Equipment and Calibration History

View Equipment |

View history entry 5774 for Licor LI-7500A sn: 1594

Date: 2022-04-08
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2021-11-30 AGC=96.7 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=30.8C, Tdew=11.2C Zero Air at 725psi, AMP CA04119 524.77 at 1600 Tdewspan=20.00 CO2 zero 0.8645 -> .8643 Offset: -.71ppm CO2 span 1.0128 -> 1.0097 Offset: 2.12ppm H2O zero 0.8877 -> .8891 Offset: 7.7ppm H2O span 1.0045 -> 1.0106 Offset: -.13ppm Notes:
Entered by: