Equipment and Calibration History

View Equipment |

View history entry 6215 for Licor LI-7550 sn: AIU-1569

Date: 2023-03-14
From: Lab - none
To: Lab - none
Title: Tested
Notes: I tested this box in the lab. It had been removed from the field because the door was left open and it got wet inside. I had cleaned it up earlier but I'm not sure if I ran it before. Today, it was running but very slowly/laggy. At first I was concerned it was the new laptop but it was the same on the old laptop. I pulled the shroud off and found a little corrosion on one of the ribbon cable connectors. I cleaned this off and it seems to be working ok.
Entered by: Joe Verfaillie