Skip to main content
rice field banner biometlab logo
   
   
 
   
   
   
   
   
 
 

Equipment and Calibration

View History

Equipment Entry

Instrument:Anemometer
Manufacture:Gill
Model No:WM 1590
Serial No:153806
Tag No:NULL
Acquisition Date:2015-09-24
Project:CDFW
Location on 2022-03-07:Loaned, NULL
Notes: 4x14-bit analog outputs, 20Hz, Firmware version 2326-701-03

Documents

No documents found

History

Rec No.DateLoc FromTitleLoc ToNotes
5716 2022-03-07 Lab none Installed Loaned none Installed on the urban flux tower on top of Berkeley Way West building
5386 2021-07-20 Bouldin Alfalfa Anemometer Removed Lab none Removed from Bouldin Alfalfa and replaced by sn 074417 so that the advection experiment could run with four inter-compared sonics.
2486 2016-10-06 Lab none Installed Bouldin Alfalfa Anemometer Reinstalled on the Bouldin Alfalfa tower replacing sn 124215. This sonic is working correctly. Spikes were caused by the soil moisture probes powering up and browning out the system.
2438 2016-08-31 Bouldin Alfalfa Anemometer Removed Lab none Removed from the Bouldin Alfalfa tower due to daytime spiking problem likely due to new firmware. Replaced by sn 124215
2420 2016-08-12 Lab none Installed Bouldin Alfalfa Anemometer Installed with the initial set up of the site.
2392 2016-08-11 Lab none Upgrade Lab none Using the Gill provided software tool we upgraded this unit's firmware from version 2329-601-03 to version 2329-701-01. This new firmware is supposed to fix the W bug.
 
 

Login