Equipment and Calibration History

Page: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 [39] 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68
Ser No.DateLoc FromTitleLoc ToNotes
eosFD 20160003 2017-10-18 Lab none Installed Tonzi Installed near the Tree 1 soil CO2 profile for three month experiment
GP-001 GP20140018 2017-08-16 Lab none Installed West Pond CO2 Probe B Installed at WP to replace a previous bad eosGP
LI-7550 AIU-1569 2017-07-14 Sherman Wetland Interface LI-7550 Upgrade Sherman Wetland Interface LI-7550 Upgraded LI-7550 software from 8.0.0 to 8.7.5.
eosFD 20160006 2017-10-18 Lab none Installed Vaira Forced Diffusion East Installed at Vaira near the East soil CO2 profile. The chamber is on a collar, fastened down with cable ties and power with 12V from a DC-DC power converter.
eosFD 20160005 2017-10-18 Lab none Installed Vaira Forced Diffusion West Installed at Vaira near the West soil CO2 profile. The chamber is on a collar, fastened down with cable ties and power with 12V from a DC-DC power converter.
LI-7500A 1594 2017-10-19 Lab none Installed Bouldin Corn Gas Analyzer CO2/H2O Installed on Bouldin Corn to replace sn 75H-2668 for calibration.
LI-7700 TG1-0423 2017-10-23 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-05-31
RSSI=74.6%, Optical RH=3%
Ambient P=101.3kPa, Tair=27.9C
Zero Air at 425psi, AMP Air 1.9497 at 1850
CH4 Zero Offset: -0.055ppm
CH4 Span Offset: 0.0197ppm
Notes:
LI-7500A 75H-2668 2017-10-19 Bouldin Corn Gas Analyzer CO2/H2O Removed Lab none Removed from Bouldin Corn for calibration. Replaced by sn 1594
LI-7550 AIU-1570 2017-08-23 Bouldin Corn Interface LI-7550 Upgrade Bouldin Corn Interface LI-7550 Upgraded the 7550 firmware from v8.0.0 to v8.7.5. I confirmed the 7550 time, time zone, sonic analog settings, and CO2 and CH4 logging config.
LAI-2050 PCA-1546 2017-10-18 Borrowed Returned Lab none Returned by Sebastian Wolf to the lab
LAI-2000 PCA-1478 2017-10-18 Borrowed Returned Lab none Returned by Sebastian Wolf sewolf@ethz.ch to the lab
LAI-2000 PCA-0846 2017-10-18 Borrowed Returned Lab none Returned by Sebastian Wolf sewolf@ethz.ch to the lab
LI-7700 TG1-0424 2017-10-13 Lab none Installed West Pond Gas Analyzer CH4 Installed on West Pond to replace sn TG1-0223 for calibration
LI-7700 TG1-0223 2017-10-13 West Pond Gas Analyzer CH4 Removed Lab none Removed from West Pond for calibration. Replaced by sn TG1-0424
LI-7500A 75H-2150 2017-10-13 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed on West Pond to replace sn 0418 for calibration
LI-7500A 0418 2017-10-13 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from West Pond for calibration. Replaced by sn 75H-2150
LI-7700 TG1-0421 2017-10-16 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-07-13
RSSI=77%, Optical RH=3%
Ambient P=100.4kPa, Tair=29.7C
Zero Air at 500psi, AMP Air 1.9497 at 2000
CH4 Zero Offset: 0.020ppm
CH4 Span Offset: -0.0137ppm
Notes:
LI-7700 TG1-0223 2017-10-16 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-03-23
RSSI=79.2%, Optical RH=2%
Ambient P=100.5kPa, Tair=28.7C
Zero Air at 500psi, AMP Air 1.9497 at 2000
CH4 Zero Offset: -0.015ppm
CH4 Span Offset: -0.0177ppm
Notes:
HFP01 000137 2017-10-13 East End-Temp3 Heat Flux Plate B Removed Lost none Cable removed from East End Boardwalk tower (temporary tower #3) with decommissioning of the site and brought back to the lab. The cable was broken and the sensor is lost to the wetland.
HFP01SC 00245 2017-10-13 East End-Temp3 Heat Flux Plate A Removed Lab none Removed from East End Boardwalk tower (temporary tower #3) with decommissioning of the site. The cable has been chewed, and the cable-head connection needs to be repaired.
HFP01 6649 2017-09-26 East End Heat Flux Plate C Lost Disposed none Cable removed from EE because the cable had been chewed through. The sensor is still lost to the wetland.
PS-15M 12501588 2017-10-16 East End-Temp2 Solar Controller Removed Lab none Removed from the East End Temporary Tower #2 - Closed Canopy with the decommissioning of the site.
LI-7500A 75H-2150 2017-10-11 Lab none Calibration Lab none Last calibration: 2017-05-30
AGC=96.8 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=23.9C, Tdew=5.5C
Zero Air at 500psi, AMP CO2 408.97 at 2000 Tdewspan=16.0
CO2 zero 0.9024 -> 0.9018 Offset: -2.42ppm
CO2 span 1.0194 -> 1.0070 Offset: 6.38ppm
H2O zero 0.8664 -> 0.8649 Offset: -10.1ppm
H2O span 0.9975 -> 1.0090 Offset: 0.24ppm
Notes:
LI-7700 TG1-0424 2017-10-11 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-03-22
RSSI=76.7%, Optical RH=4%
Ambient P=100.5kPa, Tair=23.7C
Zero Air at 500psi, AMP CH4 1.9497 at 2000
CH4 Zero Offset: 0.042ppm
CH4 Span Offset: -0.001ppm
Notes:
LI-7500A 75H-2176 2017-09-26 East End-Temp2 Gas Analyzer CO2/H2O Removed Lab none Removed from East End Temporary Tower 2 - closed canopy with the decommissioning of the tower.
U24-002-C 11020258 2017-09-26 East End-Temp2 Conductivity Sensor Removed Lab none Removed from East End Temporary Tower 2 - Closed Canopy with the decommissioning of the tower.
miniDOT 1219 2017-09-26 East End-Temp2 DO Sensor Removed Lab none Removed from East End Temporary Tower 2 - Close Canopy with the decommissioning of the tower.
CMTD-As4s0 3317052234210 2017-06-23 Lab none Installed East End-Temp3 Solar Controller Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site.
NR-LITE 041556 2017-06-23 Lab none Installed East End-Temp3 Net Radiometer Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site.
PAR-LITE 050596 2017-06-23 Lab none Installed East End-Temp3 PAR Sensor Reflected Installed at the East End Temporary Tower #3 Boardwalk location with the initial setup of the site.
GP-001 GP20160012 2017-09-21 Lab none Installed Vaira CO2 probe West Profile -0.16 Installed in the Vaira West profile at -16cm
GP-001 GP20160117 2017-09-21 Lab none Installed Vaira CO2 probe West Profile -0.08 Installed in the Vaira West profile at -8cm
GP-001 GP20170028 2017-09-21 Lab none Installed Vaira CO2 probe West Profile -0.04 Installed in the Vaira West profile at -4cm
GP-001 GP20160042 2017-09-21 Lab none Installed Vaira CO2 probe West Profile -0.02 Installed in the Vaira West profile at -2cm
GMP222 K1350102 2017-09-21 Vaira CO2 probe East Profile -0.04 Moved Vaira CO2 probe East Profile -0.16 Moved from East profile at -4cm to East profile at -16cm
GMP222 Y2009109 2017-09-21 Vaira CO2 probe East Profile -0.02 Moved Vaira CO2 probe East Profile -0.08 Moved from East profile at -2cm to East profile at -8cm
GMP222 K1310103 2017-09-21 Vaira CO2 probe North Profile -0.04 Moved Vaira CO2 probe East Profile -0.02 Moved from North profile at -4cm to East profile at -2cm
GMT222 G0230062 2017-09-21 Vaira CO2 probe North Profile -0.02 Moved Vaira CO2 probe North Profile -0.16 Moved from North profile at -2cm to North profile at -16cm
GMP222 Y2030049 2017-09-21 Vaira CO2 probe East Profile -0.08 Moved Vaira CO2 probe North Profile -0.02 Moved from East profile at -8cm to North profile at -2cm
GMP222 K1350101 2017-10-05 Lab none Installed Vaira CO2 probe North Profile -0.04 Installed at Vaira in the North profile at the -4cm location
GMP222 K1310102 2017-10-05 Lab none Installed Vaira CO2 probe East Profile -0.04 Installed at Vaira in the East profile at the -4cm location
CMTD-As4s0 3317052234210 2017-10-04 East End-Temp3 Solar Controller Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
KD140SX-UFBS 11ZV0M0855 2017-10-04 East End-Temp3 Solar Panel A Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
KD140SX-UFBS 11ZV0M0836 2017-10-04 East End-Temp3 Solar Panel B Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
CR1000 23865 2017-10-04 East End-Temp3 Data Logger Flux Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
NL115 6761 2017-10-04 East End-Temp3 Interface NL115 Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
CR1000 50889 2017-10-04 East End-Temp3 Data Logger Met Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
AM16-32A 1238 2017-10-04 East End-Temp3 Multiplexer Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
HMP60 G4540021 2017-10-04 East End-Temp3 RH & Temp Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
NR-LITE 041556 2017-10-04 East End-Temp3 Net Radiometer Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
PAR-LITE 050596 2017-10-04 East End-Temp3 PAR Sensor Reflected Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
CS451 20010046 2017-10-04 East End-Temp3 Water Level Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
LI-7700 TG1-0424 2017-10-04 East End-Temp3 Gas Analyzer CH4 Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
LI-7500 0976 2017-10-04 East End-Temp3 Gas Analyzer CO2/H2O Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
WM 1590 120907 2017-10-04 East End-Temp3 Anemometer Removed Lab none Removed with the decommissioning of Temporary Tower #3 - Boardwalk site.
LI-7550 AIU-1570 2017-10-04 Bouldin Corn Interface LI-7550 Removed Lab none Removed from the Bouldin Corn tower for testing in the lab for bad headers and 7700 drop outs during wash cycle. Replaced by sn AIU-0614.
LI-7550 AIU-0614 2017-10-04 Lab none Installed Bouldin Corn Interface LI-7550 Installed on the Bouldin Corn tower to replace sn AIU-1570 for testing in the lab for bad headers and 7700 drop outs during wash cycle.
LI-7700 TG1-0421 2017-10-04 Bouldin Corn Gas Analyzer CH4 Removed Lab none Removed from the Bouldin Corn tower for calibration and testing in the lab for drop outs when washing problem. Replaced by sn TG1-0211.
LI-7700 TG1-0211 2017-10-04 Lab none Installed Bouldin Corn Gas Analyzer CH4 Installed on the Bouldin Corn tower to replace sn TG1-0421 for calibration and testing in the lab for drop outs when washing problem.
CS451 70011236 2017-09-26 East End-Temp2 Water Level Removed Lab none Removed from the East End Temporary Tower 2 - Closed Canopy location with the decommissioning of the site.
LI-7700 TG1-0211 2017-10-02 Lab none Calibration Lab none Calibration Location: Lab
Last calibration: 2017-05-31
RSSI=73.3%, Optical RH=4%
Ambient P=99.9kPa, Tair=28.4C
Zero Air at 550psi, AMP Air 1.9497 at 2050
CH4 Zero Offset: 0.010ppm
CH4 Span Offset: -.0087ppm
Notes:
LAI-2000 PCA-0846 2017-09-29 Lab none Loaned Loaned Loaned to Sebastian Wolf sewolf@ethz.ch for a few weeks for his work a Sagehen.
LAI-2050 PCA-1546 2017-09-29 Lab none Loaned Loaned Loaned to Sebastian Wolf for a few weeks for his work a Sagehen.
LAI-2000 PCA-1478 2017-09-29 Lab none Loaned Loaned Loaned to Sebastian Wolf sewolf@ethz.ch for a few weeks for his work a Sagehen.
KD140SX-UFBS 11ZV0M0828 2017-09-13 East End-Temp2 Solar Panel A Removed Lab none Removed from the East End Temporary Tower 2 - Closed Canopy location with the decommissioning of the site.
KD140SX-UFBS 11ZV0M0829 2017-09-13 East End-Temp2 Solar Panel B Removed Lab none Removed from the East End Temporary Tower 2 - Closed Canopy location with the decommissioning of the site.
BB-Fab BB-TT 2017-09-28 Tonzi Upgrade Tonzi The Tonzi Tower battery bank was rewired to be a 24V system for the upgrade of the solar power system.
MNKID-B KB04594 2017-09-28 Lab none Installed Tonzi Installed on the Tonzi Tower to upgrade the solar system. The solar system is now three parallel strings of four panels in series approximately 12A at 80VDC. The battery bank was rewired to be a 24V system.
PS-30M 0135002X 2017-09-28 Tonzi Removed Lab none Removed from the Tonzi Tower solar system were it was serving as a Low Voltage Disconnect. The system was upgraded with a new MPPT charge controller.
C60 SCC-001 2017-09-28 Tonzi Removed Lab none Removed from the Tonzi Tower solar system. The system was upgraded with a new MPPT charge controller.
AM25T 4852 2017-09-26 East End-Temp2 Multiplexer Thermocouples Removed Lab none Brought back to the lab with the decommissioning of the East End closed canopy temporary tower 2
CR6 2115 2017-09-26 East End-Temp2 Data Logger Removed Lab none Brought back to the lab with the decommissioning of the East End closed canopy temporary tower 2
HFP01 10809 2017-09-26 East End-Temp2 Heat Flux Plate B Removed Lab none Brought back to the lab with the decommissioning of the East End closed canopy temporary tower 2.
PAR-LITE 050597 2017-09-26 East End-Temp2 PAR Sensor Reflected Removed Lab none Brought back to the lab with the decommissioning of the East End closed canopy temporary tower 2
NR-LITE 021029 2017-09-26 East End-Temp2 Net Radiometer Removed Lab none Brought back to the lab with the decommissioning of the East End closed canopy temporary tower 2. Th cable is rotten and needs to be replaced.
HMP45AC U3030046 2017-09-26 East End-Temp2 RH & Temp Removed Lab none Brought back to the lab with the decommissioning of the East End closed canopy temporary tower 2
LI-7550 AIU-0614 2017-09-26 East End-Temp2 Gas Analyzer CO2/H2O Removed Lab none Brought back to the lab with the decommissioning of the East End closed canopy temporary tower 2
LI-7700 TG1-0211 2017-09-26 East End-Temp2 Gas Analyzer CH4 Removed Lab none Brought back to the lab with the decommissioning of the East End closed canopy temporary tower 2
WM 1590 153805 2017-09-26 East End-Temp2 Anemometer Removed Lab none Brought back to the lab with the decommissioning of the East End closed canopy temporary tower 2
LI-7500A 1594 2017-09-27 Lab none Calibration Lab none Last calibration: 2017-05-30
AGC=97.2 SDM=0 DAC1=0 DAC2=0
Ambient P=100.1kPa, Tair=29.4C, Tdew=12.3C
Zero Air at 590psi, AMP Air 408.97 at 2100 Tdewspan=16.00
CO2 zero 0.8653 -> .8638 Offset: -6.17ppm
CO2 span 1.0187 -> 1.0168 Offset: 1.13ppm
H2O zero 0.8861 -> .8878 Offset: 9.5ppm
H2O span 1.0256 -> 1.0047 Offset: .44ppm
Notes: When I checked ffsets after this calibration, CO2 zero offset=-0.64 ppm, H2O zero offset=1.2 mmol/m3, and CO2 span offset=0.21 ppm. Calibrated on LI-7550 with embedded software version 8.7.5.
eosFD FD20160001 2017-09-21 Lab none Installed Twitchell Alfalfa Forced Diffusion Reinstalled at TA after testing in the lab. It is now on its own individual DC-DC converter to step down from 24V to 12V.
CR1000 50890 2017-09-21 Tonzi-Tower Upgrade Tonzi-Tower Upgraded from OS 27 to OS 31.08.
eosFD FD20160001 2017-09-19 Twitchell Alfalfa Forced Diffusion Removed Lab none Removed from the field because it has not been collecting data and some of the data were 0. Brought back to the lab for troubleshooting.
LI-7500A 0042 2017-09-19 Lab none Installed Bouldin Alfalfa Gas Analyzer CO2/H2O Installed at BA tower to replace sn 0073 for calibration.
LI-7500A 0073 2017-09-19 Bouldin Alfalfa Gas Analyzer CO2/H2O Removed Lab none Brought back to the lab for calibration and replaced by sn 0042.
A547 3585 2017-09-13 East End Interface A547 Removed Lab none Removed from EE for testing in the lab because the values were flat.
LI-7500A 0041 2017-09-13 Lab none Installed East End Gas Analyzer CO2/H2O Installed at East End to replace sn 75H-2150 for calibration.
LI-7500A 75H-2150 2017-09-13 East End Gas Analyzer CO2/H2O Removed Lab none Removed from East End for calibration. It was replaced by sn 0041.
CS547A 6494 2017-09-13 East End Conductivity Sensor Removed Lab none Removed from EE because the values were flat.
PAR-LITE 010164 2001-07-26 Factory Calibration Factory Initial factory calibration: 5.28 uV/(umol/m2/s)
GMM222 L2010070 2017-09-14 Lab none Installed Vaira CO2 probe North Profile -0.08 Reinstalled at Vaira in the North profile at -8cm after lab calibration.
GMP222 K1310103 2017-09-14 Lab none Installed Vaira CO2 probe East Profile -0.04 Reinstalled at Vaira in the North profile at -4cm after lab calibration.
GMT222 G0230062 2017-09-14 Lab none Installed Vaira CO2 probe North Profile -0.02 Reinstalled at Vaira in the North profile at -2cm after lab calibration.
GMP222 Y2030049 2017-09-14 Lab none Installed Vaira CO2 probe East Profile -0.08 Reinstalled at Vaira in the East profile at -8cm after lab calibration.
GMP222 K1350102 2017-09-14 Lab none Installed Vaira CO2 probe East Profile -0.04 Reinstalled at Vaira in the East profile at -4cm after lab calibration.
GMP222 Y2009109 2017-09-14 Lab none Installed Vaira CO2 probe East Profile -0.02 Reinstalled at Vaira in the East profile at -2cm after lab calibration.
GMP221 Y0850004 2017-09-14 Lab none Installed Tonzi CO2 Probe Tree2 Profile -0.16 Installed at Tonzi in the Tree 2 profile at -16cm after lab calibration. This is a new installation with new holes dug.
GMT221 F4960020 2017-09-14 Lab none Installed Tonzi CO2 Probe Tree2 Profile -0.08 Installed at Tonzi in the Tree 2 profile at -8cm after lab calibration. This is a new installation with new holes dug.
GMP222 K1350103 2017-09-14 Lab none Installed Tonzi CO2 Probe Tree2 Profile -0.04 Installed at Tonzi in the Tree 2 profile at -4cm after lab calibration. This is a new installation with new holes dug.
GMP222 Z2720014 2017-09-14 Lab none Installed Tonzi CO2 Probe Tree2 Profile -0.02 Installed at Tonzi in the Tree 2 profile at -2cm after lab calibration. This is a new installation with new holes dug.