017-06-05 | Bouldin Corn Narrow Band NDVI Reflected | Removed | Lab none | Removed from BC tower and replaced by a downward-facing NDVI sensor with hemispherical view. |
SRS-Nr NDVI 943203017 | 2017-06-05 | Bouldin Alfalfa Narrow Band NDVI Reflected | Removed | Lab none | Removed because we have already installed a downward-facing NDVI sensor with hemispherical view. |
BB-Fab BB-TW | 2017-06-01 | Twitchell Rice | Removed | Twitchell Rice | The Twitchell Rice solar system was disassembled and removed. Two of the batteries were dead and will be recycled. The four other batteries load tested okay and were taken to Tonzi for the Floor solar power system. The solar panels and charge controller were removed and will be used at the temporary tower at East End this summer. |
HMP45AC T2130005 | 2017-06-01 | Tonzi-Tower | Removed | Lab none | Removed from the higher middle position of CO2 profile system on the Tonzi Tower. At least one has not been working and all need to be compared against each other for drift. |
HMP45AC T1250012 | 2017-06-01 | Tonzi-Tower | Removed | Lab none | Removed from the lower middle position of CO2 profile system on the Tonzi Tower. At least one has not been working and all need to be compared against each other for drift. |
HMP45AC U3030041 | 2017-06-01 | Tonzi-Tower | Removed | Lab none | Removed from the lowest position of CO2 profile system on the Tonzi Tower. At least one has not been working and all need to be compared against each other for drift. |
BB-Fab BB-TF | 2017-06-01 | Tonzi-Floor | Maintenance | Tonzi-Floor | I disconnected all seven batteries and load tested them. Three were very weak, two dipped a little under load but still in the green and two were rock solid. I replaced the three weak ones and the two that dipped under load with four batteries we brought from the old Twitchell Rice solar system. |
LI-7550 AIU-0650 | 2017-06-01 | Lab none | Installed | East End-Temp1 Interface LI-7550 | Installed at open water temporary tower in East End wetland |
LI-7700 TG1-0423 | 2017-06-01 | Lab none | Installed | East End-Temp1 Gas Analyzer CH4 | Installed at open water temporary tower in East End wetland |
LI-7500A 0065 | 2017-06-01 | Lab none | Installed | East End Gas Analyzer T1 CO2/H2O | Installed at open water temporary tower in East End wetland |
WM 1590 153803 | 2017-06-01 | Lab none | Installed | East End-Temp1 Anemometer | Installed at open water temporary tower in East End wetland |
GP-001 GP20140009 | 2017-05-31 | Factory | Disposed | Disposed | Disposed of at Eosense factory because it was not repairable |
GP-001 GP20140010 | 2017-05-31 | Factory | Disposed | Disposed | Disposed of at Eosense factory because it was not repairable |
GP-001 GP20140021 | 2017-05-31 | Factory | Disposed | Disposed | Disposed of at Eosense factory because it was not repairable |
GP-001 GP20140022 | 2017-05-31 | Factory | Disposed | Disposed | Disposed of at Eosense factory because it was not repairable |
GP-001 GP20140023 | 2017-05-31 | Factory | Disposed | Disposed | Disposed of at Eosense factory because it was not repairable |
GP-001 GP20140024 | 2017-05-31 | Factory | Disposed | Disposed | Disposed of at Eosense factory because it was not repairable |
GP-001 GP20140025 | 2017-05-31 | Factory | Disposed | Disposed | Disposed of at Eosense factory because it was not repairable |
GP-001 GP20150100 | 2017-05-31 | Factory | Disposed | Disposed | Disposed of at Eosense factory because it was not repairable |
BB-Fab BB-TW | 2015-12-28 | Twitchell Rice | Upgrade | Twitchell Rice | I added two batteries to the battery bank. I added two batteries from the lab. One form 5/14 and one from 5/11. Both tested okay in the lab. The existing batteries were two from 2/12 and two from 6/14. |
BB-Fab BB-TW | 2016-01-21 | Twitchell Rice | Repair | Twitchell Rice | The solar charge controller was replaced. |
BB-Fab BB-TW | 2016-01-12 | Twitchell Rice | Lost | Twitchell Rice | The old 8D batteries, the solar charge controller were stolen along with the old solar panels that were on top on the big box. Power is down until the charge controller can be replaced. |
LI-7500A 0065 | 2017-05-30 | Lab none | Calibration | Lab none | Last calibration: 2016-12-06 AGC=100 SDM= DAC1= DAC2= Ambient P=99.7kPa, Tair=19.8C, Tdew=10.1C Zero Air at 1050psi, NOAA Air 402.69 at 1600 Tdewspan=16.00 CO2 zero 0.8899 -> .8901 Offset: 1.08ppm CO2 span 1.0142 -> 1.0130 Offset: -.84ppm H2O zero 0.7086 -> .7087 Offset: .6ppm H2O span 1.0127 -> 1.0166 Offset: -.08ppm Notes: Signal strength was 73.9 after thorough cleaning of lenses, so I reset the Signal Strength to 100%. |
LI-7500A 75H-2150 | 2017-05-30 | Lab none | Calibration | Lab none | Last calibration: 2017-03-14 AGC=97.2 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=19.9C, Tdew=12.2C Zero Air at 1050psi, NOAA Air 402.69 at 1600 Tdewspan=16.00 CO2 zero 0.9014 -> .9024 Offset: 4.23ppm CO2 span 1 -> 1.0194 Offset: -9.8ppm H2O zero 0.8651 -> .8664 Offset: 9.3ppm H2O span 1.0094 -> .9975 Offset: .26ppm Notes: |
LI-7500A 1594 | 2017-05-30 | Lab none | Calibration | Lab none | Last calibration: 2017-01-27 AGC=101.4 SDM= DAC1= DAC2= Ambient P=99.7kPa, Tair=19.3C, Tdew=9.5C Zero Air at 1050psi, NOAA Air 402.69 at 1600 Tdewspan=16.00 CO2 zero 0.8654 -> .8653 Offset: .79ppm CO2 span 1.0194 -> 1.0187 Offset: 0.42ppm H2O zero 0.8868 -> .8861 Offset: -4.1ppm H2O span 1.0067 -> 1.0256 Offset: -.39ppm Notes: |
LI-7700 TG1-0423 | 2017-05-31 | Lab none | Calibration | Lab none | Calibration Location: Lab Last calibration: 2016-10-11 RSSI=80.7%, Optical RH=10% Ambient P=100.3kPa, Tair=20.4C Zero Air at 1500psi, NOAA Air 1.89919 at 1600 CH4 Zero Offset: 0.017ppm CH4 Span Offset: -0.021ppm Notes: Replaced desiccant and flushed cavity with zero air. Optics RH was 26% before replacement and 10% after replacement. |
LI-7700 TG1-0211 | 2017-05-31 | Lab none | Calibration | Lab none | Calibration Location: Lab Last calibration: 2016-10-11 RSSI=71.2%, Optical RH=14% Ambient P=100.3kPa, Tair=C Zero Air at 1500psi, NOAA Air 1.89919 at 1600 CH4 Zero Offset: -0.1ppm CH4 Span Offset: -0.002ppm Notes: Replaced temperature component. Replaced desiccant and flushed cavity with zero air. Optics RH was 26% before replacement and 14% afterwards. |
CFM100 4994 | 2017-05-24 | Vaira | Removed | Lab none | Removed from the Vaira Tower. Replaced by the new Licor GHG system from Ameriflux. |
CR1000 23865 | 2017-05-24 | Vaira | Removed | Lab none | Removed from the Vaira Tower. Replaced by the new Licor GHG system from Ameriflux. |
LI-7500A 0073 | 2017-05-24 | Vaira Gas Analyzer CO2/H2O | Removed | Lab none | Removed from the Vaira Tower. Replaced by the new Licor GHG system from Ameriflux. |
1189PC 000192 | 2017-05-24 | Vaira | Removed | Lab none | Removed from the Vaira Tower. Replaced by the new Licor GHG system from Ameriflux. |
WMP 1352 000028 | 2017-05-24 | Vaira Anemometer | Removed | Lab none | Removed from the Vaira Tower. Replaced by the new Licor GHG system from Ameriflux. |
LI-7500 0976 | 2017-05-24 | Tonzi-Tower | Removed | Lab none | Removed from the Tonzi Tower. Replaced by the new Licor GHG system from Ameriflux. |