Equipment and CalibrationView History
Equipment Entry
Instrument: | Pyranometer |
Manufacture: | Kipp & Zonen |
Model No: | CMP 11 |
Serial No: | 996981 |
Tag No: | |
Acquisition Date: | 1999-09-07 |
Project: | DOE-TCP |
Location on 2023-08-03: | Lab, 23m |
Notes: |
Initial factory calibration: 5.03 uV/(W/m2) |
|
|
Documents
Calibration: cal_KippZonen_CM11_sn996981_19990907.pdf
History
Rec No. | Date | Loc From | Title | Loc To | Notes |
6355 |
2023-08-03 |
Lab none |
Repair |
Lab none |
I took this pyranometer apart and soldered a new short cable to the contacts on the bottom side of the sensor. The wiring colors are the same: red+, blue-, shield. There was a white wire that doesn't seem to be used which I replaced with a yellow so that it is accessible outside the case. It needs new desiccant and probably a longer wire spliced on to be used on a tower. There is a scratch on the outer dome and it needs an inter-comparison. |
3001 |
2017-08-25 |
Tonzi-Tower |
Removed |
Lab none |
Removed from Tonzi Tower and brought back to the lab. It is a redundant measurement with the 4-way radiometer. Cables are decayed and need to be replaced. |
325 |
2001-05-20 |
Lab none |
Installed |
Tonzi-Tower |
Installed on the top of Tonzi Tower. [Gleaned from notes] |
|
|
|