Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 [39] 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70
Ser No.DateLoc FromTitleLoc ToNotes
CR6 2115 2018-02-01 Lab none Upgrade Lab none Upgraded OS from CR6.Std.03 to CR6.Std.07.02. The older OS has a known issue where the CR6 is unresponsive until the internal lithium battery is removed and reinstalled. This new O.S. CR6.Std.07.02 requires updating your computer USB drivers using Device Config Utility 2.16 (or newer). I updated Device Config Utility on my computer from version 2.10 to 2.16. In the "Send OS" tab in Device Config Utility, the first step provides a link to update USB drivers. After upgrading the OS, I used the "Network Services" tab to enable FTP with the username tower and the usual password.
CR6 2114 2018-01-31 Factory Returned Lab none Returned from Campbell where it was evaluated, updated and returned free of charge.
CR6 2114 2018-01-22 Factory Maintenance Factory Campbell says: Tested unit on bench, unit functioned upon receipt. No problems found, but unit is running OS Rev3. This Rev is suspected to cause random freezes that require disconnection of the lithium battery. The issue has been resolved in the latest CR6 OSes. Updated the logger to latest Rev 7.02. Performed calibration, tested unit in the environmental chamber, and performed final bench test. Replaced the internal lithium battery and included a new USB cable.
HMP45AC U3030046 2018-01-30 Lab none Notes Lab none This HMP45 needs 3 o-rings replaced before field deployment.
PS-30M 04350908 2018-01-30 Lab none Loaned Loaned Cleaned up and loaned to Patty for her Eden Landing tower.
LI-7550 AIU-1570 2018-01-29 Lab none Upgrade Lab none Downgraded from embedded Licor software v8.7.5 to 8.5.0. Version 8.7.5 was causing issues with header/data mismatch. I checked 7550 time and time zone and CO2 and CH4 logging config.
LI-7500A 0035 2018-01-16 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed at BC to replace sn 75H-2176 (had PLL error)
LI-7550 AIU-0666 2018-01-16 West Pond Interface LI-7550 Upgrade West Pond Interface LI-7550 Downgraded embedded Licor software from v8.7.5 to 8.5.0 because 8.7.5 caused mismatch problems between data columns and headers
LI-7500A 75H-2180 2018-01-16 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed at WP to replace sn 75H-2150.
LI-7500A 75H-2150 2018-01-16 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from WP and brought back to the lab for calibration. it was replaced by sn 75H-2180.
KD140SX-UFBS 11ZV0M0835 2018-01-23 Lab none Installed Mayberry Installed at Mayberry to supply extra winter power.
KD140SX-UFBS 11ZV0M0856 2018-01-23 Lab none Installed Mayberry Installed at Mayberry to supply extra winter power.
LI-7500A 75H-2176 2018-01-19 Lab none Repair Factory This 7500 head has a bad chopper (PLL error). It was sent to Licor on RMA 00205846 for repair and recalibration.
CR6 2114 2018-01-19 Lab none Shipped Factory This CR6 became unresponsive for no know reason - only a green light on the RJ45 jack. It was sent to Campbell on RMA 27680 for evaluation and repair.
CR6 2114 2018-01-15 Lab none Damaged Lab none I tried to install this CR6 at Tonzi for the new co2 profile on the Tower but it was non-functional. It tested fine in the lab this morning, but was a brick in the field. The green light on the RJ45 jack comes on but no other indication that it has power.
eosFD 20160003 2018-01-15 Tonzi Removed Lab none Removed from the soil co2 profile near Tree 1 for return to Eosense.
HMP60 L2310711 2018-01-15 Lab none Installed Tonzi Insalled by a box just below the 3rd lowest location of the tower co2 profile. This sensor will be used to give real time air temperature to the GMP343s.
PTB101B B0310006 2018-01-15 Lab none Installed Tonzi Installed in a box just below the 3rd lowest location of the tower co2 profile. This sensor will be used to give real time pressure to the GMP343s.
GMP343 N2210009 2018-01-15 Lab none Installed Tonzi Installed at the 3rd lowest location in the tower co2 profile - not running yet due to data logger failure.
GMP343 N2210010 2018-01-15 Lab none Installed Tonzi Installed at the 2nd lowest location in the tower co2 profile - not running yet due to data logger failure.
GMP343 N2220007 2018-01-15 Lab none Installed Tonzi Installed at the lowest location in the tower co2 profile - not running yet due to data logger failure.
eosFD 20160004 2017-10-25 Lab none Installed Vaira Forced Diffusion North Installed near the North soil co2 profile at Vaira. This one wouldn't start because it got too hot in the sun last time.
eosFD 20160006 2018-01-15 Vaira Forced Diffusion East Removed Lab none Removed from near the East soil co2 profile for return to Eosense
eosFD 20160005 2018-01-15 Vaira Forced Diffusion West Removed Lab none Removed from near the West soil co2 profile for return to Eosense
eosFD 20160004 2018-01-15 Vaira Forced Diffusion North Removed Lab none Removed from near the North soil co2 profile for return to Eosense
LI-7500A 0035 2018-01-12 Lab none Calibration Lab none Last calibration: 2017-09-05
AGC=98.9 SDM=0 DAC1=0 DAC2=0
Ambient P=101.5kPa, Tair=20.5C, Tdew=10.2C
Zero Air at 250psi, LBL Air 408.97 at 1700 Tdewspan=16.00
CO2 zero 0.9373 -> .9378 Offset: 1.97ppm
CO2 span 1.0084 -> 1.0288 Offset: -10.27ppm
H2O zero 0.8976 -> .8978 Offset: 1.3ppm
H2O span 1.0141 -> 1.0296 Offset: -.31ppm
Notes:
HMP60 G4540021 2018-01-11 Lab none Installed Twitchell Alfalfa 0 Installed on Twitchell Alfalfa for another comparison. Last time the TA HMP60 had spiders in it and a slow fan.
AM16/32A 1236 2018-01-11 Lab none Installed Mayberry Multiplexer Met Installed on the Mayberry CR1000 met data logger to replace the existing AM16/32. There have been spikes in the data from the sensors on the existing multiplexer.
AM16/32A 1430 2018-01-11 Mayberry Multiplexer Met Removed Lab none Removed from the Mayberry Tower where is was on the CR1000 met data logger. It was removed due to spikes in mostly the radiometer channels.
BB-Fab BB-MB 2018-01-11 Mayberry Upgrade Mayberry Six flooded Super Start batteries with 07/2010 stickers were removed due to nightly low voltage power losses. These were replaced by five Outback AMG batteries from 11/2015 last used for the Temporary Towers at East End.
LI-7700 TG1-0224 2018-01-12 Lab none Adjusted Lab none This 7700 has an unlocked reference signal (the reference and signal waveforms were not synced). I talked with James at Licor and changed the Laser Start Temp from 29.8 to 25.5 in the factory settings (Help > About > Factory). To find the new start temperature, I did a Wide Scan (Help > About > Wide Scan > Start), which goes through temperature 0-50 deg C. There is a graph and table of results, where you can pinpoint the reference lock temperature. During the Wide Scan, some of the factory settings changed, so I reset those (changed Delta 1 from 100 back to 41). Licor has the original factory settings on file.
PS-30M 04350908 2018-01-11 Mayberry Removed Lab none Removed from the Mayberry solar system. This controller was full of dead ants and had some discoloring on the circuit board.
AM16-32A 1238 2018-01-10 Lab none Tested Lab none This mux works in 4x16 mode only.
LI-7500A 75H-2180 2018-01-10 Lab none Calibration Lab none Last calibration: 2017-08-16
AGC=97.8 SDM=0 DAC1=0 DAC2=0
Ambient P=101.1kPa, Tair=17.4C, Tdew=10.6C
Zero Air at 250psi, LBL Air 408.97 at 1700 Tdewspan=16.00
CO2 zero 0.9129 -> 0.9147 Offset: 7.92ppm
CO2 span 1.0145 -> 1.0362 Offset: -10.74ppm
H2O zero 0.8871 -> 0.8864 Offset: -4.2ppm
H2O span 1.0164 -> 1.0168 Offset: -0.02ppm
Notes:
LI-7700 TG1-0224 2018-01-10 Lab none Adjusted Lab none Checked the sensor in the lab, and the temperature reading was fine. However, there was an error "Reference signal is UNLOCKED," and CH4 was reading ~0.05 umol/mol. The diagnostics tab showed that the signal and reference waveforms were not synced. I changed the temperature control to manual, adjusted the laser temperature from 30.6 to 26.0 so that the waveforms were synced, and changed the temperature control back to automatic.
LI-7500A 75H-2176 2018-01-04 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from tower because of PLL error (did not hear sensor fan spinning)
LI-7500A 75H-2667 2017-12-19 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed on Bouldin Alfalfa to replace sn 0042 for calibration
LI-7500A 0042 2017-12-19 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab Gas Analyzer CO2/H2O Removed from Bouldin Alfalfa for calibration. Replaced by sn 75H-2667.
HMP45AC T1250012 2017-12-18 Lab none Installed Twitchell Alfalfa Installed at Twitchell Alfalfa for an inter-comparison of HMPs
HMP45AC U3030046 2017-12-18 Lab none Installed Twitchell Alfalfa Installed at Twitchell Alfalfa for an inter-comparison of HMPs
HMP60 L2310711 2017-12-18 Lab none Installed Twitchell Alfalfa Installed at Twitchell Alfalfa for an inter-comparison of HMPs
HMP45AC T2130005 2017-12-18 Lab none Installed Twitchell Alfalfa Installed at Twitchell Alfalfa for an inter-comparison of HMPs
HMP45AC U3030041 2017-12-18 Lab none Installed Twitchell Alfalfa Installed at Twitchell Alfalfa for an inter-comparison of HMPs
HMP45AC W0730036 2017-12-18 Lab none Installed Twitchell Alfalfa Installed at Twitchell Alfalfa for an inter-comparison of HMPs
HMP60 L1740420 2017-12-18 Twitchell Alfalfa 0 Moved Twitchell Alfalfa 0 Installed at Twitchell Alfalfa for an inter-comparison of HMPs
HMP60 G4540021 2017-12-18 Lab none Installed Twitchell Alfalfa 0 Installed at Twitchell Alfalfa for an inter-comparison of HMPs
LI-7550 AIU-1557 2017-12-18 Twitchell Alfalfa Interface LI-7550 Upgrade Twitchell Alfalfa Interface LI-7550 The software on this 7550 box was down graded from v8.7.5 to v8.5.0
LI-7500A 0035 2017-12-18 Twitchell Alfalfa Gas Analyzer CO2/H2O Removed Lab none Removed from Twitchell Alfalfa for calibration. Replaced by sn 75H-2514
LI-7500A 75H-2514 2017-12-18 Lab none Installed Twitchell Alfalfa Gas Analyzer CO2/H2O Installed at Twitchell Alfalfa to replace sn 0035 for calibration
HMP60 L1740420 2017-12-19 Twitchell Alfalfa RH & Temp Moved Twitchell Alfalfa RH & Temp Moved back up to its usual position near the 7500 from the HMP inter-comparison height.
CR23X 2303 2017-12-19 Twitchell Alfalfa Removed Lab none Removed from the HMP inter-comparison at Twitchell Alfalfa
HMP45AC T1250012 2017-12-19 Twitchell Alfalfa Removed Lab none Removed from the HMP inter-comparison at Twitchell Alfalfa
HMP45AC U3030046 2017-12-19 Twitchell Alfalfa Removed Lab none Removed from the HMP inter-comparison at Twitchell Alfalfa
HMP60 L2310711 2017-12-19 Twitchell Alfalfa Removed Lab none Removed from the HMP inter-comparison at Twitchell Alfalfa
HMP45A U2120013 2017-12-19 Twitchell Alfalfa 0 Removed Mayberry RH & Temp Removed from the HMP inter-comparison at Twitchell Alfalfa and returned to Mayberry
HMP60 L2310736 2017-12-19 Twitchell Alfalfa 0 Removed Sherman Wetland RH & Temp Removed from the HMP inter-comparison at Twitchell Alfalfa and returned to Sherman Wetland
HMP45AC T2130005 2017-12-19 Twitchell Alfalfa Removed Lab none Removed from the HMP inter-comparison at Twitchell Alfalfa
HMP60 G4540025 2017-12-19 Twitchell Alfalfa Removed West Pond RH & Temp Removed from the HMP inter-comparison at Twitchell Alfalfa and returned to West Pond
HMP60 G4540023 2017-12-19 Twitchell Alfalfa none Removed East End RH & Temp Removed from the HMP inter-comparison at Twitchell Alfalfa and returned to East End
HMP45AC U3030041 2017-12-19 Twitchell Alfalfa Removed Lab RH & Temp Removed from the HMP inter-comparison at Twitchell Alfalfa
HMP60 G4540021 2017-12-19 Twitchell Alfalfa 0 Removed Lab none Removed from the HMP inter-comparison at Twitchell Alfalfa
HMP45AC W0730036 2017-12-19 Twitchell Alfalfa Removed Lab RH & Temp Removed from the HMP inter-comparison at Twitchell Alfalfa
U24-002-C 11020258 2017-08-02 East End-Temp3 Conductivity Sensor Moved East End Conductivity Sensor The Hobo conductivity sensor was moved with the miniDOT from the T3 Boardwalk Temporary Tower to the Permanent Tower.
miniDOT 1219 2017-11-10 Lab none Installed East End DO Sensor Installed at East End hanging from the north side of the scaffolding
miniDOT 1219 2017-09-13 East End DO Sensor Moved East End-Temp2 DO Sensor Moved from the Permanent Tower to the T2 Closed Canopy Temporary Tower
U24-002-C 11020258 2017-09-13 East End Conductivity Sensor Moved East End-Temp2 Conductivity Sensor Moved from the Permanent Tower to the T2 Closed Canopy Temporary Tower
miniDOT 1219 2017-08-02 East End-Temp3 DO Sensor Moved East End DO Sensor The miniDOT was moved with the Hobo conductivity sensor from the T3 Boardwalk Temporary Tower to the Permanent Tower.
miniDOT 1219 2017-06-23 Lab none Installed East End-Temp3 DO Sensor Installed at the East End T3 Boardwalk Temporary Tower with the Hobo conductivity sensor.
LI-820 CGA-556 2017-12-14 Tonzi Removed Lab none Removed from the Tower co2 profile for cleaning and calibration. The rest of the profile equipment was turned off
eosFD 20160002 2017-12-14 Tonzi Removed Lab none Removed from the Tonzi soil CO2 profile in the Open (East) location. This FD chamber has been reporting zero flux with nearly identical CO2 readings for both cells. The pump seems to be running though.
eosFD 20160002 2017-10-25 Lab none Installed Tonzi Installed near the Open (East) soil CO2 profile for three month experiment
HMP45A U2120013 2017-12-18 Mayberry RH & Temp Removed Twitchell Alfalfa 0 Removed from Mayberry for 24hour side-by-side comparison at Twitchell Alfalfa
HMP60 L2310736 2017-12-18 Sherman Wetland RH & Temp Removed Twitchell Alfalfa 0 Removed from Sherman Wetland for 24hour side-by-side comparison at Twitchell Alfalfa
HMP60 G4540025 2017-12-18 West Pond RH & Temp Removed Twitchell Alfalfa Removed from West Pond for 24hour side-by-side comparison at Twitchell Alfalfa
LI-7550 AIU-0397 2017-12-18 East End Interface LI-7550 Upgrade East End Interface LI-7550 The embedded software was upgraded from v8.0.0 to v8.5.0
LI-7500A 75H-2669 2017-12-18 Lab none Installed East End Gas Analyzer CO2/H2O Installed on East End to replace sn 0041 for calibration. The AIU software was upgrade from v8.0.0 to v8.5.0
LI-7500A 0041 2017-12-18 East End Gas Analyzer CO2/H2O Removed Lab Gas Analyzer CO2/H2O Removed from East End for calibration. Replaced by sn 75H-2669
HMP60 G4540023 2017-12-18 East End RH & Temp Removed Twitchell Alfalfa none Removed from Eat End for a 24hour side-by-side comparison at Twitchell Alfalfa
LI-7500A 75H-2669 2017-12-15 Lab none Calibration Lab none Last calibration: 2017-08-14
AGC=103.3 SDM=0 DAC1=0 DAC2=0
Ambient P=100.7kPa, Tair=22.2C, Tdew=7C
Zero Air at 250psi, LBL Air 408.97 at 1700 Tdewspan=16.00
CO2 zero 0.9187 -> .9187 Offset: .39ppm
CO2 span 1.0144 -> 1.0157 Offset: -.73ppm
H2O zero 0.8988 -> .8988 Offset: .2ppm
H2O span 1.0091 -> 1.0033 Offset: .12ppm
Notes:
LI-7500A 75H-2667 2017-12-15 Lab none Calibration Lab none Last calibration: 2017-08-14
AGC=105.9 SDM=0 DAC1=0 DAC2=0
Ambient P=100.7kPa, Tair=21.3C, Tdew=6.46C
Zero Air at 300psi, LBL Air 408.97 at 1700 Tdewspan=16.00
CO2 zero 0.9418 -> .9415 Offset: -1.14ppm
CO2 span 1.0162 -> 1.0207 Offset: -3.47ppm
H2O zero 0.8392 -> .8372 Offset: -11.7ppm
H2O span 1.0214 -> 1.0166 Offset: .1ppm
Notes:
LI-7500A 75H-2514 2017-12-15 Lab none Calibration Lab none Last calibration: 2017-08-14
AGC=102.8 SDM=0 DAC1=0 DAC2=0
Ambient P=100.7kPa, Tair=22.1C, Tdew=7.9C
Zero Air at 250psi, LBL Air 408.97 at 1700 Tdewspan=16.00
CO2 zero 0.905 -> .9052 Offset: .8ppm
CO2 span 1.0107 -> 1.0118 Offset: -.5ppm
H2O zero 0.8378 -> .8385 Offset: 4.3ppm
H2O span 1.0094 -> 1.0103 Offset: -.01ppm
Notes:
Line Quantum LQ1570 2000-11-04 Lab none Installed Vaira Installed at Vaira with the initial setup of the site.
CR1000 50889 2017-12-11 Lab none Upgrade Lab none Upgraded O.S. from version 27 to version 31.08 (CR1000.Std.31.08.obj). I enabled FTP with the usual username (Tower) and password.
HMP45AC U3030041 2017-12-08 Lab none Repair Lab none The circuit board was corroded and the red wire was broken at the board connection. Joe repaired it. This sensor needs a new o-ring before going outdoors.
HMP60 G4540021 2017-06-23 Lab none Installed East End-Temp3 RH & Temp Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site.
LI-7500A 0418 2017-11-30 Lab none Installed Mayberry Gas Analyzer CO2/H2O Installed at MB to replace sn 75H-2669.
LI-7500A 75H-2669 2017-11-30 Mayberry Gas Analyzer CO2/H2O Removed Lab none Removed from MB and brought back to the lab for calibration.
LI-7550 AIU-1285 2017-11-30 Mayberry Interface LI-7550 Upgrade Mayberry Interface LI-7550 Downgraded from embedded software version 8.7.5 to version 8.5.0. This solves the problem of data-header mismatch we had with version 8.7.5.
LI-7700 TG1-0224 2017-11-30 Lab none Notes Lab none Temperature was reading -9C after we installed it at SW, so we brought it back to the lab...The temperature was fine earlier this week when I calibrated it.
LI-7500A 0073 2017-11-30 Lab none Installed Sherman Wetland Gas Analyzer CO2/H2O Installed at SW to replace sn 75H-2667.
LI-7500A 75H-2667 2017-11-30 Sherman Wetland Gas Analyzer CO2/H2O Removed Lab none Removed from SW and brought back to the lab for calibration
LI-7550 AIU-1569 2017-11-30 Sherman Wetland Interface LI-7550 Upgrade Sherman Wetland Interface LI-7550 Downgraded from embedded software version 8.7.5 to version 8.5.0. This should solve the issue of data-header mismatch that we found in version 8.7.5.
LI-7500A 1594 2017-11-29 Lab none Installed Tonzi-Floor Installed on the Tonzi Floor to replace sn 75H-2180 for calibration.
LI-7500A 75H-2180 2017-11-29 Tonzi Removed Lab none Removed from the Tonzi Floor for calibration. Replaced by sn 1594
LI-7500A 75H-2668 2017-11-29 Lab Gas Analyzer CO2/H2O Floor Installed Tonzi-Tower Gas Analyzer CO2/H2O Floor Installed on the Tonzi Tower to replace sn 75H-2182 for calibration.
LI-7500A 75H-2182 2017-11-29 Tonzi Removed Lab none Removed from the Tonzi Tower for calibration. Replaced by sn 75H-2668
LI-820 CGA-556 2017-11-29 Lab none Installed Tonzi Installed in the Tower CO2 profile system after lab cleaning and calibration
LI-7500A 0065 2017-11-29 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed on the Vaira tower to replace sn 75H-2514 for calibration.
LI-7500A 75H-2514 2017-11-29 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from the Vaira tower for calibration. Replaced by sn 0065
GMM222 L2010067 2017-11-29 Lab none Installed Vaira CO2 probe East Profile -0.04 Installed in the 4cm location in the East profile at Vaira to replace a probe with erratic values.