Equipment and Calibration History

View Equipment |

View history entry 5171 for Licor LI-7500A sn: 75H-2180

Date: 2021-03-08
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2020-11-24 AGC=99.7 SDM= DAC1= DAC2= Ambient P=100.6kPa, Tair=25.7C, Tdew=3.7C Zero Air at 400psi, AMP CA3923 474.19 at 550 Tdewspan=20.00 CO2 zero 0.9138 -> .9138 Offset: .18ppm CO2 span 1.0295 -> 1.0231 Offset: 3.95ppm H2O zero 0.8911 -> .8923 Offset: 7.9ppm H2O span 1.0245 -> 1.0243 Offset: 0ppm Notes: Larger H2O zero offset than I expected
Entered by: