Equipment and Calibration History

View Equipment |

View history entry 2452 for Licor LI-7500A sn: 75H-2180

Date: 2016-08-30
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2016-02-09 AGC=100.3 SDM=6 DAC1=- DAC2=- Ambient P=100.13kPa, Tair=23.80C, Tdew=13.90C Zero Air at 400psi, God CO2 421.22 at 450 Tdewspan=16.0 CO2 zero 0.9145 -> 0.9142 Offset: -1.25ppm CO2 span 1.0203 -> 1.0227 Offset: -1.22ppm H2O zero 0.8847 -> 0.8858 Offset: 7.0ppm H2O span 1.0241 -> 1.0211 Offset: 0.06ppm Notes:
Entered by: