Equipment and Calibration History

View Equipment |

View history entry 2963 for Licor LI-7500A sn: 75H-2514

Date: 2017-08-14
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-08-03 AGC=102.4 SDM=0 DAC1=0 DAC2=0 Ambient P=100.22kPa, Tair=19.9C, Tdew=13.3C Zero Air at 1100psi, AMP Air 408.97 at 2000 Tdewspan=16.00 CO2 zero 0.9038 -> .9050 Offset: 4.58ppm CO2 span 1.018 -> 1.0107 Offset: 3.72ppm H2O zero 0.8384 -> .8378 Offset: -4.7ppm H2O span 1.0081 -> 1.0094 Offset: 0.11ppm Notes:
Entered by: