Equipment and Calibration History

View Equipment |

View history entry 2450 for Licor LI-7500A sn: 1594

Date: 2016-08-30
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2016-02-22 AGC=101.4 SDM=6 DAC1=- DAC2=- Ambient P=100.12kPa, Tair=24.79C, Tdew=13.72C Zero Air at 400psi, God CO2 421.22 at 400 Tdewspan=16.0 CO2 zero 0.865 -> 0.8648 Offset: -0.70ppm CO2 span 1.0196 -> 1.0164 Offset: 1.68ppm H2O zero 0.8863 -> 0.8859 Offset: -2.6ppm H2O span 1.0139 -> 1.0113 Offset: 0.05ppm Notes:
Entered by: