Equipment and Calibration History

View Equipment |

View history entry 6213 for Licor LI-7500A sn: 1594

Date: 2023-03-14
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2023-01-25 AGC=100.7 SDM=0 DAC1=- DAC2=- Ambient P=99.1kPa, Tair=19.3C, Tdew=12.77C Zero Air at 1400psi, AMP CA04119 524.77 at 1250 Tdewspan=16.0 CO2 zero 0.8656 -> 0.8656 Offset: 0.04ppm CO2 span 1.0205 -> 1.0225 Offset: -1.37ppm H2O zero 0.8869 -> 0.8885 Offset: 9.0ppm H2O span 1.0127 -> 1.0196 Offset: -0.14ppm Notes: Using AMP loaner AIU-2312
Entered by: