Equipment and Calibration History

View Equipment |

View history entry 5771 for Gill WM 1590 sn: 153804

Date: 2022-04-06
From: Lab - none
To: Lab - none
Title: Tested
Notes: I tested this sensor briefly in the lab against sn 152601. U and V had a much smaller magnitude. I checked the analog settings and they had been set to 5m/s range. I changed U and V from 5m/s to 30m/s. I left W as 5m/s. U, V, W, SOS seem reasonable. I will be moving this intercomparison to my backyard tomorrow to run for a few days outdoors.
Entered by: Daphne Szutu