View history entry 6049 for Kipp & Zonen PAR-LITE sn: 010164

Date: 2022-10-31
From: Lab - none
To: Lab - none
Title: Repair
Notes: One of the wires was broken off right at the base of the sensor. I was able to solder on a lead and spliced a longer cable on and the sensor seemed to respond to light. From 2022-10-27 to 2022-10-31 I ran an inter-comparison out the lab window with sn 030382. The relationship was tight (r2=0.9999) and the slope was about 1.06. sn 030382 had been inter-compared on the Tonzi tower recently but in the reflected position so the range was not optimal. I think both sensors are functional and should be inter-compared with a good reference in the field to determine appropriate coefficients.