Equipment and Calibration History

View Equipment |

View history entry 2451 for Licor LI-7500A sn: 75H-2182

Date: 2016-08-30
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2016-04-11 AGC=87.48 SDM=6 DAC1=- DAC2=- Ambient P=100.13kPa, Tair=22.36C, Tdew=13.81C Zero Air at 400psi, God CO2 421.22 at 450 Tdewspan=16.0 CO2 zero 0.8832 -> 0.8831 Offset: -0.53ppm CO2 span 1.0198 -> 1.0142 Offset: 2.98ppm H2O zero 0.8466 -> 0.8464 Offset: -0.9ppm H2O span 1.0138 -> 1.0230 Offset: -0.19ppm Notes: Bandwidth setting changed from 5Hz to 10Hz
Entered by: