Ser No.DateLoc FromTitleLoc ToNotes
CS655 17130 2022-11-02 Bouldin Alfalfa-ADEX Soil Moisture Probe Edge A Removed Lab none Removed from about 2m north of the Edge tower at -10cm at the end of the 2022 advection experiment
LI-7550 AIU-0674 2022-11-02 Bouldin Alfalfa-ADEX Interface Edge LI-7550 Removed Lab none Removed from the Edge tower at the end of the 2022 advection experiment at BA
HP-127-1.0-1.3-71P PC4 2022-11-02 Bouldin Alfalfa-ADEX Heat Flux Plate Edge C Removed Lab none Removed from about 2m north of the Edge tower at -10cm at the end of the 2022 advection experiment
HP-127-1.0-1.3-71P PC3 2022-11-02 Bouldin Alfalfa-ADEX Heat Flux Plate Edge B Removed Lab none Removed from about 2m north of the Edge tower at -10cm at the end of the 2022 advection experiment
HFP01 003073 2022-11-02 Bouldin Alfalfa-ADEX Heat Flux Plate Edge A Removed Lab none Removed from about 2m north of the Edge tower at -10cm at the end of the 2022 advection experiment
LI-7500A 0042 2022-11-02 Bouldin Alfalfa-ADEX Gas Analyzer Edge CO2/H2O Removed Lab none Removed from top of the Edge tower at 3.96m alg at the end of the 2022 advection experiment at BA
CR23X 2303 2022-11-02 Bouldin Alfalfa-ADEX Data Logger Edge Met Removed Lab none Removed from the Edge tower where it was collecting 30min averages from soil senors at the end of the 2022 advection experiment at BA
NR-LITE 990353 2022-11-02 Bouldin Alfalfa-ADEX Net Radiometer Bouncer Removed Lab none Removed from top of the Bouncer tower at 3.34m alg at the end of the 2022 advection experiment at BA
SQ75 020854E-6-03-02-06832 2022-11-02 Bouldin Alfalfa-ADEX Solar Panel Bouncer B Removed Lab none Removed from about 10m south of the Bouncer tower at the end of the 2022 advection experiment
SQ75 020854E-6-03-02-06825 2022-11-02 Bouldin Alfalfa-ADEX Solar Panel Bouncer A Removed Lab none Removed from about 10m south of the Bouncer tower at the end of the 2022 advection experiment
PS-15 03220377 2022-11-02 Bouldin Alfalfa-ADEX Solar Controller Bouncer Removed Lab none Removed from the Bouncer tower at the end of the 2022 advection experiment at BA
ML3 M003371 2022-11-02 Bouldin Alfalfa-ADEX Soil Moisture Probe Edge B Removed Lab none Removed from about 2m north of the Edge tower at -10cm at the end of the 2022 advection experiment
CS655 17131 2022-11-02 Bouldin Alfalfa-ADEX Soil Moisture Probe Bouncer A Removed Lab none Removed from about 2m south of the Bouncer tower at -10cm at the end of the 2022 advection experiment
LI-7550 AIU-2312 2022-11-02 Bouldin Alfalfa-ADEX Interface Bouncer LI-7550 Removed Lab none Removed from the Bouncer tower at the end of the 2022 advection experiment at BA
HP-127-1.0-1.3-71P PC2 2022-11-02 Bouldin Alfalfa-ADEX Heat Flux Plate Bouncer C Removed Lab none Removed from about 2m south of the Bouncer tower at -10cm at the end of the 2022 advection experiment
HP-127-1.0-1.3-71P PC1 2022-11-02 Bouldin Alfalfa-ADEX Heat Flux Plate Bouncer B Removed Lab none Removed from about 2m south of the Bouncer tower at -10cm at the end of the 2022 advection experiment
HFP01 003072 2022-11-02 Bouldin Alfalfa-ADEX Heat Flux Plate Bouncer A Removed Lab none Removed from about 2m south of the Bouncer tower at -10cm at the end of the 2022 advection experiment
LI-7500RS 75H-0128 2022-11-02 Bouldin Alfalfa-ADEX Gas Analyzer Bouncer CO2/H2O Removed Lab none Removed from top of the Bouncer tower at 3.96m alg at the end of the 2022 advection experiment at BA
CR23X 2300 2022-11-02 Bouldin Alfalfa-ADEX Data Logger Bouncer Met Removed Lab none Removed from the Bouncer tower where it was collecting 30min averages from soil senors at the end of the 2022 advection experiment at BA
LI-7500A 0073 2022-11-15 Lab Calibration Lab Last calibration: 2022-04-08
AGC=101.2 SDM= DAC1= DAC2=
Ambient P=101.2kPa, Tair=23.2C, Tdew=1.69C
Zero Air at 1850psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00
CO2 zero 0.9163 -> .9165 Offset: .68ppm
CO2 span 0.9983 -> 1.0085 Offset: -7.22ppm
H2O zero 0.8246 -> .8296 Offset: 28.9ppm
H2O span 1.0147 -> 1.0258 Offset: -.25ppm
Notes: Used AMP LI-7550 sn AIU-1039
LI-7500A 0042 2022-11-15 Lab Calibration Lab Last calibration: 2022-04-19
AGC=99.1 SDM= DAC1= DAC2=
Ambient P=101.2kPa, Tair=23.0C, Tdew=2.87C
Zero Air at 1850psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00
CO2 zero 0.9483 -> .9481 Offset: -.73ppm
CO2 span 1.0087 -> 1.0127 Offset: -2.76ppm
H2O zero 0.8444 -> .847 Offset: 16.3ppm
H2O span 1.0191 -> 1.0292 Offset: -.22ppm
Notes: Used AMP LI-7550 sn AIU-1039
LI-7500A 0041 2022-11-15 Lab Calibration Lab Last calibration: 2022-03-30
AGC=93.5 SDM= DAC1= DAC2=
Ambient P=101.2kPa, Tair=22.9C, Tdew=.1C
Zero Air at 1900psi, AMP CA04119 524.77 at 1350 Tdewspan=20.00
CO2 zero 0.9969 -> .9998 Offset: 9.35ppm
CO2 span 1.0189 -> 1.0178 Offset: .23ppm
H2O zero 0.8214 -> .8213 Offset: -.5ppm
H2O span 1.0337 -> 1.0308 Offset: .07ppm
Notes: Reset SS from 93.5 to 100 before calibrating. Used AMP LI-7550 sn AIU-1039.
EXO fDOM 21K102591 2022-11-03 Gilbert Tract Calibration Gilbert Tract fDOM (QSU and RFU), Last cal: 10/31/2022 4:14:35 PM
EXO sn 22C100485
EXO/Sensor Firmware: 1.0.86/3.0.0
UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift.
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,0.51 QSU,0.01 QSU,0.00 QSU,16.967 °C
Pt2,316.50 QSU,219.51 QSU,316.49 QSU,0.00 QSU,10.855 °C

,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,0.12 RFU,0.00 RFU,0.00 RFU,15.878 °C
Pt2,105.50 RFU,74.86 RFU,105.50 RFU,0.00 RFU,11.257 °C
EXO fDOM 21K102592 2022-11-03 Gilbert Tract Calibration Gilbert Tract fDOM (QSU and RFU), Last cal: 10/31/2022 4:14:35 PM
EXO sn 22C100485
EXO/Sensor Firmware: 1.0.86/3.0.0
UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift.
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,1.87 QSU,0.00 QSU,0.00 QSU,16.796 °C
Pt2,316.50 QSU,218.07 QSU,316.49 QSU,0.00 QSU,10.735 °C

,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,0.20 RFU,0.00 RFU,0.00 RFU,15.801 °C
Pt2,105.50 RFU,71.69 RFU,105.50 RFU,0.00 RFU,11.214 °C
EXO fDOM 22C100780 2022-11-03 Gilbert Tract Calibration Gilbert Tract fDOM (QSU and RFU), Last cal: 10/24/2022 8:08:51 AM
EXO sn 22C100485
EXO/Sensor Firmware: 1.0.86/3.0.0
UCSC sensors sn 591 and sn 592 were really off even though calibrated recently. Will accept calibration and just run them at the GT channel for a week to see if they drift.
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,-1.28 QSU,0.00 QSU,0.00 QSU,16.696 °C
Pt2,316.50 QSU,315.87 QSU,316.54 QSU,0.00 QSU,10.444 °C

,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,0.11 RFU,0.00 RFU,0.00 RFU,10.610 °C
Pt2,105.50 RFU,92.32 RFU,105.51 RFU,0.00 RFU,11.275 °C
EXO Turbidity 22C100485 2022-11-01 Lab none Calibration Lab none Turbidity (FNU), Last cal: Unknown
EXO sn 22C100485
EXO/Sensor Firmware: 1.0.86/3.0.4
A factory reset was done first before this calibration was done in the lab.
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 FNU,0.24 FNU,0.00 FNU,0.00 FNU,20.384 °C
Pt2,124.00 FNU,112.11 FNU,123.99 FNU,0.00 FNU,20.389 °C
Pt3,1010.00 FNU,939.77 FNU,1009.99 FNU,0.00 FNU,20.565 °C
EXO pH-ORP 22D102641 2022-09-23 Lab none Calibration Lab none ORP (mV), Last cal: Unknown
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,223.0 mV,220.6 mV,223.0 mV,0.0 mV,25.000 °C
EXO pH-ORP 22D102641 2022-09-23 Lab none Calibration Lab none pH, Last cal: 7/12/2022 2:26:48 PM
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.00 pH,6.96 pH,7.00 pH,-4.1 mV,24.500 °C
Pt2,10.01 pH,10.15 pH,10.01 pH,-181.2 mV,24.500 °C
Pt3,4.01 pH,3.83 pH,4.02 pH,169.1 mV,24.500 °C
EXO Turbidity 22E106072 2022-09-23 Lab none Calibration Lab none Turbidity (FNU), Last cal: Unknown
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.4
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 FNU,-0.19 FNU,0.00 FNU,0.00 FNU,24.000 °C
Pt2,124.00 FNU,124.85 FNU,124.00 FNU,0.00 FNU,24.000 °C
EXO fDOM 22E100649 2022-09-23 Lab none Calibration Lab none fDOM (QSU and RFU), Last cal: Unknown
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,-1.65 QSU,0.00 QSU,0.00 QSU,25.000 °C
Pt2,308.00 QSU,301.85 QSU,307.99 QSU,0.00 QSU,15.000 °C

,Standard,PreCal,PostCal,RawValue,Temp
Pt1,103.00 RFU,98.95 RFU,103.00 RFU,0.00 RFU,15.000 °C
Pt2,0.00 RFU,-0.68 RFU,0.00 RFU,0.00 RFU,25.000 °C
EXO2 22C103416 2022-09-23 Lab none Calibration Lab none Depth (m), Last cal: 6/27/2022 1:49:10 PM
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.1.3
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.000 m,-0.008 m,0.000 m,0.000 m,22.000 °C
EXO pH-ORP 22D102641 2022-07-12 Lab none Calibration Lab none pH, Last cal: Unknown
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
Slope is lower than what it should be for a new sensor,Standard,PreCal,PostCal,RawValue,Temp
Pt1,7.00 pH,7.02 pH,7.00 pH,-1.4 mV,20.000 °C
Pt2,10.00 pH,9.89 pH,10.00 pH,-167.8 mV,20.000 °C
Pt3,4.00 pH,4.39 pH,4.02 pH,151.1 mV,20.000 °C
EXO2 22C103416 2022-06-27 Lab none Calibration Lab none Depth (m), Last cal: Unknown
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.1.3
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.000 m,10.300 m,0.000 m,0.000 m,18.800 °C
PAR-LITE 030382 2022-10-31 Lab none Tested Lab none This sensor was removed because there were spikes in its data on two different mux ports. I pulled apart the splice in the cable and didn't see any problems. From 2022-10-27 to 2022-10-31 I ran an inter-comparison out the lab window with sn 010164. The relationship was tight (r2=0.9999) and the slope was about 1.06. sn 030382 had been inter-compared on the Tonzi tower recently but in the reflected position so the range was not optimal. I think both sensors are functional and should be inter-compared with a good reference in the field to determine appropriate coefficients.
PAR-LITE 010164 2022-10-31 Lab none Repair Lab none One of the wires was broken off right at the base of the sensor. I was able to solder on a lead and spliced a longer cable on and the sensor seemed to respond to light. From 2022-10-27 to 2022-10-31 I ran an inter-comparison out the lab window with sn 030382. The relationship was tight (r2=0.9999) and the slope was about 1.06. sn 030382 had been inter-compared on the Tonzi tower recently but in the reflected position so the range was not optimal. I think both sensors are functional and should be inter-compared with a good reference in the field to determine appropriate coefficients.
CR23X 2143 2022-10-27 Lab none Disposed Disposed none I took this 23x apart and found a surface mount capacitor that had blown. I tried to repair the board trace and replaced the cap with one from a dead CR10x - no way to tell if the components were the same. After putting it back together it was still bricked. We will keep it for parts for a while.
HYDROS21 H21G20001202 2022-10-19 Lab none Installed Gilbert Tract Water Level Installed at GT off of the boardwalk where it is co-located with the eosGP and Campbell conductivity sensor. When the staff gauge read 21cm, the bottom of the sensors was 19cm below the water.
PA-II-SD 84f3ebd5ffb1 2022-10-18 Bouldin Corn Air Quality Sensor Removed Lab none Removed from BC tower and brought back to the lab for further inspection. Data over the summer seemed spiky.
Mark II C013311 2022-10-18 Lab none Installed Bouldin Alfalfa Weather Station C Installed at BA (eastern position) to replace sn C003240, which was removed b/c of a broken T/RH sensor.
Mark II C003240 2022-10-18 Bouldin Alfalfa Weather Station C Removed Lab none Removed from BA (eastern position) because its T/RH sensor is broken. Brought back to the lab to ship back to Arable. It was replaced by sn C013311.
Mark II C013433 2022-10-18 Lab none Installed Bouldin Alfalfa Weather Station B Installed at BA (western position) to replace sn C006164, which was removed b/c of a broken T/RH sensor.
Mark II C006164 2022-10-18 Bouldin Alfalfa Weather Station B Removed Lab none Removed from BA (western position) because its T/RH sensor is broken. Brought back to the lab to ship back to Arable. It was replaced by sn C013433.
OS136-1-MVC NO1107536 2022-10-20 Tonzi none Removed Lab none This sensor has not been sending good data from a while. Seems a woodpecker destroyed its window. Removed and brought in for disposal
SI-131 7385 2022-10-20 Lab none Installed Tonzi Infrared Temperature Ref Panel A Install at the Tonzi cal panel for calibration. But I damaged the data logger so not collecting data until next time.
CR23X 2143 2022-10-20 Lab none Damaged Lab none I tried to install this at Tonzi to collect data from and IRT at the cal panel but fried it with 24V instead of 12V. It flashed "Hello" before something burned on the board.
PAR-LITE 990025 2022-10-20 Lab none Installed Tonzi PAR Sensor Reflected Installed at Tonzi to replace sn 030382 that has been spiky.
PAR-LITE 030382 2022-10-20 Tonzi PAR Sensor Reflected Removed Lab none Removed from Tonzi due to random spikiness. Replaced by sn 990025.
SI-131 7385 2022-10-19 Experiment none Removed Lab none I removed this sensor from the bike kit to calibrate it against the cal panel at Tonzi.
PAR-LITE 990025 2022-10-19 Experiment none Removed Lab none I removed this sensor from the bike kit to use at Tonzi. I also spliced about 3m of extra length on to its cable.
CR23X 2143 2022-10-19 Experiment none Removed Lab none I removed this data logger from the bike kit and prepared it to install at Tonzi to do a calibration on the IRT.
CTD-10 1782303006 2022-10-18 Factory none Disposed Disposed none Apparently the sensor was unrepairable and a new sensor was sent to us a replacement since this sensor was still under warranty. No info on the defect. The new sensor is a new model Hydros21 sn H21G200001202.
CTD-10 1782303006 2022-09-20 Lab none Shipped Factory none After testing in the lab reproduce the errors seen in the field an RMA-008376 was received and this sensors was shipped to METER for repair - it should be under warranty.
SD130BN 0030F4D214A4 2022-10-07 Factory none Returned Lab none Returned from Stardot after inspection determined that a power surge on the ethernet port cooked the innards. Will use the lens on a new body
GMP222 K1310102 2022-10-05 Tonzi-SoilCO2 CO2 Probe Open Profile -0.16 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Open -16cm
GMP221 K1340006 2022-10-05 Tonzi-SoilCO2 CO2 Probe Open Profile -0.08 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Open -8cm
GMP222 Z2720015 2022-10-05 Tonzi-SoilCO2 CO2 Probe Open Profile -0.04 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Open -4cm
GMP222 K1310104 2022-10-05 Tonzi-SoilCO2 CO2 Probe Open Profile -0.02 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Open -2cm
GMP221 Y2009100 2022-10-05 Tonzi-SoilCO2 CO2 Probe Tree1 Profile -0.16 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Tree1 -16cm
GMT221 G0150036 2022-10-05 Tonzi-SoilCO2 CO2 Probe Tree1 Profile -0.08 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Tree1 -8cm
GMT221 G0150035 2022-10-05 Tonzi-SoilCO2 CO2 Probe Tree1 Profile -0.04 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Tree1 -4cm
GMM222 L2010069 2022-10-05 Tonzi-SoilCO2 CO2 Probe Tree1 Profile -0.02 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field: Tree1 -2cm
CR23X 2301 2022-10-05 Tonzi Data Logger Soil CO2 Removed Lab none The soil co2 gradient system has not been in good condition for a long time. Finally getting around to removing the equipment from the field.
SX20U C1031127-2306254 2022-08-24 Lab none Installed Hill Slough none Installed with a CR23X and lithium battery at the Hill Slough channel to collect data from water sensors installed in the channel
GP-001 GP20160127 2022-08-24 Hill Slough CO2 Probe Moved Hill Slough CO2 Probe Moved from the Hill Slough main tower to the channel about 40m east of the main tower. At the tower, the sensor was above water during low tide so we moved the sensor into deeper water where it will remain submerged during low tide.
CTD-10 1782503941 2022-08-24 Hill Slough Water Level Moved Hill Slough Water Level Moved from the Hill Slough main tower to the channel about 40m east of the main tower. At the tower, the sensor was above water during low tide so we moved the sensor into deeper water where it will remain submerged during low tide.
CR23X 2301 2002-06-20 Lab none Installed Tonzi Data Logger Soil CO2 Installed with the initiation of soil CO2 gradient measurements
CR23X 2071 2018-10-11 Lab none Returned Returned none Returned to Robin in Allen Goldstein's lab
miniDOT-usb 931528 2022-10-05 Lab none Installed Hill Slough DO Sensor Installed after cleaning and measurement check in the lab.
EXO pH-ORP 22D102641 2022-09-29 Lab none Installed Hill Slough pH-ORP Channel Installed as part of the EXO2 sonde in the channel at Hill Slough.
EXO fDOM 22E100649 2022-09-29 Lab none Installed Hill Slough fDOM Channel Installed as part of the EXO2 sonde in the channel at Hill Slough.
EXO Turbidity 22E106072 2022-09-29 Lab none Installed Hill Slough Turbidity Channel Installed as part of the EXO2 sonde in the channel at Hill Slough.
EXO pH-ORP 22D102641 2022-09-16 Hill Slough pH-ORP Channel Removed Lab none Removed as part of the EXO2 channel deployed in the channel at HS. Brought back to the lab for cleaning and calibration.
EXO fDOM 22E100649 2022-09-16 Hill Slough fDOM Channel Removed Lab none Removed as part of the EXO2 channel deployed in the channel at HS. Brought back to the lab for cleaning and calibration.
EXO Turbidity 22E106072 2022-09-16 Hill Slough Turbidity Channel Removed Lab none Removed as part of the EXO2 sonde deployed in the channel at HS. Brought back to the lab for cleaning and calibration.
EXO pH-ORP 22D102641 2022-07-13 Lab none Installed Hill Slough pH-ORP Channel Installed as part of the EXO2 sonde at HS in the channel off of the boardwalk, about 30m east of the tower.
EXO fDOM 22E100649 2022-07-13 Lab none Installed Hill Slough fDOM Channel Installed as part of the EXO2 sonde at HS in the channel off of the boardwalk, about 30m east of the tower.
EXO Turbidity 22E106072 2022-07-13 Lab none Installed Hill Slough Turbidity Channel Installed as part of the EXO2 sonde at HS in the channel off of the boardwalk, about 30m east of the tower.
miniDOT-usb 931528 2022-09-29 Hill Slough DO Sensor Removed Lab none Removed from the HS channel and brought back to the lab for cleaning. There are calcareous deposits (some sort of encrusting worm) on and around the sensing foil. I soaked it in distilled white vinegar for 3 days. The deposits came off easily with a soft toothbrush. I rinsed the sensor well under DI water.
EXO2 22C103416 2022-09-29 Lab none Installed Hill Slough Water Level Installed in the channel at Hill Slough. The sonde is now wired to the CR23X, which queries it for data every 10sec. The sonde is measureing and storing data internally every 5min.
HMP45AC W0510026 2011-02-03 Tonzi Removed Lab none Removed from under the trees near the floor tower.
NR-LITE 041556 2010-03-16 Lab none Installed Tonzi none Installed at the request of Hideki in an unshaded opening near the east end of the radiation tram.
EXO2 22C103416 2022-09-16 Hill Slough Removed Lab none Removed from the channel at Hill Slough. It was brought back to the lab for cleaning, calibration, and to convert its output to SDI-12 so we can wire it to the channel CR23X.
LI-7500A 75H-2667 2022-09-15 Lab none Installed East End Gas Analyzer CO2/H2O Installed at East End to replace sn 0041 for calibration.
LI-7500A 0041 2022-09-15 East End Gas Analyzer CO2/H2O Removed Lab none Removed from East End for calibration. Replaced by sn 75H-2667.
LI-7500A 0035 2022-09-15 Lab none Installed West Pond Gas Analyzer CO2/H2O Installed at West Pond to replace sn 0073 for calibration.
LI-7500A 0073 2022-09-15 West Pond Gas Analyzer CO2/H2O Removed Lab none Removed from West Pond for calibration. Replaced by sn 0035.
S2-411 1050 2022-09-15 Gilbert Tract Removed Lab none Removed from GT tower after 1 month of intercomparison with the SRS NDVI sensors
S2-411 1049 2022-09-15 Gilbert Tract Removed Lab none Removed from GT tower after 1 month of intercomparison with the SRS NDVI sensors
LI-7500A 0065 2022-09-14 Lab none Installed Vaira Gas Analyzer CO2/H2O Installed at Vaira to replace sn 75H-2667 for calibration.
LI-7500A 75H-2667 2022-09-14 Vaira Gas Analyzer CO2/H2O Removed Lab none Removed from Vaira for calibration. Replaced by sn 0065.
CR23X 2143 2022-08-10 Lab none Installed Experiment none Installed on the back of Robert's e-bike for his urban heat island transect in Oakley
LI-6400 PSC-0785 2022-09-13 Borrowed none Returned Lab none LI-6400, all components, and batteries were returned from Rizwan.
HFP01 000139 2001-01-05 Lab none Installed Tonzi Heat Flux Plate B Likely first installation date and location estimated from notes and data.
HFP01 000138 2001-01-05 Lab none Installed Tonzi Heat Flux Plate A Likely first installation date and location estimated from notes and data.
HFP01 000140 2001-01-05 Lab none Installed Tonzi Heat Flux Plate C Likely first installation date and location estimated from notes and data.
HFP01 000144 2001-01-05 Lab none Installed Tonzi Heat Flux Plate E Likely first installation date and location estimated from notes and data.
HFP01SC 00068 2001-01-05 Lab none Installed Tonzi Heat Flux Plate D Likely first installation date and location estimated from notes and data.
LI-7500A 0418 2022-09-12 Lab Calibration Lab Last calibration: 2022-08-30
AGC=94.7 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=21.8C, Tdew=15.1C
Zero Air at 1900psi, AMP CA04119 524.77 at 1400 Tdewspan=20.0
CO2 zero 0.9082 -> 0.9086 Offset: 1.4ppm
CO2 span 1.0000 -> 1.0050 Offset: -3.47ppm
H2O zero 0.9313 -> 0.9317 Offset: 2.4ppm
H2O span 1.0008 -> 1.0065 Offset: -0.12ppm
Notes: This head had a drift of SS (and CO2, H2O) after refit by Licor in March. It was calibrated/tested in the lab several times in August. Full calibration on 8/30 but not all the information written down. 8/30 coefficients entered as 'existing' and it seems stable since then. We will try it in the field again.
LI-7500A 0065 2022-09-12 Lab Calibration Lab Last calibration: 2022-04-08
AGC=101.0 SDM=0 DAC1=- DAC2=-
Ambient P=100.4kPa, Tair=22.8C, Tdew=15.2C
Zero Air at 1900psi, AMP CA04119 524.77 at 1400 Tdewspan=20.0
CO2 zero 0.8908 -> 0.8918 Offset: 4.5ppm
CO2 span 1.0198 -> 1.0227 Offset: -1.97ppm
H2O zero 0.7308 -> 0.7316 Offset: 6.7ppm
H2O span 1.016 -> 1.0167 Offset: -0.01ppm
Notes:
SRS-Pi PRI 913003012 2022-09-12 Lab none Tested Lab none Tested in the lab with the lab CR1000. Could not connect to datalogger when this sensor was plugged in, even though the datalogger works with other sensors. This sensor is bad and probably not repairable.