Equipment and Calibration History

View Equipment |

View history entry 2960 for Licor LI-7500A sn: 75H-2514

Date: 2017-08-03
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2017-06-20 AGC=101.4 SDM= DAC1= DAC2= Ambient P=101.3kPa, Tair=24.9C, Tdew=14.3C Zero Air at 2100psi, AMP Air 408.97 at 2100 Tdewspan=16.00 CO2 zero 0.9041 -> .9038 Offset: -1.36ppm CO2 span 1.0115 -> 1.018 Offset: -3.31ppm H2O zero 0.8378 -> .8384 Offset: 4.7ppm H2O span 1.0019 -> 1.0081 Offset: -.34ppm Notes:
Entered by: