Equipment and Calibration History

View Equipment |

View history entry 2931 for Licor LI-7700 sn: TG1-0421

Date: 2017-07-13
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Calibration Location: Lab Last calibration: 2017-04-03 RSSI=72.1%, Optical RH=3% Ambient P=100.3kPa, Tair=22.4C Zero Air at 1400psi, NOAA Air 1.89919 at 1500 CH4 Zero Offset: 0.010ppm CH4 Span Offset: 0.0519ppm Notes: First calibration after replacing the circuit power board. Ambient CH4 = 1.98
Entered by: