Skip to main content
rice field banner biometlab logo
   
   
 
   
   
   
   
   
 
 

Equipment and Calibration

View History

Equipment Entry

Instrument:Moisture Point
Manufacture:Environmental Sensors Inc.
Model No:PR K Type 8
Serial No:00000FED7C67
Tag No:NULL
Acquisition Date:2007-02-23
Project:NULL
Location on 2010-09-03:Disposed, NULL
Notes: Received with sn FEDD01B, FEDDAA1, FED8688

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
197 2010-09-03 Factory Disposed Disposed Report from vendor not repairable:
s/n 00000FED7C67
Condition on arrival: several cracks in segment 1, part of one rail partially corroded away. Resistance checks fail(open circuits in RF signal, RF gnd, ctrl gnd and diode pins). No valid diode waveforms from any of the 5 diode boards. No Id reading from probe. Probe scans fail(all segments return error readings).
173 2010-07-07 Lab none Repair Factory Sent to ESI - Environmental Sensors Inc (www.esica.com)for evaluation and possible repair.
150 2010-06-24 Sherman Island Removed Lab none This probe was removed from the Sherman Island cow pasture. It was giving bad readings.
 
 